该横断面研究使用Manifold对比视力计和qCSF算法,在标准和低亮度条件下评估不同严重程度干性年龄相关性黄斑变性(AMD)患者的对比敏感度功能。结果显示随着疾病严重程度增加,AULCSF进行性下降,且低亮度条件下组间差异更显著。低亮度缺陷在视力和AULCSF之间无相关性,提示亮度与对比敏感度的交互反映不同于视力与亮度交互的视觉功能方面。
要点
- qCSF在标准和低亮度下测量的对比敏感度与干性AMD进展阶段相关。
- 低亮度条件下中间型AMD与SDD、GA组间AULCSF差异显著。
- 低亮度缺陷在视力和AULCSF之间无相关性。
核心结论
- 通过qCSF在标准和低亮度条件下测量的对比敏感度与干性AMD进展阶段相关。
- 亮度与对比敏感度的交互似乎反映不同于视力与亮度交互的视觉功能方面。
研究设计与样本
- 横断面研究,纳入26例中间型AMD、19例AMD伴视网膜下玻璃膜疣样沉积(SDD)、20例地图样萎缩(GA)患者和23例年龄匹配对照。
- 使用Manifold对比视力计和qCSF算法在标准和低亮度条件下测试。
主要发现
- 随疾病严重程度增加,AULCSF进行性下降。
- 标准亮度下,对照与中间型AMD、SDD、GA组间以及中间型AMD与GA组间AULCSF差异显著(P < .0005和P < .005)。
- 低亮度下,对照与各疾病组间AULCSF差异均显著(均P < .0005),中间型AMD与SDD、GA组间差异显著(P < .005)。
- 任何AMD组中低亮度缺陷的视力与低亮度缺陷的AULCSF均无相关性。
临床怎么用
- qCSF可在标准和低亮度条件下评估干性AMD患者的对比敏感度功能。
- 低亮度对比敏感度可能提供不同于视力的视觉功能信息。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To assess the effectiveness of an active learning approach to measuring the contrast sensitivity function (CSF) in patients with various degrees of dry age-related macular degeneration (AMD) under multiple luminance conditions. DESIGN: Cross-sectional study. METHODS: Patients with AMD (26 intermediate AMD, 19 AMD with subretinal drusenoid deposits [SDD], 20 geographic atrophy [GA]) and 23 age-matched controls were tested with the Manifold Contrast Vision Meter (Adaptive Sensory Technology) and the qCSF algorithm, which applies active learning to estimate a model of the CSF's global shape. Testing was performed under conditions of standard and low luminance. For each AMD severity, the area under log CSF (AULCSF) and contrast sensitivities at individual spatial frequencies were calculated for analysis. Low-luminance deficits (LLDs) for visual acuity (VA) and AULCSF were calculated as the difference between standard and low luminance values. RESULTS: Progressive decreases in AULCSF were observed as disease severity increased. For standard luminance, pairwise comparisons revealed significant differences between control/intermediate AMD (P < .0005), control/SDD (P < .0005), control/GA (P < .0005), and intermediate AMD/GA (P < .005). Similarly, for low luminance, pairwise comparisons revealed significant differences between the controls and each disease group (all P < .0005), in addition to significant differences between intermediate AMD/SDD (P < .005), and intermediate AMD/GA (P < .005). No correlations were found between LLD VA and LLD AULCSF in any AMD groups. CONCLUSIONS: Contrast sensitivity measured via qCSF under both standard- and low-luminance conditions correlates with advancing stages of dry AMD. The interaction between luminance and contrast sensitivity appears to reflect a different aspect of visual function than the interaction between luminance and VA.
引用
- Ou WC, Lesmes LA, Christie AH, et al. Normal- and Low-Luminance Automated Quantitative Contrast Sensitivity Assessment in Eyes With Age-Related Macular Degeneration. Am J Ophthalmol. 2021. PMID: 33529583. doi:10.1016/j.ajo.2021.01.017
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。