超广角眼底照相检测青光眼视网膜神经纤维层缺损的新应用
本研究评估了非散瞳超广角眼底照相(UWP)及其数字转换绿色分离图像(G-UWP)检测视网膜神经纤维层缺损(RNFLD)的效能,并与传统无赤光RNFL照相(RFP)比较。纳入196只眼(84正常对照、25疑似青光眼、87青光眼),结果显示G-UWP敏感性与RFP相当但特异性较低,UWP敏感性低于G-UWP和RFP。结论认为非散瞳UWP可用于检测RNFLD,G-UWP可作为临床筛查青光眼的便捷工具。
要点
- G-UWP检测RNFLD的敏感性(94.6%)与RFP(92.9%)相当,均高于UWP(82.1%)。
- G-UWP和UWP的特异性(78.6%、75.0%)均低于RFP(98.8%)。
- 非散瞳G-UWP可作为临床筛查青光眼的便捷诊断工具。
核心结论
- 非散瞳超广角眼底照相(UWP)可用于检测视网膜神经纤维层缺损(RNFLD)。
- 非散瞳G-UWP检测RNFLD的敏感性与传统RFP相当,但特异性较低。
- 非散瞳G-UWP可作为临床筛查青光眼的便捷有用工具。
研究设计与样本
- 研究连续纳入2019年9月至2020年4月符合标准的青光眼、疑似青光眼及正常对照眼。
- 共获得196名参与者(84正常对照、25疑似青光眼、87青光眼)的三类图像集。
- 评估传统RFP、非散瞳UWP及G-UWP图像检测RNFLD的敏感性和特异性。
主要发现
- G-UWP检测RNFLD的敏感性为94.6%(95%CI 88.7-98.0),RFP为92.9%(95%CI 86.4-96.9),均高于UWP的82.1%(95%CI 73.8-88.7;p<0.05)。
- G-UWP与RFP的敏感性相当。
- G-UWP特异性为78.6%(95%CI 68.3-86.8),UWP为75.0%(95%CI 64.4-83.8),两者相当,但均低于RFP的98.8%(95%CI 93.5-100.0;p<0.05)。
临床怎么用
- 在临床环境中,非散瞳G-UWP可作为筛查青光眼的便捷工具,用于检测RNFLD。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND/AIMS: Evaluation of the retinal nerve fibre layer (RNFL) is important for identifying glaucomatous damage. Ultrawide-field fundus photography (UWP) imaging is increasingly used in the ophthalmological field; however, it is unknown whether it can be used for detecting RNFL defects (RNFLDs). We investigated whether RNFLD can be detected with UWP images and compared the clinical effectiveness of three types of images for detecting RNFLD: conventional red-free RNFL photography (RFP), non-mydriatic UWP and digitally converted green separation of non-mydriatic UWP (G-UWP). METHODS: Eyes with glaucoma or glaucoma suspect and normal control eyes meeting the eligibility criteria were consecutively enrolled from September 2019 to April 2020. Their conventional RFP, non-mydriatic UWP and G-UWP images were assessed for detecting RNFLD to evaluate the sensitivity and specificity for detecting RNFLD. RESULTS: Three image sets of 196 participants (84 normal control, 25 glaucoma suspect and 87 glaucoma) were obtained. The sensitivity of G-UWP (94.6%; 95% CI 88.7 to 98.0) and RFP (92.9%; 95% CI 86.4 to 96.9) was higher than that of UWP (82.1%; 95% CI 73.8 to 88.7; p<0.05). The sensitivities of G-UWP and RFP are comparable. The specificity of G-UWP (78.6%; 95% CI 68.3 to 86.8) and UWP (75.0%; 95% CI 64.4 to 83.8) was comparable, but both were lower than that of RFP (98.8%; 95% CI 93.5 to 100.0; p<0.05). CONCLUSION: Non-mydriatic UWP images can be used to detect RNFLD. Non-mydriatic G-UWP showed comparable sensitivity but lower specificity to conventional RFP. Non-mydriatic G-UWP could be used as a convenient and useful diagnostic tool for screening glaucoma in clinical settings.