文献速递关节外科

现代全膝关节置换术后组件设计与固定方式对应力遮挡的影响

The Impact of Component Design and Fixation on Stress Shielding After Modern Total Knee Arthroplasty.
PubMed 文献 · J Arthroplasty · 2022-02-18

这项前瞻性研究纳入101例全膝关节置换术,比较DePuy Attune与Stryker Triathlon两种设计在骨水泥固定与压配固定下的应力遮挡。2年内胫骨侧各设计间无差异,但股骨侧存在差异;压配胫骨组件应力遮挡相似或更少。结论认为需更长期随访。

要点

  • 2年内胫骨侧应力遮挡在Stryker Triathlon与DePuy Attune间无差异,骨水泥与压配固定均如此。
  • 股骨侧应力遮挡在两种设计间存在差异。
  • Stryker Triathlon与DePuy Attune的压配胫骨组件应力遮挡相似或少于骨水泥固定 counterparts。

核心结论

  • 比较两种胫骨托厚度不同的全膝关节置换植入物,未发现设计间胫骨应力遮挡的显著差异;股骨侧应力遮挡存在设计间差异,提示需更长期随访。

研究设计与样本

  • 经机构审查委员会批准的前瞻性研究,共实施101例全膝关节置换术,分为4个队列:DePuy Attune骨水泥固定与压配固定、Stryker Triathlon骨水泥固定与压配固定。

主要发现

  • 2年期间,胫骨侧应力遮挡在Stryker Triathlon与DePuy Attune之间,骨水泥固定或压配固定均无差异;但股骨侧存在差异。
  • Stryker Triathlon与DePuy Attune的压配胫骨组件在2年期间的应力遮挡与各自骨水泥固定组件相比相似或更少。

关键数值与推荐

  • Attune胫骨托较厚;骨水泥固定的胫骨与股骨组件均为钴铬合金。
  • DePuy Attune压配型采用钴铬烧结珠多孔涂层,Stryker Triathlon采用三维打印高孔隙率钛合金。
  • 所有患者在基线(术后4-6周)及术后1年、2年接受定量双能X线吸收法扫描。
  • 通过比较2年内11个放射学区域的骨密度百分比变化评估应力遮挡。

局限性

  • 原文结论指出股骨侧应力遮挡的设计间差异提示需更长期随访。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Total knee arthroplasty (TKA) tibial baseplate thickness, metallurgy, and implant fixation with cement may influence stress shielding. The purpose of the present study is to compare bone mineral density of 2 cemented and press-fit TKA designs with differing tibial baseplate thicknesses and metallurgy over a 2-year period to assess for changes in stress shielding. METHODS: One-hundred one TKAs were performed in this Institutional Review Board-approved, prospective study. There were 4 cohorts: DePuy Attune cemented and press-fit, and Stryker Triathlon cemented and press-fit. The Attune tibial baseplate was thicker; both cemented tibial and femoral components were cobalt-chromium. The DePuy Attune press-fit had a cobalt-chromium sintered bead porous coating while the Stryker Triathlon was 3-dimensional printed highly porous titanium alloy. All patients had quantitative dual-energy X-ray absorptiometry scans performed at baseline (4-6 weeks postoperatively) and at 1 and 2 years postoperatively. Stress shielding was evaluated by comparing percent change in bone mineral density in 11 radiographic zones over 2 years. RESULTS: Over a 2-year period, there were no differences in stress shielding on the tibial side in either cemented or press-fit between Stryker Triathlon and DePuy Attune; however, there were differences on the femoral side. The press-fit tibial components of the Stryker Triathlon and DePuy Attune had either similar or less stress shielding over a 2-year period compared to their cemented counterparts. CONCLUSION: This study comparing 2 TKA implants with differing tibial tray thickness did not find significant differences in tibial stress shielding between designs. There was a difference in stress shielding on the femoral side between designs, suggesting that longer term follow-up is warranted.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Bendich I, Lawrie CM, Riegler V, et al. The Impact of Component Design and Fixation on Stress Shielding After Modern Total Knee Arthroplasty. J Arthroplasty. 2022. PMID: 35249788. doi:10.1016/j.arth.2022.01.074

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。