介质混浊对OCTA中心凹无血管区自动与手动测量的不同影响

Different effect of media opacity on automated and manual measurement of foveal avascular zone of optical coherence tomography angiographies.
PubMed 文献 · Br J Ophthalmol · 2020-07-23

研究用中性密度滤光片在30只健康眼模拟介质混浊,评估其对Cirrus 5000与Triton OCTA测量中心凹无血管区(FAZ)的影响。结果显示介质混浊加重时自动FAZ面积减小、手动FAZ面积增大,提示两种测量方式受介质混浊影响方向相反。

要点

  • 介质混浊加重时自动FAZ面积减小,手动FAZ面积增大。
  • 信号强度与自动FAZ面积正相关,与手动FAZ面积负相关。
  • 解读介质混浊眼的FAZ测量结果需谨慎。

核心结论

  • 介质混浊对FAZ定量测量的影响在自动与手动测量之间不同。
  • 介质混浊眼解读FAZ测量时须谨慎。

研究设计与样本

  • 研究纳入30名健康正常受试者的30只眼,使用Cirrus 5000 OCTA和Triton OCTA以3×3 mm扫描协议成像FAZ。
  • 用中性密度滤光片模拟介质混浊,Cirrus 5000的OD为0.10-0.48,Triton的OD为0.15-0.51。
  • FAZ面积、周长和圆度分别用内置软件和定制MATLAB程序自动测量,并另进行手动测量。
  • 采用Spearman相关分析OD、SS/SSI与FAZ指标的相关性。

主要发现

  • OD增加与SS/SSI降低显著相关(rs=-0.602和-0.925,均p<0.001)。
  • OD增加与自动FAZ面积减小显著相关(rs=-0.344和-0.766,均p<0.001)。
  • OD增加与手动FAZ面积增大显著相关(rs=0.423和0.543,均p<0.001)。
  • FAZ周长和圆度也发现类似结果。
  • SS/SSI与自动FAZ面积正相关,与手动FAZ面积负相关。

临床怎么用

  • 对存在介质混浊的眼(如白内障),解读OCTA的FAZ测量结果时应考虑测量方式的影响。
  • 自动与手动FAZ测量在介质混浊条件下可能给出方向相反的结果,临床需谨慎判断。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Optical coherence tomography angiography (OCTA) provides not only visualisation but also quantitative measurement of foveal avascular zone (FAZ). Media opacity is common in elderly subjects with cataracts. This study aimed to investigate the impact of media opacity on automated and manual FAZ measurement. METHODS: Cirrus 5000 OCTA and Triton OCTA were used to image FAZ using a 3×3 mm scanning protocol from 30 eyes of 30 healthy normal subjects. Media opacity was simulated with neutral-density filters (optical density (OD): 0.10-0.48 in Cirrus 5000 and 0.15-0.51 in Triton). Signal strength (SS) and signal strength intensity (SSI) were provided by the built-in software in Cirrus 5000 and Triton, respectively. FAZ area, perimeter and circularity were measured automatically using the built-in software as well as a customised MATLAB program. FAZ metrics were also measured manually. The correlations between the OD, SS/SSI and FAZ metrics were analysed using Spearman correlation. RESULTS: Increased OD significantly correlated with decreased SS/SSI (rs =-0.602 and -0.925, respectively, both p<0.001), decreased automated FAZ area (rs =-0.344 and -0.766, respectively, both p<0.001), but increased manual FAZ area in both Cirrus 5000 and Triton (rs =0.423 and 0.543, respectively, both p<0.001). Similar results were found for FAZ perimeter and circularity. There was a positive correlation between SS/SSI with the automated FAZ area but negative correlation with the manual FAZ area. CONCLUSIONS: The effect of media opacity on quantitative measurement of FAZ is different between automated and manual measurements. Cautions must be taken when interpreting FAZ measurement in eyes with media opacity.