文献速递白内障与屈光手术

人工晶状体屈光力公式、生物测量与术中像差测量的综述

Intraocular Lens Power Formulas, Biometry, and Intraoperative Aberrometry: A Review.
PubMed 文献 · Ophthalmology · 2020-08-13

该综述检索PubMed近10年文献1313篇,纳入166篇,评估人工晶状体屈光力公式、生物测量设备及术中像差测量对白内障术后屈光结局的影响。Kane公式在全眼轴谱及短眼、长眼亚组中最准确;各生物测量设备重复性好且对IOL屈光力选择影响小;术中像差测量在未使用特定公式时改善球性与散光IOL结局。

要点

  • Kane公式在全眼轴谱及短眼(眼轴≤22.0mm)、长眼(眼轴≥26.0mm)亚组中最准确。
  • 短眼亚组中Olsen(4因子)、Haigis、Hill-RBF 1.0表现良好;长眼亚组中BUII、Olsen(4因子)、Holladay 1联合Wang-Koch调整表现良好。
  • 各生物测量设备测量可重复性高,设备间平均测量差异对IOL屈光力选择影响小;扫频OCT设备在致密白内障中测量成功率更高。

核心结论

  • Kane公式在全眼轴谱及短眼、长眼亚组中最为准确,但术中像差测量的增益取决于是否已使用特定公式。

研究设计与样本

  • 综述检索PubMed数据库近10年文献共1313篇,最终纳入166篇。

主要发现

  • Kane公式在全眼轴谱以及短眼(眼轴≤22.0mm)和长眼(眼轴≥26.0mm)亚组中最为准确。
  • 短眼亚组中Olsen(4因子)、Haigis和Hill-RBF 1.0表现良好;长眼亚组中BUII、Olsen(4因子)或Holladay 1联合Wang-Koch调整表现良好。
  • 所有生物测量设备均提供高度可重复的测量,多数比较研究显示设备间各生物测量变量平均测量值差异小,对IOL屈光力选择影响小。
  • 设备间主要差异在于致密白内障的测量成功率,扫频OCT设备优于部分相干干涉和光学低相干反射设备。
  • 术中像差测量在未使用BUII、Hill-RBF、Barrett散光计算器或Barrett True-K公式时,对有或无屈光手术史的球性与散光IOL眼改善结局。

关键数值与推荐

  • 短眼定义为眼轴≤22.0mm,长眼定义为眼轴≥26.0mm。
  • 当使用BUII、Hill-RBF、Barrett散光计算器或Barrett True-K公式时,术中像差测量未带来更好结局。

临床怎么用

  • 选择IOL屈光力公式时可优先考虑Kane公式,短眼和长眼亚组可参考Olsen(4因子)、Haigis、Hill-RBF 1.0、BUII或Holladay 1联合Wang-Koch调整。
  • 生物测量设备选择对IOL屈光力影响小,但致密白内障患者可优先选用扫频OCT设备以提高测量成功率。
  • 考虑术中像差测量前应评估是否已使用BUII、Hill-RBF、Barrett散光计算器或Barrett True-K公式。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

The refractive outcome of cataract surgery is influenced by the choice of intraocular lens (IOL) power formula and the accuracy of the various devices used to measure the eye (including intraoperative aberrometry [IA]). This review aimed to cover the breadth of literature over the previous 10 years, focusing on 3 main questions: (1) What IOL power formulas currently are available and which is the most accurate? (2) What biometry devices are available, do the measurements they obtain differ from one another, and will this cause a clinically significant change in IOL power selection? and (3) Does IA improve refractive outcomes? A literature review was performed by searching the PubMed database for articles on each of these topics that identified 1313 articles, of which 166 were included in the review. For IOL power formulas, the Kane formula was the most accurate formula over the entire axial length (AL) spectrum and in both the short eye (AL, ≤22.0 mm) and long eye (AL, ≥26.0 mm) subgroups. Other formulas that performed well in the short-eye subgroup were the Olsen (4-factor), Haigis, and Hill-radial basis function (RBF) 1.0. In the long-eye group, the other formulas that performed well included the Barrett Universal II (BUII), Olsen (4-factor), or Holladay 1 with Wang-Koch adjustment. All biometry devices delivered highly reproducible measurements, and most comparative studies showed little difference in the average measures for all the biometric variables between devices. The differences seen resulted in minimal clinically significant effects on IOL power selection. The main difference found between devices was the ability to measure successfully through dense cataracts, with swept-source OCT-based machines performing better than partial coherence interferometry and optical low-coherence reflectometry devices. Intraoperative aberrometry generally improved outcomes for spherical and toric IOLs in eyes both with and without prior refractive surgery when the BUII and Hill-RBF, Barrett toric calculator, or Barrett True-K formulas were not used. When they were used, IA did not result in better outcomes.

来源:Ophthalmology,查看原文。版权归原作者所有。

引用

  1. Kane JX, Chang DF. Intraocular Lens Power Formulas, Biometry, and Intraoperative Aberrometry: A Review. Ophthalmology. 2020. PMID: 32798526. doi:10.1016/j.ophtha.2020.08.010

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。