基于微型惯性导航系统与生物电阻抗分析的新型椎弓根导航器辅助脊柱手术椎弓根螺钉置入:猪模型研究
该研究在12只五指山小型猪中随机分组,比较基于微型惯性导航系统(MINS)与生物电阻抗分析(BIA)的新型椎弓根导航器与徒手技术辅助椎弓根螺钉置入的效果。共置入216枚螺钉,导航组突破15处、徒手组31处,突破检出率分别为93.3%与80.6%,组间差异有统计学意义;导航组螺钉轨迹在轴位和矢状面的角度偏差均小于徒手组,差异有统计学意义。结论认为该便携导航器可能有助于提高或维持精度并减少整体辐射暴露。
要点
- 导航组椎弓根突破检出率为14/15(93.3%),徒手组为25/31(80.6%),组间差异有统计学意义。
- 导航组轴位角度偏差为4.5°±5.5°、矢状位为4.8°±3.3°,徒手组分别为7.0°±5.1°和7.7°±4.7°,差异有统计学意义。
- 基于MINS和BIA的便携导航器可能有助于提高或维持精度并减少整体辐射暴露。
核心结论
- 基于MINS和BIA的新型便携导航器可能有助于提高或维持精度并减少整体辐射暴露。
研究设计与样本
- 研究为猪模型实验,使用12只年龄和体重相近的健康雄性五指山小型猪,随机分为MINS-BIA导航组或徒手(FH)组。
- 将包含MINS的小型化集成框架安装于椎弓根探子的空心手柄内,内芯补充高强度电极用于测量生物电阻抗。
- 根据CT图像上的改良Gertzbein-Robbins分级系统判定椎弓根螺钉置入情况,探针检测阻抗等于软组织基线值定义为皮质骨穿孔。
主要发现
- 12只小型猪共置入216枚螺钉,导航组椎弓根突破15处,徒手组31处。
- 导航组突破检出率为14/15(93.3%),徒手组为25/31(80.6%),组间差异有统计学意义。
- 导航组螺钉轨迹计划与术后测量倾斜角的平均偏差在轴位平面为4.5°±5.5°、矢状面为4.8°±3.3°,徒手组分别为7.0°±5.1°和7.7°±4.7°,差异有统计学意义。
关键数值与推荐
- 椎弓根螺钉位置依据改良Gertzbein-Robbins分级系统在CT图像上判定。
- 探针检测阻抗等于软组织基线值定义为皮质骨穿孔。
临床怎么用
- 原文未据此给出临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
STUDY DESIGN: A porcine model. OBJECTIVE: The study aims to design a novel pedicle navigator based on micro-inertial navigation system (MINS) and bioelectrical impedance analysis (BIA) to assist place pedicle screw placement and validate the utility of the system in enhancing pedicle screw placement. SUMMARY OF BACKGROUND DATA: The incidence of pedicle screw malpositioning in complicated spinal surgery is still high.Procedures such as computed tomography image-guided navigation, and robot-assisted surgery have been used to improve the precision of pedicle screw placement, but it remains an unmet clinical need. METHODS: The miniaturized integrated framework containing MINS was mounted inside the hollow handle of the pedicle finder. The inner core was complemented by a high-intensity electrode for measuring bioelectric impedance. Twelve healthy male Wuzhishan minipigs of similar age and weight were used in this experiment and randomized to the MINS-BIA or freehand (FH) group. Pedicle screw placement was determined according to the modified Gertzbein-Robbins grading system on computed tomography images. An impedance detected by probe equal to the baseline value for soft tissue was defined as cortical bone perforation. RESULTS: A total of 216 screws were placed in 12 minipigs. There were 15 pedicle breaches in the navigator group and 31 in the FH group; the detection rates of these breaches were 14 of 15 (93.3%) and 25 of 31 (80.6%), respectively, with a statistically significant difference between groups. The mean offsets between the planned and postoperatively measured tilt angles of the screw trajectory were 4.5° ± 5.5° in the axial plane and 4.8° ± 3.3° in the sagittal plane with the navigator system and 7.0° ± 5.1° and 7.7° ± 4.7°, respectively, with the FH technique; the differences were statistically significant. CONCLUSION: A novel and portable navigator based on MINS and BIA could be beneficial for improving or maintaining accuracy while reducing overall radiation exposure.
引用
- Lin W, Xie F, Zhao S, et al. Novel Pedicle Navigator Based on Micro Inertial Navigation System (MINS) and Bioelectric Impedance Analysis (BIA) to Facilitate Pedicle Screw Placement in Spine Surgery: Study in a Porcine Model. Spine (Phila Pa 1976). 2022. PMID: 35238856. doi:10.1097/BRS.0000000000004348