文献速递关节外科

术中压力传感器改善全膝关节置换术中的软组织平衡但未改善临床结局:一项多中心随机对照试验

Intraoperative pressure sensors improve soft-tissue balance but not clinical outcomes in total knee arthroplasty: a multicentre randomized controlled trial.
PubMed 文献 · Bone Joint J · 2022-01-01

这项多中心随机对照试验比较了传感器平衡与手动平衡在全膝关节置换术中的效果,共纳入250例患者(285膝)。主要结局ΔKOOS4在两组间无显著差异(平均差0.4分,95%CI -4.6至5.4;p=0.869)。传感器组六分钟步行距离显著增加(平均差29米;p=0.015),手动平衡组不平衡膝数量为传感器组的四倍(36.8% vs 9.4%;p<0.001)。

要点

  • 传感器平衡与手动平衡在两年ΔKOOS4上无显著差异(平均差0.4分;p=0.869)。
  • 传感器组六分钟步行距离显著增加(平均差29米;p=0.015)。
  • 手动平衡组不平衡膝比例显著高于传感器组(36.8% vs 9.4%;p<0.001)。

核心结论

  • 尽管定量软组织平衡得到改善,术中传感器使用并未在TKA后两年差异化改善临床或功能结局。

研究设计与样本

  • 多中心随机对照试验,比较传感器平衡与手动平衡,纳入250例患者(285膝TKA)。

主要发现

  • 两年ΔKOOS4两组间无显著差异(平均差0.4分(95%CI -4.6至5.4);p=0.869)。
  • 多元回归显示传感器平衡与ΔKOOS4无显著关联(增加0.2分(95%CI -5.1至4.6);p=0.924)。
  • 其他患者报告结局指标两组间无显著差异。
  • 六分钟步行距离传感器组显著增加(平均差29米;p=0.015)。
  • 手动平衡组不平衡TKA数量为传感器组四倍(36.8% MB vs 9.4% SB;p<0.001)。

关键数值与推荐

  • 主要结局为术前与两年KOOS四个子量表平均差异(ΔKOOS4)。
  • 无论分组如何,提高ICPD阈值定义平衡时,任何PROM均无差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: Intraoperative pressure sensors allow surgeons to quantify soft-tissue balance during total knee arthroplasty (TKA). The aim of this study was to determine whether using sensors to achieve soft-tissue balance was more effective than manual balancing in improving outcomes in TKA. METHODS: A multicentre randomized trial compared the outcomes of sensor balancing (SB) with manual balancing (MB) in 250 patients (285 TKAs). The primary outcome measure was the mean difference in the four Knee injury and Osteoarthritis Outcome Score subscales (ΔKOOS4) in the two groups, comparing the preoperative and two-year scores. Secondary outcomes included intraoperative balance data, additional patient-reported outcome measures (PROMs), and functional measures. RESULTS: There was no significant difference in ΔKOOS4 between the two groups at two years (mean difference 0.4 points (95% confidence interval (CI) -4.6 to 5.4); p = 0.869), and multiple regression found that SB was not associated with a significant ΔKOOS4 (0.2-point increase (95% CI -5.1 to 4.6); p = 0.924). There were no significant differences between groups in other PROMs. Six-minute walking distance was significantly increased in the SB group (mean difference 29 metres; p = 0.015). Four-times as many TKAs were unbalanced in the MB group (36.8% MB vs 9.4% SB; p < 0.001). Irrespective of group assignment, no differences were found in any PROM when increasing ICPD thresholds defined balance. CONCLUSION: Despite improved quantitative soft-tissue balance, the use of sensors intraoperatively did not differentially improve the clinical or functional outcomes two years after TKA. These results question whether a more precisely balanced TKA that is guided by sensor data, and often achieved by more balancing interventions, will ultimately have a significant effect on clinical outcomes. Cite this article: Bone Joint J 2022;104-B(5):604-612.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. MacDessi SJ, Wood JA, Diwan A, et al. Intraoperative pressure sensors improve soft-tissue balance but not clinical outcomes in total knee arthroplasty: a multicentre randomized controlled trial. Bone Joint J. 2022. PMID: 35491577. doi:10.1302/0301-620X.104B5.BJJ-2021-1299.R2

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。