文献速递角膜与眼表疾病

准静态光学相干弹性成像表征人角膜生物力学特性

Quasi-Static Optical Coherence Elastography to Characterize Human Corneal Biomechanical Properties.
PubMed 文献 · Invest Ophthalmol Vis Sci · 2020-01-01

该研究应用准静态光学相干弹性成像(OCE)评估人供体角膜的生物力学特性,并探讨Bowman层在角膜生物力学中的作用。23枚角膜按钮在10至30 mmHg压力调制下进行光谱域OCT扫描,结果显示角膜应变分布较既往假设更复杂,前基质与后基质机械响应存在根本差异。

要点

  • 角膜组织在压力升高时向上移动(486-585 nm/mm Hg),压力降低后未完全恢复(Δ=2.63至8.64 µm)。
  • 垂直角膜应变分布前部为负、后部为正,表明同时存在角膜压缩和扩张。
  • Bowman层仅引起角膜应变分布的轻微局部差异。

核心结论

  • 角膜应变分布比既往假设更复杂,前基质与后基质的机械响应存在根本差异。

研究设计与样本

  • 研究纳入23枚人角膜供体按钮,其中11枚通过20 µm基质准分子激光消融Bowman层。

主要发现

  • 压力升高过程中角膜组织向上移动(486-585 nm/mm Hg),压力降低后未完全恢复(Δ=2.63至8.64 µm)。
  • 垂直角膜应变分布前部为负、后部为正,表明同时存在角膜压缩和扩张。
  • Bowman层引起角膜应变分布的轻微局部差异。

临床怎么用

  • OCE技术或可用于监测角膜扩张性疾病的进展及判断角膜交联的成功与否。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Quasi-static optical coherence elastography (OCE) is an emerging technology to investigate corneal biomechanical behavior in situations similar to physiological stress conditions. Herein OCE was applied to evaluate previously inaccessible biomechanical characteristics of human corneal tissue and to study the role of Bowman's layer in corneal biomechanics. METHODS: Human corneal donor buttons (n = 23) were obtained and Descemet's membrane and endothelium were removed. In 11 corneas, Bowman's layer was ablated by a 20 µm stromal excimer laser ablation. Buttons were mounted on an artificial anterior chamber and subjected to a pressure modulation from 10 to 30 mm Hg, and back to 10 mm Hg, in steps of 1 mm Hg. At each step, a spectral-domain optical coherence tomography scan was obtained. Displacements were analyzed by optical flow tracking, and strain over the entire stromal depth was retrieved from the phase gradient of the complex interference signal. RESULTS: During pressure increase, corneal tissue moved upward (486-585 nm/mm Hg) but did not fully recover (Δ= 2.63 to 8.64 µm) after pressure decrease. Vertical corneal strain distribution was negative in the anterior and positive in the posterior cornea, indicating simultaneous corneal compression and expansion, respectively. Bowman's layer caused minor localized differences in corneal strain distribution. CONCLUSIONS: Corneal strain distribution is more complex than previously assumed, with a fundamental difference in mechanical response between the anterior and posterior stroma. Clinically, OCE technology might be used to monitor the progression of corneal ectatic diseases and to determine the success of corneal cross-linking.

来源:Invest Ophthalmol Vis Sci,查看原文。版权归原作者所有。

引用

  1. Kling S, Torres-Netto EA, Spiru B, et al. Quasi-Static Optical Coherence Elastography to Characterize Human Corneal Biomechanical Properties. Invest Ophthalmol Vis Sci. 2020. PMID: 32539132. doi:10.1167/iovs.61.6.29

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。