光学相干断层扫描作为特发性颅内高压颅内压测量替代指标的研究
本研究基于三项随机临床试验的纵向队列数据,纳入104例活动性特发性颅内高压女性患者,评估OCT视神经头参数与颅内压的相关性。结果发现视神经头中央厚度与颅内压显著相关,且纵向随访证实该关联,提示OCT可用于无创监测颅内压变化。
要点
- OCT视神经头中央厚度与颅内压相关,右眼r=0.60,左眼r=0.73。
- 纵向随访12和24个月证实中央厚度与颅内压的相关性。
- 中央厚度减少50μm与颅内压下降5 cm H2O相关。
核心结论
- OCT视神经头体积测量(尤其是中央厚度)可重复地与颅内压相关,并能提示颅内压变化。
- OCT不仅可监测视乳头水肿,还可无创预测特发性颅内高压的颅内压水平。
研究设计与样本
- 纵向队列研究,数据来自2014年4月1日至2019年8月1日开展的三项随机临床试验。
- 纳入英国5家国民健康服务医院的活动性特发性颅内高压女性患者104例。
- 队列1(n=15)使用持续坐位遥测颅内压监测;队列2(n=89)在基线及3、12、24个月接受腰椎穿刺评估颅内压。
主要发现
- 队列1中视神经头中央厚度与颅内压最相关(右眼r=0.60,P=.02;左眼r=0.73,P=.002)。
- 队列2纵向证实中央厚度与颅内压的相关性(12和24个月)。
- Bootstrap替代分析显示中央厚度与颅内压变化在所有时间点呈正相关,如12个月时中央厚度减少50μm与颅内压下降5 cm H2O相关。
临床怎么用
- OCT视神经头中央厚度可作为特发性颅内高压患者无创评估颅内压的辅助指标。
- 临床可利用OCT监测视乳头水肿并预测颅内压变化,减少有创检查需求。
局限性
- 研究仅纳入女性特发性颅内高压患者,结果可能不适用于其他人群。
- 样本量较小(共104例),需更大规模研究验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
IMPORTANCE: There is an unmet need for noninvasive biomarkers of intracranial pressure (ICP), which manifests as papilledema that can be quantified by optical coherence tomography (OCT) imaging. OBJECTIVE: To determine whether OCT of the optic nerve head in papilledema could act as a surrogate measure of ICP. DESIGN, SETTING, AND PARTICIPANTS: This longitudinal cohort study used data collected from 3 randomized clinical trials that were conducted between April 1, 2014, and August 1, 2019. Participants who were female and had active idiopathic intracranial hypertension were enrolled from 5 National Health Service hospitals in the UK. Automated perimetry and OCT imaging were followed immediately by ICP measurement on the same day. Cohort 1 used continuous sitting telemetric ICP monitoring (Raumedic Neurovent P-tel device) on 1 visit. Cohort 2 was evaluated at baseline and after 3, 12, and 24 months and underwent lumbar puncture assessment of ICP. MAIN OUTCOMES AND MEASURES: Optical coherence tomography measures of the optic nerve head and macula were correlated with ICP levels, Frisén grading, and perimetric mean deviation. The OCT protocol included peripapillary retinal nerve fiber layer, optic nerve head, and macular volume scans (Spectralis [Heidelberg Engineering]). All scans were validated for quality and resegmented manually when required. RESULTS: A total of 104 women were recruited. Among cohort 1 (n = 15; mean [SD] age, 28.2 [9.4] years), the range of OCT protocols was evaluated, and optic nerve head central thickness was found to be most closely associated with ICP (right eye: r = 0.60; P = .02; left eye: r = 0.73; P = .002). Subsequently, findings from cohort 2 (n = 89; mean [SD] age, 31.8 [7.5] years) confirmed the correlation between central thickness and ICP longitudinally (12 and 24 months). Finally, bootstrap surrogacy analysis noted a positive association between central thickness and change in ICP at all points (eg, at 12 months, a decrease in central thickness of 50 μm was associated with a decrease in ICP of 5 cm H2O). CONCLUSIONS AND RELEVANCE: In this study, optic nerve head volume measures on OCT (particularly central thickness) reproducibly correlated with ICP and surrogacy analysis demonstrated its ability to inform ICP changes. These data suggest that OCT has the utility to not only monitor papilledema but also noninvasively prognosticate ICP levels in idiopathic intracranial hypertension.
引用
- Vijay V, Mollan SP, Mitchell JL, et al. Using Optical Coherence Tomography as a Surrogate of Measurements of Intracranial Pressure in Idiopathic Intracranial Hypertension. JAMA Ophthalmol. 2020. PMID: 33090189. doi:10.1001/jamaophthalmol.2020.4242
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。