评估Goldmann压平眼压计测量中医师与技术员眼压测量的一致性
该前瞻性研究在青光眼门诊比较6名技术员与2名医师使用Goldmann压平眼压计测量眼压的一致性,并评估教育干预对短期和长期不一致的影响。共评估529只眼,基线时医师间以>2 mmHg定义的不一致率为17%,技术员与医师平均为25%;教育干预后即刻和6个月均无显著改善。结论提示Goldmann眼压测量存在重要局限性。
要点
- 医师间以>2 mmHg为界的不一致率为17%,技术员与医师平均为25%。
- 眼压>20 mmHg时不一致更明显,教育干预未改善一致性。
- Goldmann眼压测量的一致性局限需在临床解读中考虑。
核心结论
- 两名医师在同一房间测量同一患者眼压时,以>2 mmHg为界的不一致率为17%,医师与技术员间不一致更高。
- 教育干预未改善技术员与医师间眼压测量的一致性,凸显Goldmann眼压测量的重要局限性。
研究设计与样本
- 前瞻性研究,旨在提高测量可靠性,设置于大学医院青光眼门诊。
- 研究对象为6名技术员和2名医师,共评估529只眼。
主要发现
- 医师测量平均眼压为16.4 mmHg(SD=5.9)。
- 基线时医师间以>2 mmHg和>3 mmHg定义的不一致率分别为17%和7%,技术员与医师平均分别为25%和13%。
- 眼压>20 mmHg时不一致更大,干预后即刻和6个月不一致频率无显著变化。
关键数值与推荐
- 不一致定义为眼压差>2 mmHg或>3 mmHg,在基线、干预后即刻和6个月评估。
- 每个技术员-医师配对在每个数据收集期评估30只眼。
临床怎么用
- 临床解读Goldmann眼压测量结果时应考虑医师间及医师与技术员间存在的不一致性。
- 教育干预未必能改善眼压测量一致性,需关注测量标准化和重复测量。
局限性
- 研究仅纳入6名技术员和2名医师,样本量有限。
- 研究设置于单一大学医院青光眼门诊,结果外推需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To examine IOP measurement disagreement between technicians and physicians and the impact of an educational intervention on the short and long-term disagreement in IOP measurement using Goldmann applanation tonometry. DESIGN: Prospective study designed to enhance measurement reliability. SETTING: A glaucoma clinic at a university hospital. StudyPopulation: 6 technicians and 2 physicians. INTERVENTION: An educational intervention was implemented for the technicians to improve IOP measurement agreement with physicians. MainOutcomeMeasures: Frequency of IOP measurement disagreement between physicians and technicians, defined as a difference in IOP of >2 or >3 mm Hg and assessed at baseline and immediately and 6 months postintervention. RESULTS: IOP was evaluated for a total of 529 eyes (physician measured mean IOP = 16.4 mm Hg [SD = 5.9]), 30 per technician-physician pair for each data collection period: baseline, immediately postintervention and 6 months postintervention. At baseline, physicians disagreed 17% and 7% of the time when measuring IOP using >2 and >3 mm Hg to define disagreement, respectively, whereas the average disagreement between technicians and physicians was 25% and 13%. Disagreement was greater at IOPs greater than 20 mm Hg. No significant changes were noted in the frequency of disagreement between technicians and physicians immediately or 6 months postintervention. CONCLUSIONS: Two physicians measuring the same patient in the same room disagreed by >2 mm Hg in 17% of patients' eyes, and this amount of disagreement was even higher when comparing physicians to certified technicians. An educational intervention did not improve agreement in IOP measurements between technicians and physicians. This highlights an important limitation of Goldmann tonometry.
引用
- Mihailovic A, Varadaraj V, Ramulu PY, et al. Evaluating Goldmann Applanation Tonometry Intraocular Pressure Measurement Agreement Between Ophthalmic Technicians and Physicians. Am J Ophthalmol. 2020. PMID: 32640253. doi:10.1016/j.ajo.2020.06.041