皮质骨同种异体移植物高周疲劳寿命与辐射灭菌剂量相关:一项体外研究

The High-cycle Fatigue Life of Cortical Bone Allografts Is Radiation Sterilization Dose-dependent: An In Vitro Study.
PubMed 文献 · Clin Orthop Relat Res · 2022-02-17

该体外研究将来自3名供者的6根人股骨皮质骨标本分配至0、10、17.5、25 kGy四组(每组n=13),在35 MPa生理应力下进行高周疲劳测试。25 kGy组对数转换平均高周疲劳寿命为5.39±0.32,低于其他各组(0 kGy:6.20±0.50;10 kGy:6.35±0.79;17.5 kGy:6.01±0.53;p=0.001)。作者认为皮质骨高周疲劳寿命随γ辐射灭菌剂量增加呈非线性剂量依赖性下降。

要点

  • 25 kGy组对数疲劳寿命5.39±0.32,低于0 kGy组6.20±0.50、10 kGy组6.35±0.79和17.5 kGy组6.01±0.53(p=0.001)。
  • 25 kGy组较0、10、17.5 kGy组更易出现脆性断裂表面模式(p=0.04)。
  • 10 kGy组无序胶原与有序胶原比值由辐照前1.522±0.025降至辐照后1.489±0.024(p=0.01)。

核心结论

  • 皮质骨高周疲劳寿命在临床相关剂量范围(0-25 kGy)内随γ辐射灭菌剂量增加呈非线性、剂量依赖性下降。
  • 高周疲劳寿命的明显下降似出现在17.5 kGy与25 kGy之间,对应平均失效循环数约六倍降低。

研究设计与样本

  • 体外研究,使用来自3名供者(2男分别为63岁和61岁,1女48岁)的6根人股骨皮质骨标本。
  • 标本按供者和取材解剖部位分配至0 kGy(对照)、10 kGy、17.5 kGy和25 kGy四组,每组n=13。
  • 在35 MPa生理应力水平、37°C磷酸盐缓冲盐水浴中,使用定制设计的旋转弯曲疲劳装置进行高周疲劳测试至失效,记录失效循环数。
  • 采用拉曼光谱检测不同γ辐射剂量暴露后拉曼光谱生物标志物的变化。

主要发现

  • 25 kGy组对数转换平均高周疲劳寿命为5.39±0.32,低于0 kGy组6.20±0.50、10 kGy组6.35±0.79和17.5 kGy组6.01±0.53(p=0.001)。
  • 25 kGy组较0 kGy、10 kGy和17.5 kGy组更易出现脆性断裂表面模式而非韧性断裂表面模式(p=0.04)。
  • 无序胶原与有序胶原相对量比值的拉曼生物标志物在10 kGy组由辐照前1.522±0.025降至辐照后1.489±0.024(p=0.01),未观察到其他拉曼生物标志物变化。

临床怎么用

  • 作者指出,工业标准25 kGy辐射灭菌对皮质骨高周疲劳寿命有实质性负面影响。
  • 作者认为,将辐射灭菌剂量降至25 kGy以下可能实质性提高高周疲劳寿命并改善皮质骨同种异体移植物的整体功能寿命。
  • 作者指出,需要在上述临床相关剂量下进一步研究辐射灭菌方法,以在保持无菌的同时帮助保留皮质骨同种异体移植物的高周疲劳寿命。

