文献速递关节外科

全髋关节置换术中氧化锆与钴铬关节配对的聚乙烯磨损十年放射立体分析

A Ten-Year Radiostereometric Analysis of Polyethylene Wear Between Oxidized Zirconium and Cobalt Chrome Articulations in Total Hip Arthroplasty.
PubMed 文献 · J Arthroplasty · 2022-03-01

本研究采用放射立体分析(RSA)比较氧化锆(OxZi)/高交联聚乙烯(XLPE)与钴铬(CoCr)/XLPE两种关节配对在全髋关节置换术后至少10年的聚乙烯磨损率,两组各23例患者按年龄、性别、体重指数和诊断匹配。结果显示全队列总头端穿透率差异有统计学意义(OxZi 0.048±0.021 mm/y vs CoCr 0.035±0.017 mm/y,P=.02),但仅28-mm头亚组无统计学差异(P=.066);稳态磨损率在全队列和28-mm头亚组均无显著差异。除OxZi组Harris髋关节评分中位数较高外,患者报告结局无统计学差异。

要点

  • 全队列总头端穿透率OxZi组0.048±0.021 mm/y vs CoCr组0.035±0.017 mm/y,差异有统计学意义(P=.02)
  • 仅28-mm头亚组总头端穿透率无统计学差异(OxZi 0.045±0.016 mm/y vs CoCr 0.034±0.017 mm/y,P=.066)
  • 稳态磨损率全队列(P=.24)和28-mm头亚组(P=.574)均无显著差异,两种配对磨损率均远低于骨溶解阈值

核心结论

  • 使用RSA评估10年体内头端穿透,OxZi与CoCr关节配对的稳态磨损率无统计学显著差异。
  • 两种轴承组合的磨损率均远低于骨溶解阈值。

研究设计与样本

  • 对两个队列各23例接受全髋关节置换术的患者进行放射立体分析(RSA),分别使用OxZi/XLPE或CoCr/XLPE,随访至少10年。
  • 队列按年龄、性别、体重指数(BMI)和诊断进行匹配。

主要发现

  • 全队列平均总头端穿透率OxZi组0.048±0.021 mm/y vs CoCr组0.035±0.017 mm/y,差异有统计学意义(P=.02)。
  • 仅28-mm头亚组平均总头端穿透率OxZi组0.045±0.016 mm/y vs CoCr组0.034±0.017 mm/y,无统计学显著差异(P=.066)。
  • 全队列平均稳态磨损率OxZi组0.031±0.021 mm/y vs CoCr组0.024±0.019 mm/y,无显著差异(P=.24)。
  • 28-mm头亚组平均稳态磨损率OxZi组0.028±0.019 mm/y vs CoCr组0.024±0.019 mm/y,无显著差异(P=.574)。
  • 除Harris髋关节评分中位数OxZi组较高外,患者报告结局指标(SF12、HHS、WOMAC)无统计学差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Oxidized zirconium (OxZi) femoral heads combine the decreased abrasive properties of ceramics with the toughness of metal alloys to lower wear and increase the durability of total hip arthroplasty. The purpose of this study was to compare the 10-year polyethylene wear rates between OxZi/XLPE and cobalt chrome (CoCr)/XLPE articulations. METHODS: A radiostereometric analysis (RSA) was performed on two cohorts of 23 patients who underwent total hip arthroplasty using either OxZi/XLPE or CoCr/XLPE at a minimum of 10-year follow-up. Cohorts were matched for age, gender, body mass index (BMI), and diagnosis. Polyethylene wear was measured using RSA to determine total and steady-state wear rates for both cohorts. Preoperative and postoperative patient-reported outcome measures (SF12, HHS, and Western Ontario and McMaster Universities Arthritis Index scores) were compared. RESULTS: The mean total head penetration rate was found to be statistically different between the entire cohorts (OxZi 0.048 ± 0.021 mm/y, CoCr 0.035 ± 0.017 mm/y, P = .02) but not when 28-mm heads only (OxZi 0.045 ± 0.016 mm/y, CoCr 0.034 ± 0.017 mm/y, P = .066) were directly compared. The mean steady-state wear rate was not significantly different between the entire cohorts (OxZi 0.031 ± 0.021 mm/y, CoCr 0.024 ± 0.019 mm/y, P = .24) or 28-mm head cohorts (OxZi 0.028 ± 0.019 mm/y, CoCr 0.024 ± 0.019 mm/y, P = .574). Outcome measures showed no statistical difference except for the Harris Hip Score where the OxZi cohort demonstrated higher median scores. CONCLUSION: Using RSA to evaluate the 10-year in-vivo head penetration, there was no statistically significant difference in steady-state wear rates between OxZi and CoCr articulations. Both bearing combinations demonstrated wear rates well below the threshold for osteolysis.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Salipas A, Poole AS, Teeter MG, et al. A Ten-Year Radiostereometric Analysis of Polyethylene Wear Between Oxidized Zirconium and Cobalt Chrome Articulations in Total Hip Arthroplasty. J Arthroplasty. 2022. PMID: 35245689. doi:10.1016/j.arth.2022.02.099

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。