紫外线空气消毒装置能否降低全膝关节置换术中的颗粒物污染?
研究在40例初次全膝关节置换术中比较使用与不使用紫外线空气消毒装置时手术部位污染率,并在空手术室进行10次额外试验。结果显示使用与不使用紫外线装置之间污染率无统计学差异,而手术室内人员数量和手术室大小是影响污染率的最大变量。
要点
- 使用与不使用紫外线空气消毒装置的全膝关节置换术污染率无统计学差异。
- 空手术室平均污染率为22 ± 1.1 CFUs/m2/hr,使用紫外线装置为21.3 ± 4.6 CFUs/m2/hr,不使用为20.3 ± 4.9 CFUs/m2/hr。
- 手术室内人员数量(6 vs 7人:P = .036,6 vs 8人:P = .004)与污染率差异相关。
核心结论
- 使用或不使用紫外线空气消毒装置的污染率无统计学差异。
- 这40例手术显示影响污染率的最大变量是手术室内人员数量和手术室大小。
研究设计与样本
- 在40例初次全膝关节置换术过程中测量活菌颗粒的沉降率,其中一半手术使用紫外线空气消毒装置。
- 在手术室5个位置用硝酸纤维素膜收集空气颗粒,培养后计数所有微生物菌落并以CFUs/m2/hr报告沉降率。
- 额外在无人员在场的空手术室进行10次试验。
主要发现
- 所有部位平均污染率:空手术室22 ± 1.1 CFUs/m2/hr vs 使用紫外线装置21.3 ± 4.6 CFUs/m2/hr vs 不使用20.3 ± 4.9 CFUs/m2/hr(P = .03,P = .03,P = .964)。
- 无菌区域发现活菌污染物的比例:使用紫外线装置组25% vs 不使用组45%,差异无统计学意义。
- 污染率差异与手术室内人员数量相关:6人 vs 7人 P = .036,6人 vs 8人 P = .004。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Prosthetic joint infections have become the leading cause of joint replacement failure. The primary sources of contamination are skin flora and bacteria from airborne particles. Portable ultraviolet air disinfection units are used in the Operating Room (OR) to prevent contamination from airborne particles; however, their effectiveness is not proven. The purpose of this study was to compare the rate of contamination of sites with and without Ultraviolet (UV) air disinfection units during active surgeries. METHODS: Sedimentation rates of viable particles were measured during 40 primary TKA procedures. Half of the procedures were performed with ultraviolet air disinfection units. Air-borne particles were collected on nitrocellulose membranes at 5 locations within the OR. After incubation, all microbial colonies were counted and the sedimentation rates were reported in CFUs/m2/hr. 10 additional trials were performed in an empty OR with no staff present. RESULTS: The average contamination rate of all sites was 22 ± 1.1 CFUs/m2/hr in the empty OR vs. 21.3 ± 4.6 CFUs/m2/hr with UV units and 20.3 ± 4.9 CFUs/m2/hr without (P = .03, P = .03, P = .964). Viable contaminates were found in the sterile field in 25% of UV cases vs 45% non-UV. These differences were not statistically significant. There were differences found however, according to the number of staff in the room (6 vs 7 staff: P = .036, 6 vs 8 staff: P = .004). CONCLUSION: There was no statistical difference in contamination rate with the usage or non-usage of UV units. These 40 cases shows that the largest variables affecting the contamination rate were the number of staff present and size of the OR.
引用
- Jones HL, Braly HL, Parsley BS, et al. Do Ultraviolet Air Disinfection Units Reduce Contamination by Particulates in Total Knee Replacement Procedures? J Arthroplasty. 2022. PMID: 35219817. doi:10.1016/j.arth.2022.02.070
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。