多中心研究:Pentacam HR在圆锥角膜测量中的可靠性与重复性
该多中心研究纳入158例圆锥角膜患者的230只眼,每只眼连续进行5次Pentacam HR角膜断层扫描,按最大后表面高度分为3级,计算组内相关系数(ICC)和重复性限。结果显示所有参数ICC均达0.9,关键参数重复性限较高,且随圆锥角膜严重程度增加而增大(除彗差轴外)。
要点
- Pentacam HR所有断层及像差参数ICC均达0.9,可靠性高。
- 关键参数重复性限:屈光力>0.5D,散光轴>20°,最薄角膜厚度>10µm。
- 重复性限随圆锥角膜分级加重而增大,彗差轴除外。
核心结论
- ICC提示Pentacam HR可用于圆锥角膜诊断,但重复性限提示需建立考虑实际测量误差的新进展监测标准。
研究设计与样本
- 多中心研究,纳入158例圆锥角膜患者的230只眼,每只眼连续5次Pentacam HR角膜断层扫描,按最大后表面高度分为3级。
主要发现
- 所有参数的ICC均为0.9。
- 屈光力参数重复性限>0.5D,角膜散光轴>20°,最薄角膜厚度>10µm。
- 像差值重复性限>0.1µm,彗差轴总体>15°。
- 所有重复性限随圆锥角膜严重程度增加而增大,但彗差轴随分级增加而下降。
关键数值与推荐
- 分级标准:max_BFS_post为40µm为1级,41-75µm为2级,>75µm为3级。
临床怎么用
- 临床可依据ICC结果将Pentacam HR用于圆锥角膜诊断。
- 监测进展时应考虑各参数重复性限,避免将测量误差误判为病情变化。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To assess the repeatability and reliability of the most important tomographic parameters for characterising keratoconus measured with a Pentacam HR (high resolution). METHODS: Overall, 230 eyes in 158 patients with keratoconus were analysed. We performed five consecutive corneal tomography examinations for each eye with a Pentacam HR in patients with keratoconus. Study eyes were classified into three groups depending on the maximum posterior elevation (max_BFS_post): grade 1 for cases of keratoconus with a max_BFS_post of 40 µm; grade 2 for those with a max_BFS_post of between 41 and 75 µm and grade 3 for those with a max_BFS_post of over 75 µm. We calculated the intraclass correlation coefficients (ICCs) and repeatability limits of parameters from tomography and aberrometry. RESULTS: All the parameters were found to have excellent ICCs (0.9). The repeatability limits for the key parameters were higher than 0.5D for the power parameters, 20° for the axis of corneal astigmatism and 10 µm for the thinnest corneal thickness. Further, we obtained repeatability limits of above 0.1 µm for the aberrometry values and overall greater than 15° for the coma axis. All the values increase with the severity of keratoconus, except for that of the coma axis which falls with keratoconus grade. CONCLUSIONS: The reliability indicated by ICCs supports the view that the Pentacam HR is useful for the diagnosis of keratoconus. The repeatability limits suggest that new criteria should be established for monitoring progression taking into account the real measurements that can be made using this system.
引用
- de Luis Eguileor B, Arriola-Villalobos P, Pijoan Zubizarreta JI, et al. Multicentre study: reliability and repeatability of Scheimpflug system measurement in keratoconus. Br J Ophthalmol. 2020. PMID: 32217537. doi:10.1136/bjophthalmol-2019-314954