退伍军人人群中羟基磷灰石涂层与非羟基磷灰石涂层全髋关节置换植入物的生存率比较

Survival of Hydroxyapatite-Coated Versus Non-hydroxyapatite-Coated Total Hip Arthroplasty Implants in a Veteran Population.
PubMed 文献 · J Arthroplasty · 2022-02-25

该研究基于2000至2017年退伍军人健康管理局医院接受非骨水泥初次全髋关节置换术的患者,比较同类型羟基磷灰石涂层与非涂层假体的翻修风险。共分析262个涂层髋臼杯、4580个非涂层杯、4767个涂层股骨柄和9397个非涂层柄,平均随访分别为9.01年和7.13年。结果显示,一种髋臼杯的涂层版本生存率显著更低且翻修风险升高,其余假体对的生存率和翻修风险未受涂层影响。结论认为羟基磷灰石涂层未改善植入物生存率,不支持其普遍使用。

要点

  • 退伍军人人群中,一种羟基磷灰石涂层髋臼杯的翻修风险显著高于同型非涂层杯。
  • 其余四种股骨柄配对及另一种髋臼杯配对中,涂层与否对生存率和翻修风险无显著影响。
  • 总体而言,羟基磷灰石涂层未改善全髋关节置换植入物生存率。

核心结论

  • 在退伍军人人群中,羟基磷灰石涂层未改善全髋关节置换植入物的生存率。
  • 尽管涂层假体通常更昂贵,这些结果不支持其普遍使用。

研究设计与样本

  • 回顾性队列研究,纳入2000至2017年在任何退伍军人健康管理局医院接受非骨水泥初次全髋关节置换术且植入物有涂层或非涂层型号可供选择的患者。
  • 终点为因任何原因取出目标部件。
  • 对每个股骨柄和髋臼杯,使用未调整和调整(年龄和体重指数)的Cox比例风险模型估计同类型涂层与非涂层植入物之间取出风险的HR。

主要发现

  • 共分析262个羟基磷灰石涂层髋臼杯、4580个非涂层杯、4767个涂层股骨柄和9397个非涂层柄。
  • 髋臼杯平均随访9.01年(总计43,627植入物年),股骨柄平均随访7.13年(总计101,004植入物年)。
  • 所审查的两种髋臼杯中的一种,涂层版本生存率显著更低且取出HR升高。
  • 五种股骨柄配对及另一种髋臼杯中,植入物生存率和取出HR未受羟基磷灰石涂层影响。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Hydroxyapatite (HA) coatings were introduced to improve uncemented implant osteointegration and to prevent loosening and osteolysis. However, data regarding these implants have been inconsistent. We studied the effect of HA coating of femoral stems and acetabular cups on component revision after primary total hip arthroplasty (THA) in the veteran population. METHODS: We identified patients who underwent uncemented primary THA at any Veterans Health Administration (VHA) hospital from 2000 to 2017 and who had implants that were available as either HA-coated or non-HA-coated models. The endpoint was removal of the component of interest for any reason. For each individual stem and shell, unadjusted and adjusted (for age and body mass index) Cox proportional hazards models were used to estimate hazard ratios for the risk of explantation between HA-coated and non-HA-coated implants of the same type. RESULTS: A total of 262 HA-coated cups, 4580 non-HA-coated cups, 4767 HA-coated stems, and 9397 non-HA-coated stems were available for analysis. The mean follow-up time was 9.01 years (43,627 total implant-years) for cups and 7.13 years (101,004 total implant-years) for stems. One of the two shells reviewed had significantly lower survivorship and an elevated hazard ratio for explantation with HA coating. Among the five implant pairs of stems, and the other shell, implant survival and hazard ratios for explantation were not affected by HA coating. CONCLUSION: HA coating did not improve THA implant survival in our veteran population. Although HA-coated versions of hip implants tend to be more costly than their noncoated counterparts, these results do not support their general use.