PubMed 文献 · Bone Joint J · 2022-01-01

本研究比较了两种金属对金属(MoM)铰链全膝关节置换(TKA)系统及一种氮化钛(TiN)涂层变体的血清钴和铬离子浓度。共63例患者(65个植入物)随访至少12个月,GenuX组钴和铬离子水平显著高于BPKS组(钴:16.3 vs 9.4 µg/l;铬:9.5 vs 5.2 µg/l),TiN涂层未见改善金属离子水平。结论认为MoM铰链TKA导致临界高水平钴和铬离子浓度。

要点

  • GenuX组平均钴和铬离子水平显著高于BPKS组(钴:16.3 vs 9.4 µg/l;铬:9.5 vs 5.2 µg/l)。
  • GenuX组81%(29/36)患者金属离子浓度超过7 µg/l,BPKS组为41%(11/27)。
  • TiN涂层植入物未显示改善金属离子水平。

核心结论

  • 使用MoM铰链的铰链TKA导致临界高水平钴和铬离子浓度。
  • BPKS铰链TKA相比GenuX TKA显示显著更低的金属离子浓度。
  • 使用TiN涂层未观察到益处。
  • 不同的承重力学可能影响组件材料的磨损。
  • 更高的工作负荷和体力活动可能影响铬水平。

研究设计与样本

  • 比较两种不同MoM铰链TKA系统及一种TiN涂层变体的血清钴和铬离子浓度的研究。
  • 共63例患者(65个植入物),接受MoM涂层(n=29)或TiN涂层(n=7)铰链TKA(GenuX mobile bearing, MUTARS; Implantcast, Germany) versus BPKS(Brehm, Germany)铰链TKA(n=27)。
  • 血清钴和铬离子浓度在至少12个月随访后评估,同时评估功能结果和生活质量。

主要发现

  • 组间平均年龄(69岁,40至86)、平均年龄调整Charlson合并症指数(3.1(SD 1.4))、平均BMI(29.2 kg/m2(SD 5.8))或其他植入物数量无差异。
  • 所有组的结果评分和疼痛水平均显著改善,生活质量(12项简短健康调查问卷(SF-12))无差异。
  • GenuX组平均钴和铬离子水平显著高于BPKS组(GenuX vs BPKS:钴:16.3 vs 9.4 µg/l;铬:9.5 vs 5.2 µg/l)。
  • TiN涂层植入物未显示改善金属离子水平。
  • 金属离子浓度超过7 µg/l在GenuX组为81%(29/36),BPKS组为41%(11/27)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant. METHODS: A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life. RESULTS: No differences in mean age (69 years, 40 to 86), mean age adapted Charlson Comorbidity Index (3.1 (SD 1.4)), mean BMI (29.2 kg/m2 (SD 5.8)), or number of other implants were observed between groups. Significant improvements in outcome scores and pain levels were achieved for all groups, and there was no difference in quality of life (12-Item Short-Form Health Survey questionnaire (SF-12)). Mean cobalt and chromium ion levels were significantly higher for the GenuX versus the BPKS hinged TKA (GenuX vs BPKS: cobalt: 16.3 vs 9.4 µg/l; chromium: 9.5 vs 5.2 µg/l). The TiN-coated implants did not appear to confer improvement in the metal ion levels. Metal ion concentrations above 7 µg/l were detected in 81%(29/36) of GenuX patients versus 41% (11/27) in the BPKS group. No GenuX patients had normal levels under 2 µg/l, versus 22% of BPKS patients. No significant reduction in outcome scores was observed regardless of the metal ion levels, whereas higher work-related activity was correlated with higher chromium concentrations. CONCLUSION: Hinged TKA, using MoM hinges, resulted in critically high cobalt and chromium ion concentrations. The BPKS hinged TKA showed significantly lower metal ion concentrations compared with the GenuX TKA. No benefits were observed using TiN coating. The different weightbearing mechanics might influence the wear of the component materials. Higher workloads and physical activity could influence chromium levels. Cite this article: Bone Joint J 2022;104-B(3):376-385.