文献速递角膜与眼表疾病青光眼

用巩膜Schiotz眼压值估算1型人工角膜眼的Goldmann压平眼压

Estimation of Goldmann applanation tonometer intraocular pressure (IOP) from scleral Schiotz IOP values in eyes with type-1 keratoprostheses.
PubMed 文献 · Br J Ophthalmol · 2020-12-08

这项前瞻性横断面研究在253只正常眼前段眼和100只1型人工角膜眼中,用非线性模型由巩膜Schiotz眼压预测Goldmann压平眼压(GAT)。正常眼中预测值与实际GAT值平均差异0.05 mmHg,一致性界限-5.39至5.5;人工角膜眼中模型预测分类与手指触压评估的一致性Kappa=0.81(95%CI 0.69至0.93),准确率91%(95%CI 0.84至0.96)。

要点

  • 正常眼中巩膜Schiotz预测GAT与实际GAT平均差异仅0.05 mmHg。
  • 1型人工角膜眼中模型预测与手指触压评估分类一致性Kappa=0.81。
  • 模型评估眼压准确率为91%(95%CI 0.84至0.96)。

核心结论

  • 巩膜Schiotz眼压值结合预测模型可作为1型人工角膜眼评估眼压的客观替代方法,替代手指触压评估。

研究设计与样本

  • 前瞻性横断面研究,队列1为253只正常眼前段眼,队列2为100只1型人工角膜眼。
  • 用巩膜Schiotz眼压测量值通过非线性模型预测GAT眼压值。
  • 正常眼用实际GAT值评估预测准确性,人工角膜眼用经验丰富的青光眼专家手指触压评估作参照。

主要发现

  • 正常眼实际GAT眼压范围6至62 mmHg(均值22±10.05 mmHg),预测与实际GAT平均差异0.05 mmHg,Bland-Altman一致性界限-5.39至5.5。
  • 100只1型人工角膜眼中,68只被分类为手指触压正常眼压,28只为临界,4只为高眼压。
  • 手指触压分类与模型预测GAT分类一致性Kappa=0.81(95%CI 0.69至0.93)。
  • 模型评估眼压准确率为91%(95%CI 0.84至0.96)。

关键数值与推荐

  • 正常眼预测GAT与实际GAT平均差异0.05 mmHg,一致性界限-5.39至5.5。
  • 人工角膜眼分类一致性Kappa=0.81(95%CI 0.69至0.93)。
  • 模型准确率91%(95%CI 0.84至0.96)。

临床怎么用

  • 对1型人工角膜眼,可用巩膜Schiotz眼压测量结合预测模型客观估算GAT眼压,辅助眼压评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To validate estimation of Goldmann applanation tonometer (GAT) intraocular pressure (IOP) from scleral Schiotz IOP measurements using a regression model in normal eyes and eyes with type-1 keratoprostheses. METHODS: In this prospective cross-sectional study, cohort-1 had 253 normal anterior segment eyes, and cohort-2 had 100 eyes with type-1 keratoprostheses. Scleral Schiotz IOP measurements were used (in a non-linear model) to predict GAT IOP values for these eyes. Accuracy of predicted GAT IOP values was assessed using actual GAT IOP values for normal eyes, while for type-1 keratoprosthetic eyes, finger tension (FT) IOP assessments by an experienced glaucoma specialist were used. Primary outcome was agreement between FT IOP (assessed by an experienced glaucoma specialist) and predicted GAT IOP-derived clusters. RESULTS: The actual values of GAT IOP measurements in normal eyes (n=253; mean age ±SD, 51.35±15.56 years) ranged between 6 mm Hg and 62 mm Hg (mean=22±10.05 mm Hg). Estimated and actual GAT IOP values for normal eyes were very similar (mean difference=0.05 mm Hg with limits of agreement: -5.39 to 5.5 by Bland-Altman plot). Of the 100 eyes with type-1 keratoprostheses, 68 were classified as having digitally normal IOP, 28 as borderline and 4 as high. The agreement between classification by FT assessment and model-predicted GAT IOP values was substantial (Kappa=0.81, 95% CI 0.69 to 0.93). The accuracy of the model in assessing IOP was found to be 91% (95% CI 0.84 to 0.96). CONCLUSION: Scleral Schiotz IOP values along with our predictive model can be an alternative objective method to FT IOP in assessing IOP in eyes with type-1 keratoprostheses.