两种频域OCT检测青光眼黄斑GCIPL纵向变化的比较
本研究在68只青光眼眼中比较了Spectralis与Cirrus两种OCT设备检测黄斑GCIPL厚度变化率及恶化/改善比例。结果显示Spectralis检测到的GCIPL变薄速率更快,且显著负向变化比例更高,提示两种OCT设备在检测黄斑结构进展时不可互换。
要点
- Spectralis OCT检测到的GCIPL变薄速率较Cirrus更快,在鼻上及颞上扇区差异显著。
- Spectralis OCT在全球及颞下/颞上扇区显示更高比例的显著负向变化率。
- 两种OCT设备在检测黄斑结构进展时不可互换。
核心结论
- Spectralis OCT的纵向容积扫描比Cirrus OCT更频繁地检测到黄斑GCIPL变化。
- 两种OCT设备在检测黄斑结构进展方面不可互换。
研究设计与样本
- 纵向队列研究,纳入68只随访≥2年且OCT图像≥4幅的青光眼眼。
- 从两种OCT设备导出黄斑容积扫描,进行配准和分割,提取中央4.8×4.0-mm区域整体和扇区GCIPL数据。
- 使用线性回归估计GCIPL变化率,置换分析控制特异性,以2.5百分位截断点定义「真实」恶化。
主要发现
- 平均(标准差)24-2视野平均偏差为-9.4(6.1)dB,中位(四分位距)随访时间为3.8(3.3-4.2)年,OCT图像数为6(5-8)幅。
- Spectralis OCT的GCIPL变薄速率较Cirrus更负,在鼻上(P = .03)和颞上(P = .04)扇区差异显著。
- Spectralis OCT在全球及颞下/颞上扇区观察到更高比例的显著负向变化率(P < .04)。
- 置换分析证实Spectralis OCT在全球和扇区负向变化率比例更高(P < .001)。
关键数值与推荐
- 使用2.5百分位截断点定义「真实」恶化。
- 主要结局包括两种OCT设备间整体/扇区GCIPL变化率差异及负向 vs 正向变化率比例(P < .05)。
临床怎么用
- 临床监测青光眼黄斑结构进展时,不应互换使用Spectralis和Cirrus OCT设备。
- 使用Spectralis OCT可能更频繁地检测到GCIPL变化,需结合设备特性解读随访数据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: We compared rates of change of macular ganglion cell/inner plexiform (GCIPL) thickness and proportion of worsening and improving rates from 2 optical coherence tomography (OCT) devices in a cohort of eyes with glaucoma. DESIGN: Longitudinal cohort study. METHODS: In a tertiary glaucoma clinic we evaluated 68 glaucoma eyes with ≥2 years of follow-up and ≥4 OCT images. Macular volume scans from 2 OCT devices were exported, coregistered, and segmented. Global and sectoral GCIPL data from the central 4.8 × 4.0-mm region were extracted. GCIPL rates of change were estimated with linear regression. Permutation analyses were used to control specificity with the 2.5 percentile cutoff point used to define "true" worsening. Main outcome measures included differences in global/sectoral GCIPL rates of change between 2 OCT devices and the proportion of negative vs positive rates of change (P < .05). RESULTS: Average (standard deviation) 24-2 visual field mean deviation, median (interquartile range) follow-up time, and number of OCT images were -9.4 (6.1) dB, 3.8 (3.3-4.2) years, and 6 (5-8), respectively. GCIPL rates of thinning from Spectralis OCT were faster (more negative) compared with Cirrus OCT; differences were significant in superonasal (P = .03) and superotemporal (P = .04) sectors. A higher proportion of significant negative rates was observed with Spectralis OCT both globally and in inferotemporal/superotemporal sectors (P < .04). Permutation analyses confirmed the higher proportion of global and sectoral negative rates of change with Spectralis OCT (P < .001). CONCLUSIONS: Changes in macular GCIPL were detected more frequently on Spectralis' longitudinal volume scans than those of Cirrus OCT. OCT devices are not interchangeable with regard to detection of macular structural progression.