锥形带槽股骨柄的骨干接触长度高度依赖于髓腔形态

The Length of Diaphyseal Contact of Tapered Fluted Stems is Highly Dependent on Canal Morphology.
PubMed 文献 · J Arthroplasty · 2022-03-17

本研究基于33例Dorr B和C型股骨的CT扫描建立三维模型,模拟近端骨缺损后虚拟植入锥形带槽股骨柄,测量柄与骨内膜皮质的接触长度。结果显示接触长度在标本间差异很大(66.5±16.6 mm,范围21-98 mm),且随峡部位置越靠近远端、峡部与损伤区远端边缘距离越大而增加。作者认为在峡部位置更靠近近端的标本中接触有限,可考虑辅助钢板固定和/或使用更长、弯曲的假体。

要点

  • 33例Dorr B和C型股骨三维模型模拟近端骨缺损后植入锥形带槽股骨柄,接触长度为66.5±16.6 mm(范围21-98 mm)。
  • 接触长度随峡部低于小转子的深度增加而增加(r²=0.473,P=.005),随峡部与损伤区远端边缘距离增加而增加(r²=0.508,P=.002)。
  • 逐步回归识别扩髓长度、骨折与峡部距离、峡部直径为接触长度的独立预测因子(r=0.643)。

核心结论

  • 在峡部位置更靠近近端的标本中,锥形带槽股骨柄的骨干接触长度有限。
  • 对于这些病例,作者认为可考虑使用钢板和/或更长、弯曲的假体进行辅助固定。

研究设计与样本

  • 基于33例Dorr B和C型股骨的CT扫描创建三维股骨模型,去除近端120 mm以模拟近端骨缺损。
  • 虚拟植入锥形带槽股骨柄(3度,150 mm)模型前进行髓腔扩髓,测量柄与骨内膜皮质接触长度及其他变量。
  • 使用Spearman相关分析评估变量间关系,logistic回归分析识别接触长度的预测因子(P<.05)。

主要发现

  • 接触长度在标本间差异很大:66.5±16.6 mm,范围21-98 mm。
  • 接触长度随峡部低于小转子的深度增加而增加(范围55-155 mm;r²=0.473,P=.005)。
  • 接触长度随峡部与损伤区远端边缘的距离增加而增加(范围-9至96 mm;r²=0.508,P=.002)。
  • 逐步回归识别扩髓长度、骨折与峡部距离、峡部直径为接触长度的独立预测因子(r=0.643)。

临床怎么用

  • 作者认为,在峡部位置更靠近近端、接触有限的病例中,可考虑使用钢板和/或更长、弯曲的假体进行辅助固定。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Diaphyseal fixation remains the mainstay of revision THA. The stability of diaphyseal fixation can be quantified by the extent of contact between the stem and the endosteal cortex. This is highly affected by the morphology of the proximal femur. The purpose of this study was to examine factors affecting diaphyseal contact in the revision THA and to identify preoperative predictors of adequate fixation. METHODS: Three-dimensional femur models were created from CT scans of 33 Dorr B and C femora. The proximal 120 mm of the femur was omitted to mimic proximal bone deficiency. A tapered fluted stem (3 degrees, 150 mm) model was virtually implanted after reaming of the medullary canal. The contact length between stem and endosteal cortex was measured, in addition to other variables. The relationship between variables was evaluated using Spearman's correlation, and logistic regression analysis was used to identify predictors of the contact length (P < .05). RESULTS: The contact length varied widely between specimens (66.5 ± 16.6 mm, range: 21-98 mm). Contact increased with the depth of the isthmus below the lesser trochanter (range: 55-155 mm; r2 = 0.473, P = .005) and the distance between the isthmus and the distal edge of the damage zone (range: -9 to 96 mm; r2 = 0.508, P = .002). Stepwise regression identified the reaming length, distance between fracture and the isthmus, and isthmus diameter as independent predictors of contact length (r = 0.643). CONCLUSIONS: Contact is limited in specimens where the isthmus is more proximally located. In these cases, supplementary fixation using plating and/or longer, curved prosthesis may be considered.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Frangie R, Wagstaff PD, Ismaily SK, et al. The Length of Diaphyseal Contact of Tapered Fluted Stems is Highly Dependent on Canal Morphology. J Arthroplasty. 2022. PMID: 35307531. doi:10.1016/j.arth.2022.03.041

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。