标准自动视野计中的假阳性反应分析

False Positive Responses in Standard Automated Perimetry.
PubMed 文献 · Am J Ophthalmol · 2021-07-18

研究发现假阳性率与视野结果仅呈弱相关,假阳性率每增加10个百分点,MD仅增加0.3至0.6 dB,VFI变化更小;但在严重青光眼中假阳性效应更大,MD增加1.1至2.0 dB。GH与视野结果的相关性远强于假阳性率。

要点

  • 假阳性率与视野结果仅弱相关,假阳性率每增10个百分点MD仅增0.3至0.6 dB。
  • 严重青光眼中假阳性效应更大,假阳性率每增10个百分点MD增1.1至2.0 dB。
  • GH与视野结果的相关性远强于假阳性率,或更适合识别不可靠结果。

核心结论

  • 在三种SITA阈值算法中,假阳性率与视野检测结果仅呈弱相关,严重青光眼除外。
  • 当前关于可接受假阳性范围的建议可能需要修订。
  • GH或其他分析可能比假阳性率更适合识别频繁按压反应按钮但未感知刺激的不可靠结果。

研究设计与样本

  • 每位患者在两次访视中分别接受SITA Standard、SITA Fast和SITA Faster检测。

主要发现

  • 假阳性值增加与三种算法MD值升高相关,但效应较小,假阳性率每增10个百分点MD增0.3至0.6 dB。
  • VFI变化更小,假阳性率每增10个百分点VFI变化0.6%至1.4%。
  • 访视间差异仅小部分由假阳性解释,r²值为0.00至0.11。
  • 严重青光眼中假阳性效应更大,假阳性率每增10个百分点MD增1.1至2.0 dB,r²值0.04至0.33。
  • 显著压低的总偏差点数仅轻微受影响,模式偏差概率图一般不受影响。

关键数值与推荐

  • 假阳性率每增加10个百分点,MD增加0.3至0.6 dB,VFI变化0.6%至1.4%。
  • 严重青光眼中假阳性率每增加10个百分点,MD增加1.1至2.0 dB。
  • 假阳性解释访视间差异的r²值为0.00至0.11,严重青光眼中为0.04至0.33。

临床怎么用

  • 解读视野结果时不应仅依赖假阳性率判断可靠性,尤其需结合GH等指标。
  • 在严重青光眼中应更关注假阳性率对MD的影响。
  • 当前可接受假阳性范围可能需重新评估。

局限性

  • 仅纳入确诊或疑似青光眼患者,结果可能不适用于其他人群。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To analyze the relationship between rates of false positive (FP) responses and standard automated perimetry results. DESIGN: Prospective multicenter cross-sectional study. METHODS: One hundred twenty-six patients with manifest or suspect glaucoma were tested with Swedish Interactive Thresholding Algorithm (SITA) Standard, SITA Fast, and SITA Faster at each of 2 visits. We calculated intervisit differences in mean deviation (MD), visual field index (VFI), and number of statistically significant test points as a function of FP rates and also as a function of general height (GH). RESULTS: Increasing FP values were associated with higher MD values for all 3 algorithms, but the effects were small, 0.3 dB to 0.6 dB, for an increase of 10 percentage points of FP rate, and for VFI even smaller (0.6%-1.4%). Only small parts of intervisit differences were explained by FP (r2 values 0.00-0.11). The effects of FP were larger in severe glaucoma, with MD increases of 1.1 dB to 2.0 dB per 10 percentage points of FP, and r2 values ranging from 0.04 to 0.33. The numbers of significantly depressed total deviation points were affected only slightly, and pattern deviation probability maps were generally unaffected. GH was much more strongly related to perimetric outcomes than FP. CONCLUSIONS: Across 3 different standard automated perimetry thresholding algorithms, FP rates showed only weak associations with visual field test results, except in severe glaucoma. Current recommendations regarding acceptable FP ranges may require revision. GH or other analyses may be better suited than FP rates for identifying unreliable results in patients who frequently press the response button without having perceived stimuli.