青光眼患者眼内压遥测传感器的短期与长期变异性
结果显示日内7个时段的短期组内相关系数为0.52至0.66,长期组内相关系数为0.29至0.51。结论认为眼内压短期变异性中等、长期变异性高,单次测量不能反映逐日变化。
要点
- 眼内压短期变异性中等,长期变异性高。
- 单次眼内压测量不能反映逐日变化。
- 持续眼内压监测对青光眼患者有重要价值。
核心结论
- 持续眼内压监测显示青光眼患者眼内压短期变异性中等、长期变异性高。
- 单次眼内压测量不能反映逐日变化,持续监测在青光眼患者中具有重要性。
研究设计与样本
- 要求患者每日至少获取4次眼内压测量,按眼与用药分组,将一天分为7个时段。
- 短期变异性定义为3个月内不同日期同一时段测量的变异性,长期变异性定义为1年或以上不同日期同一时段测量的变异性,采用组内相关系数评估。
主要发现
- 每只眼测量次数范围为每日1次至每日277次。
- 日内7个时段短期组内相关系数范围为0.52(上午)至0.66(清晨)。
- 长期组内相关系数范围为0.29(夜间)至0.51(深夜)。
临床怎么用
- 临床应认识到单次眼内压测量不足以刻画逐日眼内压变化。
- 对青光眼患者可考虑持续眼内压监测以评估眼内压变异性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To evaluate the short-term and long-term variability of intraocular pressure (IOP) in eyes with primary open-angle glaucoma. DESIGN: Prospective study. PARTICIPANTS: Twenty-two patients previously implanted with a sulcus-based IOP sensor (EyeMate, Implandata GmbH, Germany). METHODS: Twenty-two patients previously implanted with the EyeMate were requested to obtain at least 4 IOP measurements daily. Data were grouped according to the eye and the medication so that an eye treated with a particular medication was considered as one group, and the same eye treated with a different medication during the observation period was considered as a different group. A day was divided into 7 periods: night, midnight to 5:59 am; early, 6 am to 7:59 am; morning, 8 am to 10:59 am; noon, 11 am to 1:59 pm; afternoon, 2 pm to 5:59 pm; evening, 6 pm to 8:59 pm; and late, 9 pm to 11:59 pm. Short-term variability during a particular period was defined as the variability in IOP measurements obtained during that period on different days within 3 months of each other. Long-term variability was defined as the variability in IOP measurements obtained during a particular period on different days over a period of 1 year or more. Variability was assessed using intraclass correlation coefficients (ICCs). RESULTS: The mean age of study participants was 67.8 ± 6.8 years and 36.4% were women. The mean follow-up duration of patients was 19.2 ± 21.3 months (median, 9 months; range, 1-58 months). Overall, 92 860 IOP measurements over 15 811 measurement days were obtained and analyzed during the study period. The number of measurements obtained from each eye ranged from 1 daily to 277 daily. Intraclass correlation coefficients for short-term variability among the 7 periods during the day ranged from 0.52 (morning) to 0.66 (early). Long-term ICCs ranged from 0.29 (night) to 0.51 (late). CONCLUSIONS: Continual IOP monitoring showed that IOP has moderate short-term and high long-term variability in glaucoma patients. These findings demonstrate that single IOP measurements do not characterize day-to-day variations in IOP. Moreover, they show the importance of continual IOP monitoring in glaucoma patients.
引用
- Mansouri K, Rao HL, Weinreb RN, et al. Short-Term and Long-Term Variability of Intraocular Pressure Measured with an Intraocular Telemetry Sensor in Patients with Glaucoma. Ophthalmology. 2020. PMID: 32663530. doi:10.1016/j.ophtha.2020.07.016
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。