角膜上皮厚度作为角膜缘干细胞缺乏症诊断参数的AS-OCT研究
该前瞻性横断面研究用前段OCT测量角膜中央上皮厚度(CET)和角膜缘最大上皮厚度(mLET),比较角膜缘干细胞缺乏症(LSCD)眼与正常眼。发现LSCD组CET和mLET均变薄,三点法OCT-CET3与活体共聚焦显微镜CET相关性好(r=0.91),诊断LSCD的cutoff值为46.6 μm,且其敏感性和特异性最高。
要点
- LSCD组OCT-CET3为41.6±10.8 μm,低于对照组55.0±3.0 μm(P<.001)。
- OCT-CET3与IVCM-CET相关性(r=0.91)优于OCT-CET1(r=0.87,P=.001)。
- OCT-CET3诊断LSCD的cutoff值为46.6 μm,敏感性及特异性最高。
核心结论
- LSCD患者中央角膜上皮和角膜缘上皮均较正常人变薄。
- OCT-CET3是临床怀疑LSCD时确认诊断的可靠参数。
研究设计与样本
- 用AS-OCT测量中央角膜上皮厚度(OCT-CET1和OCT-CET3)及四个象限角膜缘最大上皮厚度(mLET),并与IVCM-CET比较。
主要发现
- OCT-CET3在对照组为55.0±3.0 μm(范围50.6-62.0 μm),LSCD组为41.6±10.8 μm(范围0-56.3 μm),P<.001。
- OCT-CET3与IVCM-CET相关性r=0.91,OCT-CET1为r=0.87(P=.001)。
- OCT-CET3减少程度随LSCD临床分期进展而增加(所有P<.001)。
- 与对照组相比,LSCD组四个角膜缘区域mLET均降低(所有P<.001)。
关键数值与推荐
- OCT-CET3提示LSCD的cutoff值为46.6 μm。
- OCT-CET3在检测LSCD时的敏感性和特异性在所有mLET参数中最高。
临床怎么用
- 临床怀疑LSCD时可用AS-OCT测量OCT-CET3辅助确诊。
- OCT-CET3与IVCM-CET相关性好,可作为无创评估角膜上皮厚度的可靠指标。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Using anterior segment optical coherence tomography (AS-OCT), we investigated the epithelial thickness (ET) of the central cornea and limbal regions in patients with limbal stem cell deficiency (LSCD) as a diagnostic and staging parameter. DESIGN: Prospective, cross-sectional study. METHODS: The central corneal epithelium thickness (CET) and maximum limbal epithelium thickness (mLET) were measured in the superior, inferior, nasal, and temporal limbus on AS-OCT images of the normal and eyes with LSCD. CET was obtained by 1-point (OCT-CET1) and 3-point measurement (OCT-CET3). The values of OCT-CET1 and OCT-CET3 were compared to the CET obtained with in vivo confocal microscopy (IVCM-CET). RESULTS: Sixty-eight eyes of 50 patients with LSCD and 52 eyes of 34 normal subjects were included. The mean (±standard deviation) OCT-CET3 was 55.0 ± 3.0 μm (range, 50.6-62.0 μm) in the control group and 41.6 ± 10.8 μm (range, 0-56.3 μm) in the LSCD group (P < .001). OCT-CET3 had a better correlation with IVCM-CET (r = 0.91) than did OCT-CET1 (r = 0.87, P = .001). The degree of reduction in OCT-CET3 increased in more advanced clinical stages of LSCD (all P < .001). The OCT-CET3 cutoff value that suggests LSCD was 46.6 μm. Compared with the control group, the LSCD group had decreases in mLET in all 4 limbal regions (all P < .001). The sensitivity and specificity of OCT-CET3 is the highest among all mLET in detecting LSCD. CONCLUSIONS: Both CET and mLET were thinner in patients with LSCD than in normal subjects. OCT-CET3 appears to be a reliable parameter to confirm LSCD when there is clinical suspicion.
引用
- Liang Q, Le Q, Cordova DW, et al. Corneal Epithelial Thickness Measured Using Anterior Segment Optical Coherence Tomography as a Diagnostic Parameter for Limbal Stem Cell Deficiency. Am J Ophthalmol. 2020. PMID: 32283095. doi:10.1016/j.ajo.2020.04.006
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。