植入式微传感器连续眼压遥测提示注视方向与眼睑闭合影响眼压的初步研究
本研究利用植入式遥测眼压传感器,在11例原发性开角型青光眼患者中探索注视方向和眼睑闭合对眼压的影响。结果显示眼压随注视方向可重复变化,上注视时升高,鼻下注视和眼睑闭合时降低,但存在患者间变异。
要点
- 上注视25°时眼压平均升高4.4±4.9 mmHg,鼻下注视25°时降低1.6±0.8 mmHg。
- 注视方向与偏心度存在显著交互作用(P=0.003),鼻下注视各偏心度均显著降低眼压。
- 眼睑闭合时眼压降低,且显著低于下注视25°时的眼压变化(P=0.014)。
核心结论
- 眼压随注视方向可重复变化,但存在患者间变异;上注视一般升高眼压,鼻下注视和眼睑闭合降低眼压。
研究设计与样本
- 11例曾植入遥测眼压传感器的原发性开角型青光眼患者,注视8个等间距目标,每个目标在10°、20°和25°三个偏心度;其中9例还进行眼睑闭合测试。
主要发现
- 最大眼压升高见于上注视25°(均值±标准差:4.4±4.9 mmHg),最大降低见于鼻下注视25°(-1.6±0.8 mmHg)。
- 注视方向与偏心度存在显著交互作用(P=0.003);鼻下注视各偏心度均显著降低眼压(10°:-0.7±0.2,P=0.007;20°:-1.1±0.2,P=0.006;25°:-1.6±0.2,P=0.006)。
- 11眼中8眼在上注视与鼻下注视25°时眼压差异显著。
- 眼睑闭合时眼压降低,且显著低于下注视25°时的眼压变化(均值±标准误:-2.1±0.3 mmHg vs. -0.7±0.2 mmHg,P=0.014)。
临床怎么用
- 临床眼压测量时应考虑注视方向和眼睑闭合的影响,以避免测量偏差。
局限性
- 样本量小(11例),为初步研究,患者间变异较大,需进一步研究眼压动态和变异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To explore the effect of gaze direction and eyelid closure on intraocular pressure (IOP). METHODS: Eleven patients with primary open-angle glaucoma previously implanted with a telemetric IOP sensor were instructed to view eight equally-spaced fixation targets each at three eccentricities (10°, 20°, and 25°). Nine patients also performed eyelid closure. IOP was recorded via an external antenna placed around the study eye. Differences of mean IOP between consecutive gaze positions were calculated. Furthermore, the effect of eyelid closure on gaze-dependent IOP was assessed. RESULTS: The maximum IOP increase was observed at 25° superior gaze (mean ± SD: 4.4 ± 4.9 mm Hg) and maximum decrease at 25° inferonasal gaze (-1.6 ± 0.8 mm Hg). There was a significant interaction between gaze direction and eccentricity (P = 0.003). Post-hoc tests confirmed significant decreases inferonasally for all eccentricities (mean ± SEM: 10°: -0.7 ± 0.2, P = 0.007; 20°: -1.1 ± 0.2, P = 0.006; and 25°: -1.6 ± 0.2, P = 0.006). Eight of 11 eyes showed significant IOP differences between superior and inferonasal gaze at 25°. IOP decreased during eyelid closure, which was significantly lower than downgaze at 25° (mean ± SEM: -2.1 ± 0.3 mm Hg vs. -0.7 ± 0.2 mm Hg, P = 0.014). CONCLUSIONS: Our data suggest that IOP varies reproducibly with gaze direction, albeit with patient variability. IOP generally increased in upgaze but decreased in inferonasal gaze and on eyelid closure. Future studies should investigate the patient variability and IOP dynamics.
引用
- van den Bosch JJON, Pennisi V, Invernizzi A, et al. Implanted Microsensor Continuous IOP Telemetry Suggests Gaze and Eyelid Closure Effects on IOP-A Preliminary Study. Invest Ophthalmol Vis Sci. 2021. PMID: 33956052. doi:10.1167/iovs.62.6.8
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。