前节扫频OCT子午线扫描用于房角关闭评估
研究在2027例社区晶状体眼受试者中比较前节扫频OCT各子午线扫描诊断房角镜房角关闭的性能;80°-260°单扫描AUC最高(0.82,95%CI 0.79-0.84),优于多数颞鼻侧扫描;连续累加至上下方近80°-85°扫描时AUC峰值0.80,继续加颞鼻侧扫描无进一步增益。
要点
- 80°-260°单扫描诊断房角镜房角关闭的AUC最高,为0.82(95%CI 0.79-0.84)。
- 上下方扫描的AUC高于颞鼻侧扫描。
- 连续累加扫描的AUC峰值0.80,继续加颞鼻侧扫描无进一步累积增益。
核心结论
- 80°-260°单次SS-OCT扫描对房角镜房角关闭的诊断性能最高,360°全面评估未必带来更优临床效用。
研究设计与样本
- 观察性研究,从社区综合诊所连续招募2027例晶状体眼受试者,行房角镜和SS-OCT检查。
主要发现
- 各单扫描AUC范围为0.73至0.82。
- 80°-260°单扫描AUC最高(0.82,95%CI 0.79-0.84),显著优于多数颞鼻侧扫描(0°-52°和153°-179°)。
- 上下方扫描的AUC高于颞鼻侧扫描。
- 连续累加单扫描时,纳入接近80°-85°的上下方扫描获得峰值AUC 0.80,继续加其余颞鼻侧扫描无进一步正向累积效应。
关键数值与推荐
- 房角镜房角关闭定义为≥180°房角范围内后部小梁网不可见。
- SS-OCT房角关闭定义为巩膜突前方的虹膜小梁网接触。
临床怎么用
- 临床可优先采用80°-260°子午线SS-OCT扫描筛查房角镜房角关闭,简化分析流程。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND/AIMS: As swept-source optical coherence tomography (SS-OCT) simultaneously obtains 128 meridional scans, it is important to identify which scans are playing the main role in classifying gonioscopic angle closure to simplify the analysis. We aimed to evaluate the diagnostic performance of every meridional scan in its ability to detect gonioscopic angle closure. METHODS: Observational study with 2027 phakic subjects consecutively recruited from a community polyclinic. Gonioscopy and SS-OCT were performed. Gonioscopic angle closure was defined as non-visibility of the posterior trabecular meshwork in ≥180° of the angle, while SS-OCT was defined as iridotrabecular contact anterior to the scleral spur. The area under the receiver operating characteristic curve (AUC) was calculated to assess the diagnostic performance of each single scan, the sequential anticlockwise cumulative effect of those single scans and different combinations of them. RESULTS: The AUCs of each scan ranged from 0.73 to 0.82. The single scan at 80°-260° had the highest AUC (0.82, 95% CI 0.79 to 0.84) and performed significantly better than most of the temporonasal scans (from 0° to 52° and from 153° to 179°). The superoinferior scans achieved higher AUCs compared with the temporonasal ones. When assessing the cumulative effect of adding individual scans consecutively, the peak AUC (0.80) was obtained when considering the superoinferior scans closer to 80°-85°, but no further positive cumulative effect was seen when adding the rest of the temporonasal scans of the circumference. CONCLUSIONS: In conclusion, the single SS-OCT scan at 80°-260° had the highest diagnostic performance. Our study suggests that the 360° evaluation may not translate to better clinical utility for detection of gonioscopic angle closure.
引用
- Porporato N, Baskaran M, Perera S, et al. Evaluation of meridional scans for angle closure assessment with anterior segment swept-source optical coherence tomography. Br J Ophthalmol. 2020. PMID: 32152140. doi:10.1136/bjophthalmol-2019-315461
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。