深度学习全容积OCT自动分割在视网膜疾病中的验证与临床适用性
定性评估显示,自动化全容积分割在73%的扫描中排名优于或等同于至少1名专家分级者,78%的模型分割获得中性或正面评价。模型提供的视网膜疾病定量测量在临床上具有实用价值。
要点
- 自动化全容积分割在73%扫描中优于或等同于至少1名专家。
- 78%模型分割获中性或正面评价,专家分级为89%。
- 模型在糖尿病黄斑水肿扫描中评价更佳(86%-92%)。
核心结论
- 基于深度学习的全容积分割工具可提供临床上有用的视网膜疾病测量,否则难以获得。
主要发现
- 自动化全容积分割在127张扫描(73%;95% CI,66%-79%)中排名优于或等同于至少1名专家分级者。
- 135个模型分割(78%;95% CI,71%-84%)和309个专家分级(89%;95% CI,86%-92%)获得中性或正面评价。
- 模型在糖尿病黄斑水肿扫描中获中性或正面评价的比例为86%至92%,在年龄相关性黄斑变性扫描中为53%至87%。
- 组内相关系数范围为0.33(95% CI,0.08-0.96)至0.96(95% CI,0.90-0.99)。
- Dice相似系数范围为0.43(95% CI,0.29-0.66)至0.78(95% CI,0.57-0.85)。
关键数值与推荐
- 两名专家分级者对视网膜内液、视网膜下液、视网膜下高反射物质和色素上皮脱离进行像素级分割,每张扫描耗时长达50小时。
临床怎么用
- 该工具可提供临床上有用的视网膜疾病定量测量,定性评估对揭示临床适用性很重要。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
IMPORTANCE: Quantitative volumetric measures of retinal disease in optical coherence tomography (OCT) scans are infeasible to perform owing to the time required for manual grading. Expert-level deep learning systems for automatic OCT segmentation have recently been developed. However, the potential clinical applicability of these systems is largely unknown. OBJECTIVE: To evaluate a deep learning model for whole-volume segmentation of 4 clinically important pathological features and assess clinical applicability. DESIGN, SETTING, PARTICIPANTS: This diagnostic study used OCT data from 173 patients with a total of 15 558 B-scans, treated at Moorfields Eye Hospital. The data set included 2 common OCT devices and 2 macular conditions: wet age-related macular degeneration (107 scans) and diabetic macular edema (66 scans), covering the full range of severity, and from 3 points during treatment. Two expert graders performed pixel-level segmentations of intraretinal fluid, subretinal fluid, subretinal hyperreflective material, and pigment epithelial detachment, including all B-scans in each OCT volume, taking as long as 50 hours per scan. Quantitative evaluation of whole-volume model segmentations was performed. Qualitative evaluation of clinical applicability by 3 retinal experts was also conducted. Data were collected from June 1, 2012, to January 31, 2017, for set 1 and from January 1 to December 31, 2017, for set 2; graded between November 2018 and January 2020; and analyzed from February 2020 to November 2020. MAIN OUTCOMES AND MEASURES: Rating and stack ranking for clinical applicability by retinal specialists, model-grader agreement for voxelwise segmentations, and total volume evaluated using Dice similarity coefficients, Bland-Altman plots, and intraclass correlation coefficients. RESULTS: Among the 173 patients included in the analysis (92 [53%] women), qualitative assessment found that automated whole-volume segmentation ranked better than or comparable to at least 1 expert grader in 127 scans (73%; 95% CI, 66%-79%). A neutral or positive rating was given to 135 model segmentations (78%; 95% CI, 71%-84%) and 309 expert gradings (2 per scan) (89%; 95% CI, 86%-92%). The model was rated neutrally or positively in 86% to 92% of diabetic macular edema scans and 53% to 87% of age-related macular degeneration scans. Intraclass correlations ranged from 0.33 (95% CI, 0.08-0.96) to 0.96 (95% CI, 0.90-0.99). Dice similarity coefficients ranged from 0.43 (95% CI, 0.29-0.66) to 0.78 (95% CI, 0.57-0.85). CONCLUSIONS AND RELEVANCE: This deep learning-based segmentation tool provided clinically useful measures of retinal disease that would otherwise be infeasible to obtain. Qualitative evaluation was additionally important to reveal clinical applicability for both care management and research.
引用
- Wilson M, Chopra R, Wilson MZ, et al. Validation and Clinical Applicability of Whole-Volume Automated Segmentation of Optical Coherence Tomography in Retinal Disease Using Deep Learning. JAMA Ophthalmol. 2021. PMID: 34236406. doi:10.1001/jamaophthalmol.2021.2273
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。