三种常用人工晶状体长期旋转稳定性的比较
这项前瞻性干预性比较临床研究纳入381眼年龄相关性白内障,随机植入Acrysof SN60WF、Tecnis ZCB00或Envista MX60人工晶状体,比较术后即刻至4-7个月的绝对旋转度。三组均出现可能需要二次干预的离群值。
要点
- Acrysof组绝对旋转最小,显著优于Envista组,与Tecnis组无显著差异。
- 三组最大旋转值分别为15.8°、38.6°、44.9°,均存在离群值。
- 术后第一小时旋转幅度最大。
核心结论
- Acrysof人工晶状体绝对旋转量最小,Tecnis与Envista组间无显著差异。
- 所有组均观察到可能需要二次干预的旋转离群值。
- 术后第一小时旋转幅度最大。
研究设计与样本
- 前瞻性干预性比较临床研究,在维也纳医科大学眼科开展。
- 共纳入381眼(199例)年龄相关性白内障患者,随机植入Acrysof SN60WF、Tecnis ZCB00或Envista MX60人工晶状体。
- 术中患者仰卧于手术台完成基线测量,术后1小时、1周、1个月和6个月随访。
- 主要结局指标为术后即刻至6个月的绝对旋转度差异。
主要发现
- 术后即刻至6个月绝对旋转:Acrysof组1.65±2.1°,Tecnis组2.65±4.1°,Envista组3.18±5.8°。
- 最大旋转值:Acrysof组15.8°,Tecnis组38.6°,Envista组44.9°。
临床怎么用
- 对于旋转稳定性要求高的患者,Acrysof可能提供更优的长期旋转稳定性。
- 术后早期尤其是第一小时应关注旋转情况,所有人工晶状体均需警惕离群旋转。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To compare rotational stability and its influencing factors in 3 different widely used hydrophobic acrylic intraocular lenses (IOLs) from the end of surgery (EoS) to 4-7 months (6 months) in over 380 eyes. DESIGN: Prospective interventional comparative clinical study. METHODS: Setting: Department of Ophthalmology, Medical University of Vienna. PATIENT POPULATION: A total of 381 eyes of 199 patients with age-related cataract received an IOL Acrysof SN60WF, Tecnis ZCB00, or Envista MX60 in a consecutive order. INTERVENTION: Implantation of an Acrysof, Tecnis, or Envista IOL randomized to the 0 ± 10, 45 ± 10, 90 ± 10, or 135 ± 10° axis in 1 or both eyes. Baseline measurement was performed with patients supine still on the operating table. Postoperative follow-ups were conducted after 1 hour, 1 week, 1 month, and 6 months. MAIN OUTCOME MEASURES: Difference of absolute rotation from the EoS to 6 months. RESULTS: Absolute rotations from the EoS to 6 months were 1.65 ± 2.1, 2.65 ± 4.1, and 3.18 ± 5.8° for the Acrysof, Tecnis, and Envista group, respectively. Rotational stability was statistically significantly superior in the Acrysof compared with the Envista group (P = .014), but not compared with the Tecnis group (P = .10). No significant difference was found between the Tecnis and Envista groups (P = .761). Maximum values of 15.8, 38.6, and 44.9° were observed for the Acrysof, Tecnis, and Envista group, respectively. CONCLUSION: The Acrysof IOL showed the least amount of absolute rotation compared with the Tecnis and Envista IOLs. Outliers possibly requiring secondary intervention were observed in all groups. The amount of rotation was greatest during the first postoperative hour.
引用
- Schartmüller D, Schwarzenbacher L, Meyer EL, et al. Comparison of Long-Term Rotational Stability of Three Commonly Implanted Intraocular Lenses. Am J Ophthalmol. 2020. PMID: 32702360. doi:10.1016/j.ajo.2020.07.019