三维OCTA指标提高糖尿病黄斑水肿眼缺血量化重复性
本研究回顾性纳入20例糖尿病视网膜病变合并糖尿病黄斑水肿患者(20眼),在同一次就诊内连续进行两次OCTA扫描,比较二维与三维OCTA指标在量化视网膜灌注方面的组内重复性。结果显示三维OCTA指标的组内相关系数(0.935–0.967)高于二维指标(0.591–0.824),变异系数更低,提示三维OCTA在糖尿病黄斑水肿眼中具有更高的重复性。
要点
- 三维OCTA指标的组内相关系数为0.935–0.967,高于二维指标的0.591–0.824。
- 三维OCTA变异系数为1.9–2.0,低于二维指标的2.2–4.2,重复性更优。
- 二维OCTA扫描间差异与平均黄斑体积相关,可能与分割错误率高有关。
核心结论
- 在糖尿病黄斑水肿眼中,三维OCTA指标的同次组内重复性高于二维OCTA指标。
研究设计与样本
- 回顾性纳入20例糖尿病视网膜病变合并糖尿病黄斑水肿患者(20眼),同一次就诊内连续两次OCTA扫描,比较二维与三维OCTA指标的组内重复性。
主要发现
- 二维OCTA指标的组内相关系数范围为0.591–0.824,三维OCTA指标为0.935–0.967。
- 三维OCTA指标的组内相关系数高于二维指标,且95%置信区间不重叠。
- 二维OCTA指标变异系数范围为2.2–4.2,三维OCTA指标为1.9–2.0。
- 二维OCTA指标扫描间差异与平均黄斑体积相关。
临床怎么用
- 在糖尿病黄斑水肿眼中量化视网膜灌注时,可优先考虑三维OCTA指标以提高重复性。
局限性
- 回顾性设计且样本量较小(20眼),结论需更大样本验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Two-dimensional (2D) optical coherence tomography angiography (OCTA) is known to be prone to segmentation errors, especially in pathologic eyes. Therefore, our aim was to systematically compare intrasession repeatability between repeated scans for 2D and three-dimensional (3D) OCTA metrics in quantifying retinal perfusion in eyes with diabetic macular edema. METHODS: Diabetic patients with diabetic retinopathy and diabetic macular edema who had two consecutive OCTA imaging scans obtained during the same visit were retrospectively included. A previously validated algorithm was applied to OCTA volume data to measure the 3D vascular volume and perfusion density. Optical coherence tomography angiography en face images were also processed to obtain 2D perfusion density metrics. RESULTS: Twenty patients (20 eyes) with diabetic retinopathy and diabetic macular edema were included. The intraclass correlation coefficient ranged from 0.591 to 0.824 for 2D OCTA metrics and from 0.935 to 0.967 for 3D OCTA metrics. Therefore, compared with the 2D OCTA analysis, the intraclass correlation coefficients of the 3D OCTA analysis were higher (without overlapping of the 95% confidential intervals). Similarly, the coefficient of variation (ranging from 2.2 to 4.2 for 2D OCTA metrics and from 1.9 to 2.0 for 3D OCTA metrics) indicated that the 3D OCTA-based quantifications had the highest interscan intrasession agreements. Differences in interscan 2D OCTA metrics' values were associated with average macular volume. CONCLUSION: Three-dimensional OCTA metrics have higher values of intrasession repeatability, as compared with 2D OCTA metrics. The latter finding seems to be related to the high rate of segmentation errors occurring in diabetic macular edema eyes.
引用
- Borrelli E, Parravano M, Costanzo E, et al. USING THREE-DIMENSIONAL OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY METRICS IMPROVES REPEATABILITY ON QUANTIFICATION OF ISCHEMIA IN EYES WITH DIABETIC MACULAR EDEMA. Retina. 2021. PMID: 33332812. doi:10.1097/IAE.0000000000003077
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。