参考扫描提升OCTA测量在正常眼与青光眼中的重复性
该横断面研究比较了参考扫描与非参考扫描在正常、可疑青光眼和青光眼眼中视盘周围灌注密度与血流指数的重复性。结果显示参考扫描的重复性系数(CRw)和变异系数(CVw)均低于非参考扫描,灌注密度的变异性低于血流指数。结论认为参考扫描的OCTA灌注密度测量更适合监测青光眼血管变化。
要点
- 参考扫描的OCTA测量重复性优于非参考扫描。
- 灌注密度的变异性低于血流指数。
- 监测青光眼血管变化时优选参考扫描的灌注密度。
核心结论
- 参考扫描的OMAG测量重复性估计值优于非参考扫描。
- 灌注密度测量的变异性低于血流指数。
- 参考扫描的OCTA灌注密度适合用于监测青光眼血管变化。
研究设计与样本
- 横断面研究,48眼(33例)接受三次非参考视盘周围OMAG扫描,43眼(25例)接受三次参考扫描。
- 参考扫描组中第二、三次扫描使用设备「track to prior scan」选项精确对准基线扫描。
主要发现
- 非参考扫描灌注密度CRw范围为2.0-4.1%,显著高于参考扫描的1.4-2.7%。
- 非参考与参考扫描的CVw范围分别为1.7-3.1%和1.2-2.1%。
- 血流指数非参考与参考扫描的CRw范围分别为4.4-5.8和3.6-4.8。
- 血流指数非参考与参考扫描的CVw范围分别为4.1-5.2和3.3-4.5。
临床怎么用
- 在青光眼随访中优先使用参考扫描模式获取OCTA灌注密度。
- 灌注密度比血流指数更稳定,更适合监测血管变化。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIM: To compare the repeatability of peripapillary perfusion density and flux index measurements on referenced and non-referenced optical microangiography (OMAG) scans in normal, glaucoma suspect and glaucoma eyes. METHODS: In a cross-sectional study, 48 eyes (33 subjects) underwent three repeat, non-referenced peripapillary OMAG scans in the same session and 43 eyes (25 subjects) underwent three referenced peripapillary OMAG scans. In the referenced scan group, repeat scans (second and the third scan) were acquired exactly on the baseline (first) scan using the 'track to prior scan' option on the device. Repeatability estimates of the mean and four-sector (temporal, superior, nasal and inferior) OMAG measurements on the non-referenced and referenced scans were assessed using within-subject coefficient of repeatability (CRw) and variation (CVw). RESULTS: CRw (%) of peripapillary perfusion density measurements (range: 2.0-4.1) on non-referenced scans were significantly higher than that on referenced scans (range: 1.4-2.7). CVw (%) on non-referenced and referenced scans ranged from 1.7 to 3.1 and from 1.2 to 2.1, respectively . CRw of flux index on non-referenced and referenced scans ranged from 4.4 to 5.8 and from 3.6 to 4.8, respectively. CVw on non-referenced and referenced scans ranged from 4.1 to 5.2 and from 3.3 to 4.5, respectively. CONCLUSIONS: Repeatability estimates of OMAG measurements were better on referenced scans compared with non-referenced scans. Perfusion density measurements had lower variability than flux index. OCTA-measured perfusion density of referenced scans is preferable for monitoring vascular change in glaucoma.
引用
- Rao HL, Dasari S, Riyazuddin M, et al. Referenced scans improve the repeatability of optical coherence tomography angiography measurements in normal and glaucoma eyes. Br J Ophthalmol. 2020. PMID: 32962991. doi:10.1136/bjophthalmol-2020-316480
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。