文献速递关节外科

全髋关节置换术中的锥度材料损失:是否受关节摩擦力影响

Taper Material Loss in Total Hip Replacements: Is It Affected by Joint Friction?
PubMed 文献 · J Bone Joint Surg Am · 2022-02-15

研究对85例翻修取出的金属对金属大头髋假体进行计量学与生物力学测试,量化头锥度材料损失和关节面磨损,并测量模拟步态下的关节摩擦力矩。结果显示锥度材料损失与摩擦力矩无统计学相关性,但与在体时间显著相关。

要点

  • 头锥度材料损失为0.04至25.57 mm³,关节面联合磨损为0.80至351.75 mm³。
  • 锥度材料损失与摩擦力矩无统计学相关性(R=-0.20至0.11,p=0.07至0.81)。
  • 锥度材料损失随时间显著增加(R=0.34,p<0.01)。

核心结论

  • 尽管对锥部磨损存在严重担忧,头锥度磨损量总体低于关节面磨损量。
  • 锥度磨损与摩擦力矩之间无统计学相关性,结果提示金属对金属假体的高摩擦力并未导致头锥度材料损失增加。

研究设计与样本

  • 研究纳入85例在翻修手术中取出的金属对金属大头髋臼杯和股骨头,在体服务时间为10至96个月。
  • 首先进行计量学量化头锥度材料损失和假体关节面磨损,然后在生物力学模型中对每对取出的头杯组合在10个周期的模拟生理步态中测量关节摩擦力矩。
  • 评估锥度材料损失与摩擦力矩、关节面磨损、头直径、头杯间隙及在体时间的相关性。

主要发现

  • 峰值合力摩擦力矩范围为9.1至26.3 Nm,平均为17.3 ± 2.7 Nm。
  • 在体服役期间微动和腐蚀损伤导致头锥度材料损失范围为0.04至25.57 mm³,而该队列头杯关节面联合磨损范围为0.80至351.75 mm³。
  • 锥度材料损失与更高摩擦力矩无相关性(R=-0.20至0.11,p=0.07至0.81)。
  • 轴向旋转产生的更高摩擦力矩与更高的头杯磨损相关(R=0.33,p<0.01)。
  • 锥度材料损失与头直径的相关性弱且未达统计学显著(R=0.20,p=0.07),与名义头杯间隙无相关性(R=0.06,p=0.6),随时间显著增加(R=0.34,p<0.01)。

临床怎么用

  • 原文未据此给出推荐。
  • 作者指出,多种因素影响锥度磨损:假体设计、材料、尺寸、表面光洁度以及患者体重和活动水平。
  • 作者认为,在本队列中,尽管锥度材料损失部位的力矩增加,金属对金属全髋关节置换的高摩擦力可能并未导致锥度界面材料损失体积增加。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Metal debris and corrosion products generated from the taper junctions of modular joint replacements have been recognized as contributors to failure. Therefore, understanding the factors associated with increased taper wear and corrosion is fundamental to improving implant performance. METHODS: A cohort of 85 large-diameter metal-on-metal heads and cups retrieved at revision surgery, after 10 to 96 months of service, was evaluated. First, metrology was conducted to quantify head taper material loss and implant articular surface wear. Then, joint frictional moments for each retrieved head-and-cup pair were measured during 10 cycles of simulated physiological gait in a biomechanical model. Taper material loss was evaluated for correlations with frictional moments, articular wear, head diameter, head-cup clearance, and time in vivo. RESULTS: Peak resultant frictional moments ranged from 9.1 to 26.3 Nm, averaging 17.3 ± 2.7 Nm. Fretting and corrosion damage during in vivo service resulted in material loss from the head tapers ranging between 0.04 and 25.57 mm3, compared with combined head and cup articular wear of 0.80 to 351.75 mm3 in this cohort. Taper material loss was not correlated with higher frictional moments (R = -0.20 to 0.11, p = 0.07 to 0.81). Higher frictional moments from axial rotation were correlated with higher head and cup wear (R = 0.33, p < 0.01). The correlation between taper material loss and head diameter was weak and did not reach statistical significance (R = 0.20, p = 0.07). Taper material loss was not correlated with nominal head-cup clearance (R = 0.06, p = 0.6). Finally, taper material loss increased significantly over time (R = 0.34, p < 0.01). CONCLUSIONS: Despite serious concerns regarding trunnionosis, volumes of head taper wear were generally lower than those of articular surface wear. There was no statistical correlation between taper wear and frictional moments. Therefore, the results suggest that high friction in metal-on-metal implants does not contribute to higher material loss at the head taper, despite high bending moments. CLINICAL RELEVANCE: The amount of metal debris and corrosion products from taper junctions of the joint arthroplasties, widely recognized as an insidious cause of failure, was not correlated with joint frictional moments. Multiple factors affect taper wear: implant design, material, size, surface finish, and patient weight and activity level. However, in the present cohort, high friction of metal-on-metal total hip replacements likely did not contribute to increased volume of material loss at taper interfaces, despite increased moments at the locations of taper material loss.

来源:J Bone Joint Surg Am,查看原文。版权归原作者所有。