双能X线吸收法随访骨密度检测中的扫描仪不匹配
该研究基于加拿大阿尔伯塔省人群数据库,纳入2009年4月1日至2018年12月31日期间完成至少两次DXA扫描的患者,评估随访DXA扫描中扫描仪不匹配的发生比例。结果显示18.9%的随访DXA扫描存在扫描仪不匹配,28.7%的患者至少经历一次扫描仪不匹配。
要点
- 18.9%的随访DXA扫描存在扫描仪不匹配
- 28.7%的患者至少经历一次扫描仪不匹配
- 扫描间隔较长和大骨松性骨折史与扫描仪不匹配风险升高相关
核心结论
- 扫描仪不匹配是常见问题,约五分之一的随访DXA扫描发生扫描仪不匹配,影响超过四分之一的患者。
- 作者认为有必要采取干预措施来减少这一较大比例的扫描仪不匹配。
研究设计与样本
- 研究使用基于人群的卫生服务数据库,纳入2009年4月1日至2018年12月31日在加拿大阿尔伯塔省完成至少两次DXA扫描的所有患者。
- 扫描仪不匹配定义为随访DXA扫描所使用的扫描仪与上一次不同且未与上一次扫描仪进行交叉校准。
- 采用多变量逻辑回归模型评估可能促成扫描仪不匹配的预测因素。
主要发现
- 共识别264,866例患者,对应470,641次随访DXA扫描。
- 扫描仪不匹配发生于18.9%的随访DXA扫描,28.7%的患者至少经历一次扫描仪不匹配。
- 扫描间隔时间较长与扫描仪不匹配风险升高相关(OR 1.25,95%CI 1.24-1.26)。
- 索引扫描前有大骨松性骨折史与扫描仪不匹配风险升高相关(OR 1.06,95%CI 1.03-1.08)。
- 索引扫描前使用骨质疏松药物(OR 0.89,95%CI 0.88-0.91)、随访扫描较近期(OR 0.98,95%CI 0.73-0.98)、女性(OR 0.97,95%CI 0.94-1.00)和末次扫描年龄(OR 0.99,95%CI 0.99-1.00)与扫描仪不匹配风险降低相关。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
UNLABELLED: Scanner mismatch occurs frequently with follow-up dual-energy x-ray absorptiometry (DXA) scans. Nearly one-in-five follow-up DXA scans were conducted on non-cross-calibrated scanners (scanner mismatch) and more than a quarter of patients who had a follow-up DXA scan had experienced scanner mismatch. INTRODUCTION: Detecting significant changes in bone mineral density (BMD) with dual-energy x-ray absorptiometry (DXA) scanners relies on the least significant change (LSC). Results from two different DXA scanners can only be compared, albeit with decreased sensitivity for change, if the LSC between the two scanners has been directly determined through cross-calibration. Performing follow-up DXA scans on non-cross-calibrated scanners (scanner mismatch) has safety and economic implications. This study aims to determine the proportion of scanner mismatch occurring at a population level. METHODS: All patients who completed at least two DXA scans between 1 April 2009 and 31 December 2018 in the province of Alberta, Canada, were identified using population-based health services databases. Scanner mismatch was defined as a follow-up DXA scan completed on a DXA scanner that differed from and was not cross-calibrated to the previous DXA scanner. Multivariate logistic regression models were used to assess predictive factors that may contribute to scanner mismatch. RESULTS: A total of 264,866 patients with 470,641 follow-up DXA scans were identified. Scanner mismatch occurred in 18.9% of follow-up DXA scans; 28.7% of patients experienced at least one scanner mismatch. Longer duration between scans (OR 1.25, 95% CI 1.24-1.26) and major osteoporotic fracture history before index scan (OR 1.06, 95% CI 1.03-1.08) increased risk of scanner mismatch. Osteoporosis medication use before index scan (OR 0.89; 95% CI 0.88-0.91), recency of follow-up scans (OR 0.98, 95% CI 0.73-0.98), female sex (OR 0.97, 95% CI 0.94-1.00), and age at last scan (OR 0.99, 95% CI 0.99-1.00) were associated with lower risk of scanner mismatch. CONCLUSION: Scanner mismatch is a common problem, occurring in one-in-five follow-up DXA scans and affecting more than a quarter of patients. Interventions to reduce this large proportion of scanner mismatch are necessary.