巨噬细胞对钴铬钼合金的体外细胞诱导腐蚀研究
研究将鼠巨噬细胞培养于有无单极电灼损伤的钴铬钼(CoCrMo)合金盘上,每12小时换液、每4天收集上清,30天后检测金属浓度。结果显示未损伤盘上活化巨噬细胞使上清钴浓度更高,未损伤盘上清钴和钼浓度高于电灼损伤盘,细胞内钴和钼浓度在活化细胞组显著高于对照组。扫描电镜显示无电灼损伤的巨噬细胞暴露表面出现微小点蚀。
要点
- 巨噬细胞可在CoCrMo合金表面重现炎症细胞诱导腐蚀(ICIC)点蚀。
- 电灼损伤未加速巨噬细胞诱导的腐蚀过程。
- 未损伤盘上清钴浓度高于电灼损伤盘,细胞内钴和钼浓度活化细胞组显著高于对照组。
核心结论
- 巨噬细胞可在体外重现CoCrMo合金表面的ICIC点蚀,且电灼损伤未增加该过程。
- 这些发现的临床意义需进一步研究,以确定其是否能解释少数全膝关节置换术不良结局。
研究设计与样本
- 体外研究:将鼠巨噬细胞培养于有无单极电灼损伤的CoCrMo合金盘上。
- 每12小时更换培养基,每4天收集上清,30天后移除、计数并消化细胞。
- 采用电感耦合等离子体光谱法比较上清和细胞内金属浓度。
主要发现
- 未损伤盘上活化巨噬细胞的上清钴浓度更高。
- 未损伤盘上清钴和钼浓度高于电灼腐蚀损伤盘。
- 活化细胞在CoCrMo盘上的细胞内钴和钼浓度显著高于对照组,与电灼损伤盘组相比无差异。
- 扫描电镜显示无电灼损伤的巨噬细胞暴露表面出现微小点蚀。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Patients have received cobalt-chromium-molybdenum (CoCrMo) implants for their joint replacement for decades. There have been reports of inflammatory cell-induced corrosion (ICIC) of these implants from retrieval studies. The goal of this study is to see if we could recreate ICIC in vitro and whether electrocautery damage to alloy surfaces may hasten this process. METHODS: Murine macrophages were cultured on CoCr disks with and without damage from a monopolar electrocautery. Culture medium was replaced every 12 hours and supernatant was collected every 4 days. After 30 days, cells were removed, counted, and digested. The metal concentrations in the supernatant and within cells were assessed using inductively coupled plasma spectrometry for comparison. RESULTS: The Co supernatant concentration was higher in the undamaged disks with activated macrophages. Higher concentrations of Co and Mo were found in the supernatant of the undamaged disks vs the electrocautery (EC) corrosion damaged disks. There was a significantly higher intracellular Co and Mo concentration with activated cells on CoCrMo disks vs the control group and no difference compared to EC damaged disk group. Scanning electron microscopy displayed microscopic pitting on the surfaces exposed to macrophages without EC damage. CONCLUSION: We found that macrophages could reproduce findings of ICIC pits on the surface of CoCrMo alloy and that the addition of EC damage to the surface did not increase the process. The clinical significance of these findings should be further investigated to determine if this could explain a small number of poor total knee arthroplasty reported outcomes.
引用
- Miller KC, Holloway MB, Morrow BR, et al. In-Vitro Cell-Induced Corrosion by Macrophages on Cobalt-Chromium-Molybdenum Alloy. J Arthroplasty. 2022. PMID: 35219574. doi:10.1016/j.arth.2022.01.062
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。