文献速递青光眼

从SITA标准转换为SITA快速算法对视野检查表现的影响

The Effect of Transitioning from SITA Standard to SITA Faster on Visual Field Performance.
PubMed 文献 · Ophthalmology · 2021-03-30

这项回顾性纵向研究纳入421例(766眼)疑似或确诊青光眼患者,比较SITA标准与SITA快速算法序贯检查的视野平均偏差变化。轻症者两种算法无显著差异,中度和晚期青光眼使用SITA快速算法后MD值更高,可能掩盖疾病进展。

要点

  • 轻度或疑似青光眼:SITA快速与标准算法间ΔMD无显著差异。
  • 中度青光眼:SITA快速算法使ΔMD较标准算法提高0.87 dB(95%CI 0.18-1.57)。
  • 晚期青光眼:SITA快速算法使ΔMD提高1.49 dB(95%CI 0.79-2.19),可能掩盖进展。

核心结论

  • 从SITA标准转换为SITA快速算法后,轻度青光眼视野表现相似,但中度和晚期青光眼MD值更高。
  • 对中重度青光眼患者,转换检查策略可能掩盖疾病进展。

研究设计与样本

  • 回顾性纵向研究,纳入421例(766眼)确诊或疑似青光眼患者,来自单中心真实临床环境。
  • 每眼按顺序接受两次SITA标准检查和一次SITA快速检查,检查间隔平均13.9个月。
  • 采用线性混合效应模型比较Standard-Standard序列与Standard-Faster序列的ΔMD差异,校正检查间隔和假阳性错误变化。

主要发现

  • 轻度或疑似青光眼:Standard-Faster与Standard-Standard序列的ΔMD差异为-0.23 dB(95%CI -0.60至0.15 dB),无统计学意义。
  • 中度青光眼:Standard-Faster序列较Standard-Standard序列ΔMD提高0.87 dB(95%CI 0.18-1.57 dB)。
  • 晚期青光眼:Standard-Faster序列较Standard-Standard序列ΔMD提高1.49 dB(95%CI 0.79-2.19 dB)。

临床怎么用

  • 对中度或晚期青光眼患者,由SITA标准转为SITA快速算法时需警惕MD值假性升高可能掩盖真实进展。
  • 轻度或疑似青光眼患者转换算法后视野表现无显著差异,可平稳过渡。

局限性

  • 回顾性设计,仅来自单中心真实临床数据。
  • 检查间隔平均13.9个月,可能影响纵向比较。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To determine the effect of transitioning from Swedish Interactive Thresholding Algorithm (SITA) Standard to SITA Faster on visual field (VF) performance in glaucomatous eyes with a broad spectrum of disease severity undergoing longitudinal VF testing in a real-world clinical setting. DESIGN: Retrospective, longitudinal study. PARTICIPANTS: A total of 421 patients (766 eyes) with manifest or suspect glaucoma followed at a single institution. METHODS: Each included eye received the following sequence of VF examinations during routine clinical care: (1) SITA Standard, (2) SITA Standard, and (3) SITA Faster (mean time between tests = 13.9 months). Intra-eye comparisons were made between the first 2 VFs (Standard-Standard sequence) and the last 2 VFs (Standard-Faster sequence). The primary dependent variable was the difference in mean deviation (MD) between the second and first VF of the sequence (ΔMD, calculated as MDVF2 - MDVF1). The primary independent variable was the VF sequence (Standard-Standard or Standard-Faster). Linear mixed-effects models were used to study the effect of testing sequence on ΔMD, adjusting for confounders including time between VFs and change in false-positive (FP) errors. Results were stratified to understand the effect of glaucoma severity on the relationship between testing sequence and ΔMD. MAIN OUTCOME MEASURES: The difference in ΔMD between Standard-Standard and Standard-Faster sequence by mild, moderate, and advanced disease severity. RESULTS: In eyes with mild or suspect glaucoma, there was no significant difference in ΔMD between Standard-Faster and Standard-Standard sequences (-0.23 decibels [dB]; 95% confidence interval [CI], -0.60 to 0.15 dB). However, the Standard-Faster sequence was associated with a 0.87 dB (95% CI, 0.18-1.57 dB) improvement in ΔMD compared with the Standard-Standard sequence in eyes with moderate glaucoma and a 1.49 dB (95% CI, 0.79-2.19 dB) improvement in ΔMD in eyes with advanced glaucoma. CONCLUSIONS: Converting to SITA Faster in eyes that were previously followed with SITA Standard led to similar VF performance in mild glaucoma but resulted in higher MD values in moderate and advanced glaucoma. For patients with moderate or severe glaucoma, this may conceal disease progression when transitioning testing strategies.