利用深度学习自动化Goldmann压平眼压计读数
这项前瞻性横断面研究开发了一种基于深度学习和固定力Goldmann压平眼压计(GAT)的自动化眼压测量方法。在测试集中,自动化与标准GAT的平均差异为-0.9 mmHg(95%一致性界限-5.4至3.6 mmHg),重复性系数分别为3.8和3.9 mmHg。初步结果表明自动化GAT与标准GAT结果相当,并可能减少偏差、提高重复性。
要点
- 自动化GAT与标准GAT的平均差异为-0.9 mmHg,一致性界限较宽。
- 自动化GAT的重复性系数为3.8 mmHg,与标准GAT的3.9 mmHg相近。
- 自动化GAT可能减少偶数次测量的偏差,并观察到眼脉冲振幅。
核心结论
- 使用深度学习自动化GAT的初步测量结果与标准GAT相当。
- 自动化GAT有潜力通过减少偏差和提高重复性来显著改进当前GAT测量标准。
研究设计与样本
- 前瞻性横断面研究,纳入学术青光眼诊所的患者。
主要发现
- 测试集中自动化与标准GAT的平均差异为-0.9 mmHg(95% LoA -5.4至3.6 mmHg)。
- 两名观察者使用标准GAT的平均差异为0.09 mmHg(LoA -3.8至4.0 mmHg)。
- 两名观察者使用自动化GAT视频的平均差异为-0.3 mmHg(LoA -4.1至3.5 mmHg)。
- 自动化和标准GAT的重复性系数分别为3.8和3.9 mmHg。
关键数值与推荐
- 自动化GAT与标准GAT的平均差异为-0.9 mmHg(95% LoA -5.4至3.6 mmHg)。
- 自动化GAT的重复性系数为3.8 mmHg,标准GAT为3.9 mmHg。
临床怎么用
- 自动化GAT可作为标准GAT的潜在替代方法,减少测量偏差并提高重复性。
- 该技术还可观察到眼脉冲振幅。
局限性
- 研究为初步测量,测试集样本量较小(50只眼)。
- 一致性界限较宽,自动化与标准GAT的个体差异可能较大。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To develop an objective and automated method for measuring intraocular pressure using deep learning and fixed-force Goldmann applanation tonometry (GAT) techniques. DESIGN: Prospective cross-sectional study. PARTICIPANTS: Patients from an academic glaucoma practice. METHODS: Intraocular pressure was estimated by analyzing videos recorded using a standard slit-lamp microscope and fixed-force GAT. Video frames were labeled to identify the outline of the reference tonometer and the applanation mires. A deep learning model was trained to localize and segment the tonometer and mires. Intraocular pressure values were calculated from the deep learning-predicted tonometer and mire diameters using the Imbert-Fick formula. A separate test set was collected prospectively in which standard and automated GAT measurements were collected in random order by 2 independent masked observers to assess the deep learning model as well as interobserver variability. MAIN OUTCOME MEASURES: Intraocular pressure measurements between standard and automated methods were compared. RESULTS: Two hundred sixty-three eyes of 135 patients were included in the training and validation videos. For the test set, 50 eyes from 25 participants were included. Each eye was measured by 2 observers, resulting in 100 videos. Within the test set, the mean difference between automated and standard GAT results was -0.9 mmHg (95% limits of agreement [LoA], -5.4 to 3.6 mmHg). Mean difference between the 2 observers using standard GAT was 0.09 mmHg (LoA,-3.8 to 4.0 mmHg). Mean difference between the 2 observers using automated GAT videos was -0.3 mmHg (LoA, -4.1 to 3.5 mmHg). The coefficients of repeatability for automated and standard GAT were 3.8 and 3.9 mmHg, respectively. The bias for even-numbered measurements was reduced when using automated GAT. CONCLUSIONS: Preliminary measurements using deep learning to automate GAT demonstrate results comparable with those of standard GAT. Automated GAT has the potential to improve on our current GAT measurement standards significantly by reducing bias and improving repeatability. In addition, ocular pulse amplitudes could be observed using this technique.