青光眼OCT血管成像中的伪影研究

OCT Angiography Artifacts in Glaucoma.
PubMed 文献 · Ophthalmology · 2021-04-02

该回顾性研究纳入368名健康、疑似青光眼及青光眼患者的649只眼,由4名专家评审5263幅OCTA图像中的伪影类型与图像质量。结果显示33.9%的图像质量差,较差视野平均偏差、高龄、男性、无眼动追踪及黄斑扫描区域与劣质图像相关;商业质量指标检测优质图像的能力有限,提示不应仅依赖仪器自动评估。

要点

  • 33.9%的OCTA图像质量差,伪影在健康眼与青光眼眼中均常见。
  • 高龄、男性、较差视野平均偏差、无眼动追踪及黄斑扫描与劣质图像相关。
  • QS与信号强度指数检测优质图像的AUC分别为0.65与0.70,不应仅依赖自动指标。

核心结论

  • OCTA中与劣质图像相关的伪影常见,其患病率受眼部与患者特征影响。
  • 不应仅依赖OCTA仪器自动提供的定量评估,应系统审阅扫描图像以确保正确解读。
  • 当图像存在明显且可纠正的伪影时,应重新采集图像。

研究设计与样本

  • 回顾性研究,纳入368名健康、疑似青光眼及青光眼患者的649只眼。
  • 由4名专家评审Angiovue高清与非高清视盘及黄斑OCTA图像中的伪影。
  • 共评估5263幅OCTA图像,按伪影有无判定图像质量为好或差。

主要发现

  • 33.9%的OCTA图像质量差。
  • 质量评分可接受(QS≥4)的图像中,76.6%无伪影,13.6%有1种伪影,9.8%有2种及以上伪影。
  • 高龄(P<0.001)、男性(P=0.045)、较差视野平均偏差(P<0.001)、无眼动追踪(P<0.001)及黄斑扫描区域(P<0.001)与获得劣质图像的可能性更高相关。
  • 在质量评分可接受的图像中,QS检测优质图像的AUC为0.65(95%CI 0.62-0.69),信号强度指数为0.70(95%CI 0.68-0.73)。

关键数值与推荐

  • 质量评分可接受定义为QS≥4。
  • QS检测优质图像的AUC为0.65(95%CI 0.62-0.69),信号强度指数为0.70(95%CI 0.68-0.73)。

临床怎么用

  • 解读OCTA图像时应系统审阅扫描,不能仅依赖仪器自动提供的QS或信号强度指数。
  • 发现明显且可纠正的伪影时应重新采集图像。
  • 对高龄、男性、视野平均偏差较差、无眼动追踪及黄斑扫描的图像应提高质量核查意识。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To determine the prevalence of different types of artifacts seen in OCT angiography (OCTA) images of healthy and glaucoma eyes and evaluate the characteristics associated with poor-quality images. DESIGN: Retrospective study. PARTICIPANTS: A total of 649 eyes of 368 healthy, glaucoma suspect, and glaucoma patients. METHODS: Angiovue (Optovue Inc) high-density (HD) and non-HD optic nerve head and macula OCTA images of participants were evaluated by 4 expert reviewers for the presence of different artifacts, including eye movement, defocus, shadow, decentration, segmentation error, blink, and Z offset in the superficial vascular layer. Each OCTA scan was designated to have good or poor quality based on the presence of artifacts. The association of demographic and ocular characteristics with the likelihood of obtaining poor-quality OCTA images was evaluated. MAIN OUTCOME MEASURES: The prevalence of OCTA artifacts and the factors associated with increased likelihood of capturing poor-quality OCTA images. RESULTS: A total of 5263 OCTA images were evaluated. Overall, 33.9% of the OCTA images had poor quality. The majority of images with acceptable quality scores (QS ≥ 4) had no artifacts (76.6%). Other images had 1 (13.6%) or 2 or more artifacts (9.8%). Older age (P < 0.001), male gender (P = 0.045), worse visual field mean deviation (P < 0.001), absence of eye tracking (P < 0.001), and macular scan area (P < 0.001) were associated with a higher likelihood of obtaining poor-quality images. In images with acceptable QS, the commercially available quality measures including QS and signal strength index had the area under the receiver operating characteristic curves of 0.65 (95% confidence interval [CI], 0.62-0.69) and 0.70 (95% CI, 0.68-0.73) to detect good-quality images, respectively. CONCLUSIONS: OCTA artifacts associated with poor-quality images are frequent, and their prevalence is affected by ocular and patient characteristics. One should not rely solely on the quantitative assessments that are provided automatically by OCTA instruments. A systematic scan review should be conducted to ensure appropriate interpretation of OCTA images. Given the high prevalence of poor-quality OCTA images, the images should be reacquired whenever an apparent and correctable artifact is present on a captured image.