全髋关节置换术中维生素E高交联聚乙烯与大号股骨头:股骨头侵入率与高交联聚乙烯的比较

Large Femoral Heads in Total Hip Arthroplasty With Vitamin E Highly Cross-Linked Polyethylene: Head Penetration Rates Compared to Highly Cross-Linked Polyethylene.
PubMed 文献 · J Arthroplasty · 2022-02-25

回顾性分析142例初次全髋关节置换术(84例维生素E高交联聚乙烯,58例高交联聚乙烯),采用Martell法测量股骨头侵入。维生素E高交联聚乙烯在中位线性股骨头侵入上数值较低但无统计学显著性,大号股骨头(≥40 mm)在早期磨合期预测较高的股骨头侵入,但1年后无影响。

要点

  • 维生素E高交联聚乙烯组中位线性股骨头侵入在4周至1年磨合期为0.383 mm,高交联聚乙烯组为0.551 mm(P = .650)。
  • 1年至末次随访期间,维生素E高交联聚乙烯组中位线性股骨头侵入为0.131 mm/y,高交联聚乙烯组为0.270 mm/y(P = .636)。
  • 大号股骨头(≥40 mm)在早期磨合期预测较高的股骨头侵入(P ≤ .025),但1年后在多变量回归中无影响。

核心结论

  • 这些发现支持维生素E高交联聚乙烯是磨合期及以后最小化股骨头侵入的最佳聚乙烯承重面的观点。
  • 大号陶瓷股骨头似乎影响初始磨合期的股骨头侵入,但1年后不增加股骨头侵入。
  • 仍需更长期的随访。

研究设计与样本

  • 回顾性审查142例连续初次全髋关节置换术,使用同一植入物系统的陶瓷股骨头(维生素E高交联聚乙烯组n = 84,高交联聚乙烯组n = 58)。
  • 采用Martell法在4周、1年和末次随访X线片上测量股骨头侵入。
  • 证据等级为III级。

主要发现

  • 维生素E高交联聚乙烯组与高交联聚乙烯组在4周至1年磨合期的中位线性股骨头侵入分别为0.383 mm与0.551 mm(P = .650)。
  • 1年至末次随访期间,维生素E高交联聚乙烯组与高交联聚乙烯组的中位线性股骨头侵入分别为0.131 mm/y与0.270 mm/y(P = .636)。
  • 髋臼杯外倾角和前倾角不影响线性或体积性股骨头侵入(P ≥ .204)。
  • 大号股骨头(≥40 mm)在早期磨合期预测较高的股骨头侵入(P ≤ .025),但1年后在多变量回归中无影响。
  • 所有病例均未观察到放射性骨溶解。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文结论指出仍需更长期的随访。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Highly cross-linked polyethylene with vitamin E (VE-HXLPE) has shown superior tribological properties and has been rapidly adopted in total hip arthroplasty. However, the majority of studies compare VE-HXLPE to conventional or moderately cross-linked polyethylene using standard femoral head sizes. This study's purpose was 2-fold: (1) compare radiographic femoral head penetration (FHP) between VE-HXLPE and HXLPE and (2) evaluate FHP in large femoral heads ≥40 mm. METHODS: One hundred forty-two consecutive primary total hip arthroplasties using ceramic femoral heads (n = 84 VE-HXLPE; n = 58 HXLPE) in a single implant system were retrospectively reviewed. FHP was measured radiographically utilizing Martell method at 4-week, 1-year, and latest radiographs. FHP, cup position, and demographic variables were compared between VE-HXLPE and HXLPE liners. RESULTS: Median linear FHP was lower for VE-HXLPE compared to HXLPE during the initial "bedding-in" period between 4-week and 1-year (0.383 vs 0.551 mm, P = .650) and between 1-year and latest follow-up (0.131 vs 0.270 mm/y, P = .636) although without statistical significance. Acetabular cup inclination and anteversion did not influence linear or volumetric FHP (P ≥ .204). Large femoral heads (≥40 mm) were predictive of higher FHP during the early bedding-in period (P ≤ .025) but did not have an effect beyond 1 year in multivariate regression with numbers available. No radiographic osteolysis was observed in any case. CONCLUSION: These findings support others that VE-HXLPE is the optimal polyethylene bearing surface to minimize FHP during the bedding-in period and beyond. Surprisingly, large ceramic femoral heads appear to influence FHP during the initial bedding-in period but do not increase FHP beyond 1 year. Further longer term follow-up remains warranted. LEVEL OF EVIDENCE: III.