非青光眼与青光眼眼视盘周围OCTA血管参数的同次重复性与跨次再现性

Intrasession repeatability and intersession reproducibility of peripapillary OCTA vessel parameters in non-glaucomatous and glaucomatous eyes.
PubMed 文献 · Br J Ophthalmol · 2020-09-11

该观察性纵向研究比较了基于光学微血管造影的OCTA在非青光眼和青光眼眼中视盘周围血管参数的同次重复性与跨次再现性,使用眼内变异系数(CVW)和组内相关系数(ICC)评估。非青光眼120眼同次CVW为VAD 4.2%、flux 5.3%、PD 1.5%、FI 2.0%,跨次CVW为6.5%、8.0%、2.0%、3.2%;青光眼118眼同次CVW为9.0%、10.3%、1.7%、2.3%,跨次CVW为12.1%、14.2%、2.3%、3.5%。结论为两组同次重复性均优于跨次再现性,且商用量化软件一致性优于自定义研究软件。

要点

  • 非青光眼与青光眼眼视盘周围OCTA血管参数的同次重复性均优于跨次再现性。
  • 商用量化软件(Cirrus 11.0)计算PD和FI的一致性优于自定义研究软件计算的VAD和flux。
  • 信号强度与两组OCTA血管参数均呈显著正相关(p<0.0001)。

核心结论

  • 在非青光眼和青光眼眼中,视盘周围OCTA血管参数的同次重复性均优于跨次再现性。
  • 内置商用量化软件的一致性优于自定义研究导向量化软件。

研究设计与样本

  • 观察性纵向研究,纳入120只非青光眼眼和118只青光眼眼。
  • 使用基于光学微血管造影的OCTA采集视盘周围扫描,以眼内变异系数(CVW)和组内相关系数(ICC)评估同次重复性与跨次再现性。
  • 图像量化使用自定义研究导向软件计算血管面积密度(VAD)和flux,以及商用临床导向软件(Cirrus 11.0, Carl Zeiss Meditec)计算灌注密度(PD)和flux指数(FI)。
  • 评估信号强度对OCTA参数可靠性的影响。

主要发现

  • 非青光眼眼同次CVW:VAD 4.2%、flux 5.3%、PD 1.5%、FI 2.0%;跨次CVW:VAD 6.5%、flux 8.0%、PD 2.0%、FI 3.2%。
  • 非青光眼眼同次ICC范围0.928至0.945,跨次ICC范围0.811至0.866。
  • 青光眼眼同次CVW:VAD 9.0%、flux 10.3%、PD 1.7%、FI 2.3%;跨次CVW:VAD 12.1%、flux 14.2%、PD 2.3%、FI 3.5%。
  • 青光眼眼同次ICC范围0.904至0.972,跨次ICC范围0.855至0.955。
  • 信号强度与两组OCTA血管参数均呈显著正相关(p<0.0001)。

关键数值与推荐

  • 非青光眼眼同次CVW:VAD 4.2%、flux 5.3%、PD 1.5%、FI 2.0%;跨次CVW:VAD 6.5%、flux 8.0%、PD 2.0%、FI 3.2%。
  • 非青光眼眼同次ICC 0.928–0.945,跨次ICC 0.811–0.866。
  • 青光眼眼同次CVW:VAD 9.0%、flux 10.3%、PD 1.7%、FI 2.3%;跨次CVW:VAD 12.1%、flux 14.2%、PD 2.3%、FI 3.5%。
  • 青光眼眼同次ICC 0.904–0.972,跨次ICC 0.855–0.955。
  • 信号强度与OCTA血管参数显著正相关(p<0.0001)。

临床怎么用

  • 解读视盘周围OCTA血管参数时应考虑同次与跨次测量差异,跨次比较需谨慎。
  • 商用量化软件(Cirrus 11.0)计算的PD和FI可能具有更好的一致性,适合临床随访。
  • 信号强度显著影响OCTA血管参数,临床采集时应关注信号强度质量。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To compare intrasession repeatability versus intersession reproducibility of the peripapillary vessel parameters using optical microangiography-based optical coherence tomography angiography (OCTA) in non-glaucomatous and glaucomatous eyes. METHODS: In an observational, longitudinal study, peripapillary OCTA scans were collected to evaluate intrasession repeatability and intersession reproducibility using within-eye coefficient of variation (CVW) and intraclass correlation coefficient (ICC). Images were quantified using a custom research-oriented quantification software calculating vessel area density (VAD) and flux and a commercially developed, clinic-oriented quantification software (Cirrus 11.0, Carl Zeiss Meditec) calculating perfusion density (PD) and flux index (FI). Effect of signal strength on the reliability of OCTA parameters was also evaluated. RESULTS: Among 120 non-glaucomatous eyes, intrasession CVW were 4.2% for VAD, 5.3% for flux, 1.5% for PD and 2.0% for FI. The intersession CVW were 6.5% for VAD, 8.0% for flux, 2.0% for PD and 3.2% for FI. The intrasession ICC ranged from 0.928 to 0.945, and intersession ICC ranged from 0.811 to 0.866. From 118 glaucomatous eyes, intrasession CVW was 9.0% for VAD, 10.3% for flux, 1.7% for PD and 2.3% for FI. The intersession CVW was 12.1% for VAD, 14.2% for flux, 2.3% for PD and 3.5% for FI. The intrasession ICC ranged from 0.904 to 0.972, and intersession ICC ranged from 0.855 to 0.955. Signal strength was significantly positively associated with OCTA vessel parameters (p<0.0001) for both groups. CONCLUSION: Peripapillary OCTA vessel parameters had greater intrasession repeatability compared to intersession reproducibility in both non-glaucomatous and glaucomatous eyes. The built-in commercially developed quantification software demonstrated greater agreement than the custom research-oriented quantification software.

来源:Br J Ophthalmol,查看原文。版权归原作者所有。