局限性

  • 作者推测10 kGy时无序胶原拉曼生物标志物下降而高周疲劳寿命未损失可能由胶原骨架非酶交联相对于胶原链断裂增加所致,但未来研究需要确定是否确实如此。
  • 作者指出,需要在上述临床相关剂量下进一步研究辐射灭菌方法,以在保持无菌的同时帮助保留皮质骨同种异体移植物的高周疲劳寿命。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Structural cortical bone allografts are a reasonable treatment option for patients with large cortical bone defects caused by trauma, tumors, or complications of arthroplasty. Although structural cortical bone allografts provide the benefit of an osteoconductive material, they are susceptible to fatigue failure (fracture) and carry a risk of disease transmission. Radiation-sterilization at the recommended dose of 25 kGy decreases the risk of disease transmission. However, previous studies demonstrated that radiation sterilization at this dose can negatively impact the high cycle-fatigue life of cortical bone. Although the effects of higher doses of radiation on cortical bone allografts are well described, the effects of lower doses of radiation on a high-cycle fatigue life of cortical bone are poorly understood. QUESTIONS/PURPOSES: (1) Does the cycle-fatigue life of human cortical allograft bone vary with gamma radiation dose levels of 0 (control), 10 kGy, 17.5 kGy, and 25 kGy? (2) What differences in Raman spectral biomarkers are observed following varying doses of gamma radiation exposure? METHODS: The high-cycle fatigue behavior of human cortical bone specimens was examined at different radiation sterilization doses under physiologic stress levels (35 MPa) and in a 37° C phosphate-buffered saline bath using a custom-designed rotating-bending fatigue device. Six human femora from three donors were obtained for this study (two male, 63 and 61 years old, respectively, and one female, 48 years old). Test specimens were allocated among four treatment groups (0 kGy [control], 10 kGy, 17.5 kGy, and 25 kGy) based on donor and anatomic location of harvest site (both length and cross-sectional quadrant of femoral diaphysis) to ensure equal variation (n = 13 per group). Specimens underwent high-cycle fatigue testing to failure. The number of cycles to failure was recorded. Raman spectroscopy (a noninvasive vibrational spectroscopy used to qualitatively assess bone quality) was used to detect whether any changes in Raman spectral biomarkers occurred after varying doses of gamma radiation exposure. RESULTS: There was a decrease in the log-transformed mean high-cycle fatigue life in specimens irradiated at 25 kGy (5.39 ± 0.32) compared with all other groups (0 kGy: 6.20 ± 0.50; 10k Gy: 6.35 ± 0.79; 17.5 kGy: 6.01 ± 0.53; p = 0.001). Specimens irradiated at 25 kGy were also more likely to exhibit a more brittle fracture surface pattern than specimens with more ductile fracture surface patterns irradiated at 0 kGy, 10 kGy, and 17.5 kGy (p = 0.04). The Raman biomarker for the ratio of the relative amount of disordered collagen to ordered collagen showed a decrease at the 10 kGy radiation level from 1.522 ± 0.025 preirradiation to 1.489 ± 0.024 postirradiation (p = 0.01); no other detectable changes in Raman biomarkers were observed. CONCLUSION: The high-cycle fatigue life of cortical bone undergoes a nonlinear, dose-dependent decrease with an increase in gamma radiation sterilization in a clinically relevant dose range (0-25 kGy). Importantly, a notable drop-off in the high-cycle fatigue life of cortical bone appeared to occur between 17.5 kGy and 25 kGy, correlating to a sixfold decrease in mean cycles to failure. We speculate that the decrease in the Raman biomarker for disordered collagen at 10 kGy with no loss in high-cycle fatigue life may be caused by an increased amount of nonenzymatic crosslinking of the collagen backbone relative to collagen chain-scission (whereas the benefits of crosslinking may be outweighed by excess scission of the collagen backbone at higher radiation doses), but future studies will need to ascertain whether this in fact is the case. CLINICAL RELEVANCE: Radiation sterilization at the industry standard of 25 kGy has a substantial negative impact on the high-cycle fatigue life of cortical bone. Given these findings, it is possible to provide a meaningful increase in the high-cycle fatigue life and improve the overall functional lifetime of cortical bone allografts by lowering the radiation-sterilization dose below 25 kGy. Future work on radiation-sterilization methods at these clinically relevant doses is warranted to aid in preserving the high cycle fatigue life of cortical bone allografts while maintaining sterility.