基于扫频OCT的深度学习算法实现360°房角评估
以压陷式房角镜为参考标准,算法诊断房角关闭的AUC为0.85(95%CI 0.80-0.90),灵敏度83%、特异度87%。该算法不依赖巩膜突识别,有望成为非接触、快速、可重复的「自动化房角镜」。
要点
- 深度学习算法基于扫频OCT 360°扫描诊断房角关闭,AUC达0.85。
- 在 circumferential angle closure >35% 的最佳截断值下,灵敏度83%、特异度87%。
- 算法不依赖巩膜突识别,为未来自动化房角镜奠定基础。
核心结论
- 深度学习算法在青光眼门诊环境下对360°扫频OCT扫描检测房角镜定义的房角关闭表现出良好的诊断性能。
研究设计与样本
- 连续招募50岁以上、无内眼手术史的青光眼门诊患者,行压陷式房角镜和暗室扫频OCT检查。
主要发现
- 受试者中约80%为中国人,女性占57.4%,房角镜诊断的房角关闭患病率为20.2%。
- 算法诊断房角镜房角关闭的AUC为0.85(95%CI 0.80-0.90),灵敏度83%、特异度87%。
关键数值与推荐
- 房角镜房角关闭定义为≥180°房角范围内后部小梁网不可见。
- 算法最佳截断值为环周房角关闭>35%。
临床怎么用
- 该算法可用于青光眼门诊中基于扫频OCT的房角关闭自动检测,辅助房角评估。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: To validate a deep learning (DL) algorithm (DLA) for 360° angle assessment on swept-source optical coherence tomography (SS-OCT) (CASIA SS-1000, Tomey Corporation, Nagoya, Japan). METHODS: This was a reliability analysis from a cross-sectional study. An independent test set of 39 936 SS-OCT scans from 312 phakic subjects (128 SS-OCT meridional scans per eye) was analysed. Participants above 50 years with no previous history of intraocular surgery were consecutively recruited from glaucoma clinics. Indentation gonioscopy and dark room SS-OCT were performed. Gonioscopic angle closure was defined as non-visibility of the posterior trabecular meshwork in ≥180° of the angle. For each subject, all images were analysed by a DL-based network based on the VGG-16 architecture, for gonioscopic angle-closure detection. Area under receiver operating characteristic curves (AUCs) and other diagnostic performance indicators were calculated for the DLA (index test) against gonioscopy (reference standard). RESULTS: Approximately 80% of the participants were Chinese, and more than half were women (57.4%). The prevalence of gonioscopic angle closure in this hospital-based sample was 20.2%. After analysing a total of 39 936 SS-OCT scans, the AUC of the DLA was 0.85 (95% CI:0.80 to 0.90, with sensitivity of 83% and a specificity of 87%) to classify gonioscopic angle closure with the optimal cut-off value of >35% of circumferential angle closure. CONCLUSIONS: The DLA exhibited good diagnostic performance for detection of gonioscopic angle closure on 360° SS-OCT scans in a glaucoma clinic setting. Such an algorithm, independent of the identification of the scleral spur, may be the foundation for a non-contact, fast and reproducible 'automated gonioscopy' in future.
引用
- Porporato N, Tun TA, Baskaran M, et al. Towards 'automated gonioscopy': a deep learning algorithm for 360° angle assessment by swept-source optical coherence tomography. Br J Ophthalmol. 2021. PMID: 33846160. doi:10.1136/bjophthalmol-2020-318275
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。