两种疏水性丙烯酸人工晶状体后囊膜混浊的3年随机试验结果
术后3年Vivinex XY1眼客观PCO评分显著低于AcrySof SN60WF眼(0.9±0.8对1.4±1.1,P<.001),Nd:YAG囊膜切开率分别为11.4%与18.6%(P=.23)。
要点
- Vivinex XY1的3年客观PCO评分显著低于AcrySof SN60WF。
- 3年Nd:YAG囊膜切开率两组分别为11.4%与18.6%,差异无统计学意义。
- 疏水材料、光滑光学面与锐利后缘共同影响PCO发生。
核心结论
- 新型疏水性丙烯酸Vivinex XY1人工晶状体较AcrySof SN60WF显示出显著更低的PCO发生率与更低的YAG率。
- 疏水材料、光滑光学面与锐利后光学边缘等多种因素的相互作用在PCO发展中起关键作用。
研究设计与样本
- 随机、前瞻性、患者与检查者设盲的临床 trial,采用眼内对照设计。
- 一眼植入Vivinex XY1人工晶状体,另一眼植入AcrySof SN60WF人工晶状体,术后3年随访。
主要发现
- Vivinex XY1眼的平均客观PCO评分为0.9±0.8,AcrySof SN60WF眼为1.4±1.1(P<.001)。
- 术后3年,Vivinex XY1眼11.4%的患者接受Nd:YAG囊膜切开,AcrySof SN60WF眼为18.6%(P=.23)。
关键数值与推荐
- PCO评分采用0-10分制,由裂隙灯主观评估并使用自动图像分析软件(AQUA)客观评估。
临床怎么用
- 在选择人工晶状体时,可将3年PCO评分与YAG囊膜切开率数据作为参考,结合材料与设计特征进行个体化决策。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To compare the incidence and intensity of posterior capsule opacification (PCO) and neodymium-yttrium-aluminum-garnet (Nd:YAG) capsulotomy rates between 2 similar open-loop single-piece hydrophobic acrylic intraocular lenses (IOLs) that differ in the proprietary material characteristics and design features, over a period of 3 years. DESIGN: Randomized, prospective, patient- and examiner-masked clinical trial with intraindividual comparison. METHODS: Setting: Department of Ophthalmology, Medical University of Vienna, Vienna, Austria. PATIENT POPULATION: Eighty patients (160 eyes) had bilateral cataract surgery and received a Vivinex XY1 IOL in 1 eye and an AcrySof SN60WF IOL in the other eye. OBSERVATION PROCEDURES: Follow-up examinations were performed 3 years after surgery. Digital retroillumination images were taken of each eye. The amount of PCO (score: 0-10) was assessed subjectively at the slit lamp and objectively using automated image analysis software (AQUA). MAIN OUTCOME MEASURE: PCO score (scale, 0-10). RESULTS: The mean objective PCO score of the Vivinex XY1 IOLs was 0.9 ± 0.8 compared to the PCO score of 1.4 ± 1.1 for the AcrySof SN60WF IOLs (P < .001). Three years postoperatively, 11.4% of patients had an Nd:YAG capsulotomy in the Vivinex XY1 eye and 18.6% had a capsulotomy in the AcrySof SN60WF eye (P = .23). CONCLUSION: The new hydrophobic acrylic Vivinex XY1 IOL showed significantly lower PCO rates and lower YAG rates compared to the AcrySof SN60WF IOL. The interaction of various factors such as hydrophobic material, smooth optic surface, and sharp posterior optic edge plays a key role in PCO development.
引用
- Leydolt C, Schartmüller D, Schwarzenbacher L, et al. Posterior Capsule Opacification With Two Hydrophobic Acrylic Intraocular Lenses: 3-Year Results of a Randomized Trial. Am J Ophthalmol. 2020. PMID: 32335056. doi:10.1016/j.ajo.2020.04.011
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。