全髋关节置换术中的三维功能性撞击:一项生物力学分析

Three-Dimensional Functional Impingement in Total Hip Arthroplasty: A Biomechanical Analysis.
PubMed 文献 · J Arthroplasty · 2022-03-08

该回顾性生物力学研究纳入16例全髋关节置换术患者(年龄61.5±12.1岁,体重指数28.3±4.9 kg/m2),基于术前与术后CT构建三维模型,测试髋臼杯、髋臼内衬、股骨柄和股骨头±5 mm偏移对两种易撞击功能体位下骨-骨撞击前最大活动度的影响。结果发现任何方式增加偏移均增大撞击前活动度,髋臼杯或内衬对旋转中心偏移的改变对假体外撞击影响最大。

要点

  • 在90°屈曲内旋时,内衬±5 mm偏移对活动度影响最大(+9°/-10°),其次为髋臼杯(+8°/-9°)、股骨头(+5°/-7°)和股骨柄(+3°/-5°)。
  • 在10°伸展外旋时,髋臼杯±5 mm偏移对活动度影响最大(+10°/-10°),其次为内衬(+9°/-9°)、股骨头(+7°/-8°)和股骨柄(+4°/-4°)。
  • 通过髋臼杯或内衬改变旋转中心偏移对假体外撞击影响最大,但部分偏移方式间的活动度变化无统计学显著差异。

核心结论

  • 任何方式增加偏移均减少撞击,通过髋臼杯或内衬改变旋转中心偏移对假体外撞击影响最大。

研究设计与样本

  • 回顾性识别16例全髋关节置换术患者(年龄61.5±12.1岁,体重指数28.3±4.9 kg/m2),具有术前和术后计算机断层扫描。
  • 利用术前扫描与术后扫描对齐构建股骨和骨盆三维模型以消除金属伪影,并使用三维扫描植入物模型再现临床植入。
  • 测试髋臼杯、髋臼内衬、股骨柄和股骨头±5 mm偏移,计算90°屈曲内旋和10°伸展外旋时骨-骨撞击前最大活动度。

主要发现

  • 在所有病例中,增加偏移均增加撞击前活动度,反之亦然。
  • 90°屈曲内旋时,内衬±5 mm偏移对活动度影响最大(+9°/-10°),其次为髋臼杯(+8°/-9°)、股骨头(+5°/-7°)和股骨柄(+3°/-5°)。
  • 10°伸展外旋时,髋臼杯±5 mm偏移对活动度影响最大(+10°/-10°),其次为内衬(+9°/-9°)、股骨头(+7°/-8°)和股骨柄(+4°/-4°)。
  • 髋臼杯与内衬偏移在屈曲时活动度变化、内衬与股骨头偏移在伸展时活动度变化、股骨柄与股骨头偏移增加在伸展时活动度变化均无统计学显著差异。

关键数值与推荐

  • 90°屈曲内旋时,内衬±5 mm偏移致活动度变化+9°/-10°。
  • 90°屈曲内旋时,髋臼杯±5 mm偏移致活动度变化+8°/-9°。
  • 10°伸展外旋时,髋臼杯±5 mm偏移致活动度变化+10°/-10°。
  • 10°伸展外旋时,内衬±5 mm偏移致活动度变化+9°/-9°。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Although component offset can affect impingement after total hip arthroplasty, the exact impact is unclear. Evaluation of offset on an anterior-posterior pelvic radiograph is different than evaluation in functional positions of impingement, namely flexion/internal rotation and extension/external rotation. We quantified the effect of acetabular (cup/liner) vs femoral (head/stem) offsets on changes in range of motion to extra-prosthetic impingement in these 2 impingement-prone functional positions. METHODS: We retrospectively identified 16 total hip arthroplasty patients (age 61.5 ± 12.1 years, body mass index 28.3 ± 4.9 kg/m2) with preoperative and postoperative computerized tomography scans. To eliminate metal artifact, femoral and pelvic 3-dimensional models were created using preoperative scans aligned with postoperative scans, and 3-dimensional scanned implant models were used to reproduce clinical implantation. We tested ±5 mm acetabular cup, acetabular liner, femoral stem, and femoral head offsets. Maximum range of motion (ROM) to bone-bone impingement was calculated for internal rotation at 90° flexion and external rotation at 10° extension. RESULTS: In all cases, increased offset increased ROM to impingement, and vice versa. During internal rotation at 90° flexion, ±5 mm liner offset had the greatest impact on ROM (+9°/-10°), followed by cup (+8°/-9°), head (+5°/-7°), and stem (+3°/-5°) offset. During external rotation at 10° extension, ±5 mm cup offset had the greatest impact on ROM (+10°/-10°), followed by liner (+9°/-9°), head (+7°/-8°), and stem (+4°/-4°) offset. However, no statistically significant differences were found in the changes to ROM in flexion obtained through cup and liner offsets, the changes to ROM in extension obtained through liner and head offsets, and the changes to ROM in extension obtained through increasing stem and head offsets. CONCLUSION: Increasing offset by any method reduces impingement. Center-of-rotation offset changes via acetabular cup or liner have the greatest impact on extra-prosthetic impingement.