圆锥角膜最大曲率点区域平均值与单点测量的重复性比较

Repeatability of Zone Averages Compared to Single-Point Measurements of Maximal Curvature in Keratoconus.
PubMed 文献 · Am J Ophthalmol · 2020-08-13

本研究纳入65例圆锥角膜患者的65只眼,用Galilei双Scheimpflug-Placido系统连续扫描3次,比较以最大曲率点(Kmax)为中心的区域平均值与单点Kmax的重复性。结果显示,瞬时曲率区域平均值较轴向区域平均值重复性改善更大,在Kmax≤50 D眼取半径至少1.5 mm、Kmax>50 D眼取2.0 mm半径时达到临床可接受水平。

要点

  • 瞬时曲率区域平均值重复性优于轴向区域平均值。
  • Kmax≤50 D眼取≥1.5 mm半径区域平均值重复性临床可接受。
  • Kmax>50 D眼取2.0 mm半径瞬时区域平均值重复性临床可接受。

核心结论

  • 以Kmax为中心的瞬时曲率区域平均值较轴向区域平均值重复性改善更大,在特定半径下达到临床可接受水平。

研究设计与样本

  • 比较可靠性分析,纳入65例圆锥角膜患者的65只眼,连续3次Galilei双Scheimpflug-Placido系统扫描。

主要发现

  • 轴向曲率区域平均值仅在Kmax≤50 D眼、半径1.5 mm和2.0 mm时重复性临床可接受(r=0.87 D和r=0.76,单点轴向Kmax r=0.91)。
  • 瞬时曲率区域平均值在Kmax≤50 D眼半径至少1.5 mm时重复性临床可接受(r=0.99 D和r=0.70 D)。
  • 瞬时曲率区域平均值在Kmax>50 D眼半径2.0 mm时重复性临床可接受(r=2.28 D和r=0.87 D)。
  • 所有眼Kmax位置的重复性界限轴向曲率为0.82 mm、瞬时曲率为0.80 mm。

关键数值与推荐

  • 轴向曲率区域平均值临床可接受重复性:Kmax≤50 D,半径1.5 mm(r=0.87 D)和2.0 mm(r=0.76)。
  • 瞬时曲率区域平均值临床可接受重复性:Kmax≤50 D,半径至少1.5 mm(r=0.99 D和r=0.70 D)。
  • 瞬时曲率区域平均值临床可接受重复性:Kmax>50 D,半径2.0 mm(r=2.28 D和r=0.87 D)。
  • Kmax位置重复性界限:轴向0.82 mm,瞬时0.80 mm。

临床怎么用

  • 在圆锥角膜Kmax测量中,可优先考虑瞬时曲率区域平均值以提高重复性。
  • 根据Kmax值选择区域半径:Kmax≤50 D用至少1.5 mm,Kmax>50 D用2.0 mm。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: To evaluate the repeatability of curvature zone averages centered on the point of maximum curvature (Kmax) compared to that of the single-point Kmax. DESIGN: Comparative reliability analysis. METHODS: Setting: American University of Beirut Medical Center, Beirut, Lebanon. STUDY POPULATION: Sixty-five eyes of 65 adult keratoconus patients. Patients with other ocular disease, history of ocular surgery or trauma, and contact lens wear within 2 weeks of image acquisition were excluded. OBSERVATION PROCEDURES: Eyes were evaluated with 3 consecutive scans using the Galilei dual Scheimpflug-Placido system. MAIN OUTCOME MEASURES: Repeatability of axial and instantaneous Kmax single points, and zone averages with radii of 0.1-2.0 mm, centered on them. Repeatability was assessed by within-subject standard deviations, repeatability limits (r), and intraclass correlation coefficients. RESULTS: Axial curvature zone averaging yielded clinically acceptable repeatability only in eyes with Kmax ≤50 diopters (D), for radii of 1.5 mm and 2.0 mm (r = 0.87 D and r = 0.76, respectively, vs r = 0.91 for the single-point axial Kmax). Compared to instantaneous Kmax, clinically acceptable repeatability was achieved with instantaneous zone averages of at least 1.5 mm radius in eyes with Kmax ≤50 D (r = 0.99 D and r = 0.70 D, respectively) and 2.0 mm radius in eyes with Kmax >50 D (r = 2.28 D and r = 0.87 D, respectively). For all eyes, the repeatability limit of the location of Kmax was 0.82 mm and 0.80 mm for axial and instantaneous curvature, respectively. CONCLUSIONS: Instantaneous curvature zone averages centered on Kmax yielded a greater improvement in repeatability than axial zone averages and reached clinical adequacy with radii of at least 1.5 mm, for eyes with Kmax ≤50 D, and with a 2.0 mm radius for eyes with Kmax >50 D.