两种飞秒激光辅助白内障手术平台临床应用对比分析
前瞻性随机对照研究纳入86例(90眼)行飞秒激光辅助白内障手术患者,随机分为LenSx组和LenSAR组各45眼。结果显示LenSAR对接便捷、结膜下出血率低,LenSx总负压吸引时间短、眼压升高幅度低,两组囊膜切开完整性及并发症无显著差异。
要点
- LenSAR组前囊膜切开时间3.00s、结膜下出血率8.9%,均优于LenSx组。
- LenSx组总负压吸引时间97.23s、眼压升高2.99mmHg,均低于LenSAR组。
- 两组囊膜切开完整性、圆度、居中性及撕裂率、后囊破裂率无显著差异。
核心结论
- LenSAR系统对接过程便捷且结膜下出血少,LenSx系统总负压吸引时间短且眼压升高较低。
研究设计与样本
- 前瞻性随机对照研究,纳入2018年4至11月武汉爱尔眼科医院行飞秒激光辅助白内障手术患者86例(90眼),随机分为LenSx组44例(45眼)和LenSAR组42例(45眼)。
主要发现
- LenSx组尝试对接次数为1(1~4)次,LenSAR组为1(1~2)次,差异有统计学意义(Z=-2.23,P<0.05)。
- LenSx组前囊膜切开时间13.00(10.00~22.00)s,LenSAR组3.00(1.00~3.00)s,差异有统计学意义(Z=-8.71,P<0.05)。
- LenSx组预劈核时间(16.67±3.36)s、总发射时间(30.49±3.53)s,LenSAR组分别为(12.38±4.36)s、(15.36±4.29)s,差异均有统计学意义(t=-5.23、-18.26,均P<0.05)。
- LenSx组总负压吸引时间(97.23±19.96)s,短于LenSAR组的(123.76±16.81)s(t=6.82,P<0.05)。
- LenSx组术后眼压升高(2.99±4.66)mmHg,低于LenSAR组的(5.64±5.42)mmHg(t=-2.49,P<0.05);LenSx组结膜下出血率33.3%(15/45),高于LenSAR组的8.9%(4/45)(χ²=6.67,P<0.05)。
关键数值与推荐
- 两组年龄和晶状体核密度差异均无统计学意义(均P>0.05)。
- 两组晶状体前囊膜完整切开率、前囊膜切口圆度、前囊膜切口全周包裹人工晶状体光学部、前囊膜切口撕裂率、晶状体后囊膜破裂率差异均无统计学意义(均P>0.05)。
临床怎么用
- 偏好对接便捷和低结膜下出血率可选LenSAR系统,偏好短负压吸引时间和低眼压升高可选LenSx系统。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Objective: To evaluate the operability and clinical application effects of femtosecond laser-assisted cataract surgery systems of LenSx and LenSAR. Methods: This was a randomized controlled study. A total of 86 patients (90 eyes) who underwent femtosecond laser-assisted cataract surgery in Wuhan Aier Eye Hospital from April 2018 to November 2018 were enrolled and divided into two groups randomly, including 44 patients (45 eyes) in the LenSx group and 42 patients (45 eyes) in the LenSAR group. During the operation, the following observation indexes were obtained. Operational indicators included the number of docking attempts, anterior capsulotomy time, nucleus pre-treatment time, total femtosecond laser emission time, and total vacuum suction duration. Clinical outcome indicators included changes in the patient's intraocular pressure during femtosecond laser surgery, the rate of subconjunctival hemorrhage, capsulotomy integrity (yes/no), roundness and centricity of the anterior capsule opening (yes/no), the rate of anterior capsule opening tear, and the rate of posterior capsule rupture. The t-test, rank-sum test or chi-square test were used for statistical analysis. Results: There were no significant differences between groups in the age and the lens density (both P>0.05). The number of docking attempts in the LenSx group was 1 (1 to 4) and in the LenSAR group was 1 (1 to 2); there was statistically significant difference (Z =-2.23, P<0.05). The difference in the anterior capsulotomy time between the two groups was statistically significant [13.00 (10.00 to 22.00) s compared with 3.00 (1.00 to 3.00) s, Z=-8.71, P<0.05]. The femtosecond laser pre-nucleation time and total femtosecond laser emission time of the LenSx group were (16.67±3.36) s and (30.49±3.53) s, and those of the LenSAR group were (12.38±4.36) s and (15.36±4.29) s, respectively; the differences between the two groups were statistically significant (t=-5.23, -18.26; both P<0.05). The total vacuum suction duration in the LenSx group was (97.23±19.96) s, shorter than that in the LenSAR group [(123.76±16.81) s] (t=6.82, P<0.05). The intraocular pressure after femtosecond laser surgery in both groups was higher than that before surgery. The increase of intraocular pressure in the LenSAR group was (5.64±5.42) mmHg (1 mmHg=0.133 kPa), higher than that in the LenSx group [(2.99±4.66) mmHg] (t=-2.49, P<0.05). The rate of subconjunctival hemorrhage in the LenSx group was 33.3% (15/45), while it was 8.9% (4/45) in the LenSAR group; the difference between the two groups was statistically significant (χ²=6.67, P<0.05). There were no significant differences between groups in capsulotomy integrity, roundness and centricity of the anterior capsule opening, the rate of anterior capsule opening tear, and the rate of posterior capsule rupture (all P>0.05). Conclusion: The docking process of the LenSAR system is convenient, and there is less subconjunctival hemorrhage; the total vacuum suction duration of LenSx is short, and the increase of intraocular pressure is low. (Chin J Ophthalmol, 2020, 56: 530-535). 目的: 评价LenSx和LenSAR两种不同飞秒激光辅助白内障摘除手术操作平台的临床应用效果。 方法: 前瞻性随机对照研究。选择2018年4至11月在武汉爱尔眼科医院行飞秒激光辅助白内障摘除手术患者86例(90只眼),采用随机信封法分为LenSx组44例(45只眼),LenSAR组42例(45只眼)。手术过程中获得下列观察指标结果并进行组间比较。操作性指标包括尝试对接次数、飞秒激光前囊膜切开时间、预劈核时间、飞秒激光发射总时间、负压吸引眼球的总时长;临床效果指标包括飞秒激光前后的眼压变化、结膜下出血率、晶状体前囊膜是否切开完整、前囊膜切口圆度、前囊膜切口居中性(前囊膜切口是否全周包裹人工晶状体光学部)、前囊膜切口撕裂率、晶状体后囊膜破裂率。采用t检验、秩和检验及χ²检验进行统计学分析。 结果: 两组年龄、晶状体核硬度差异均无统计学意义(均P>0.05)。LenSx组和LenSAR组尝试对接次数分别为1(1~4)、1(1~2)次,飞秒激光前囊膜切开时间分别为13.00(10.00~22.00)、3.00(1.00~3.00)s,飞秒激光预劈核时间分别为(16.67±3.36)、(12.38±4.36)s,飞秒激光发射总时间分别为(30.49±3.53)、(15.36±4.29)s,负压吸引固定眼球的总时长分别为(97.23±19.96)、(123.76±16.81)s,两组间差异均有统计学意义(Z=-2.23,-8.71;t=-5.23,-18.26,6.82;均P<0.05)。两组患眼飞秒激光术后眼压平均值均较术前升高,LenSx组和LenSAR组术后与术前眼压差值分别为(2.99±4.66)、(5.64±5.42)mmHg(1 mmHg=0.133 kPa);结膜下出血率分别为33.3%(15/45)、8.9%(4/45),两组间差异均有统计学意义(t=-2.49,χ²=6.67;均P<0.05)。两组晶状体前囊膜完整切开率、前囊膜切口圆度、前囊膜切口全周包裹人工晶状体光学部、前囊膜切口撕裂率、晶状体后囊膜破裂率差异均无统计学意义(均P>0.05)。 结论: LenSAR对接过程便捷,结膜下出血少;LenSx总负压吸引时间短,眼压升高较低。(中华眼科杂志,2020,56:530-535).
引用
- Xu YX, Wang Y, Sun M, et al. [Comparative analysis of the clinical application of two femtosecond laser-assisted cataract operation platforms]. Zhonghua Yan Ke Za Zhi. 2020. PMID: 32842336. doi:10.3760/cma.j.cn112142-20191113-00571
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。