智能移动应用远程监测糖尿病视网膜病变与年龄相关性黄斑变性的视觉功能:CLEAR研究
这项前瞻性多中心可靠性分析纳入108例年龄相关性黄斑变性、糖尿病视网膜病变或无视网膜病变的健康受试者,比较Checkup移动应用与临床标准视力及Amsler格测试。结果显示Checkup与参考测试的视力一致性r=0.86,Amsler格灵敏度93%、特异度92%,家庭与临床测试一致性r=0.96,支持其作为视网膜病变监测工具。
要点
- Checkup应用与临床参考视力测试一致性r=0.86。
- Amsler格测试灵敏度93%、特异度92%。
- 家庭与临床测试一致性r=0.96,无严重不良事件。
核心结论
- Checkup移动应用与临床视力及Amsler格测试结果一致性良好,支持其用于视网膜病变的临床和家庭监测。
研究设计与样本
- 前瞻性多中心可靠性分析,纳入108例近矫正视力优于20/200的年龄相关性黄斑变性、糖尿病视网膜病变患者或健康受试者。
主要发现
- Checkup与参考测试的视力一致性r=0.86,Amsler格灵敏度93%、特异度92%。
- 家庭与临床测试一致性r=0.96,患者报告家庭使用成功,无严重不良事件或中止。
- 患者对Checkup可用性评价为优秀。
关键数值与推荐
- 视力一致性r=0.86,Amsler格灵敏度93%、特异度92%,家庭与临床一致性r=0.96。
临床怎么用
- Checkup可用于糖尿病视网膜病变和年龄相关性黄斑变性患者的远程视力及Amsler格监测。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: The purpose of this study was to determine if a mobile application, the Checkup Vision Assessment System, could reliably monitor visual acuity (VA) and metamorphopsia remotely versus standard VA reference tests in the clinic. With the current COVID-19 pandemic, an even greater need for remote monitoring exists. Mobile tools enhance the ability to monitor patients virtually by enabling remote monitoring of VA and Amsler grid findings. DESIGN: Prospective, multicenter reliability analysis. METHODS: Participants: Patients (N = 108) with near corrected VA better than 20/200 and a diagnosis of age-related macular degeneration, diabetic retinopathy, or healthy patients without retinal disease (best-corrected visual acuity [BCVA] of 20/32 or better). INTERVENTION: participants were tested using the Checkup, reference VA, and Amsler tests, with the order of testing (Checkup or reference) randomized. Patients monitored their vision using Checkup at least twice a week at home between office visits. Main outcome measurements were near corrected VA and Amsler grid test results. RESULTS: Agreement was strong between Checkup and reference tests for VA (r = 0.86) and Amsler grid (sensitivity: 93%; specificity: 92%). Home versus clinic testing showed excellent agreement (r = 0.96). Patients reported successful home use. There were no serious adverse events or discontinuations. Patients rated the usability of Checkup to be excellent. CONCLUSIONS: There was good agreement between Checkup and in-clinic test results for VA and Amsler grid. The low variance of Checkup testing, agreement between in-clinic and home results, and excellent usability support Checkup as a reliable method for monitoring retinal pathology in clinic and home settings.
引用
- Khurana RN, Hoang C, Khanani AM, et al. A Smart Mobile Application to Monitor Visual Function in Diabetic Retinopathy and Age-Related Macular Degeneration: The CLEAR Study. Am J Ophthalmol. 2021. PMID: 33831342. doi:10.1016/j.ajo.2021.03.033
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。