新型接触镜传感器首次实现24小时眼压与眼脉动连续测量
这项前瞻性、开放标签、单中心、非随机研究纳入青光眼与健康受试者,评估压力测量接触镜(PMCL)设备连续24小时测量眼压(IOP)和眼脉动振幅(OPA)的可行性。八名受试者获得24小时IOP与OPA曲线,PMCL与标准眼压计在同一眼测量值差异多在±5 mmHg内,饮水试验中PMCL与动态轮廓眼压计(DCT)均检测到IOP升高。结论认为该设备无创且与标准眼压测量具有良好可比性,为24小时连续监测提供概念验证。
要点
- PMCL可连续24小时获取IOP与OPA曲线,共8名受试者完成。
- 同一眼PMCL与Goldmann压平眼压计(GAT)差值在±5 mmHg内者占75%。
- 同一眼PMCL与DCT差值在±5 mmHg内者占87.5%。
核心结论
- 首次人体研究证明PMCL用于24小时连续测量IOP和OPA的概念可行,设备无创且与标准眼压计可比。
研究设计与样本
- 前瞻性、开放标签、单中心、非随机研究,纳入青光眼与健康受试者,共8名受试者获得24小时IOP和OPA曲线。
主要发现
- 同一眼PMCL与GAT的IOP平均差值在±5 mmHg内者占75%,与DCT在±5 mmHg内者占87.5%。
- 饮水试验中PMCL与DCT均检测到IOP变化,平均升高分别为2.43 mmHg和1.85 mmHg。
- 对侧眼PMCL与DCT的IOP变化差值在±5 mmHg内的时间点占97.2%,OPA差值在±5 mmHg内的时间点占85.5%。
临床怎么用
- PMCL有望用于24小时连续IOP与OPA监测,但当前仅为概念验证,尚不能替代标准眼压测量。
局限性
- 样本量小(8名受试者),为非随机单中心研究,结果需更大样本验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND/AIMS: This study assessed the feasibility of a novel contact lens device for intraocular pressure (IOP) and ocular pulse amplitude (OPA) continuous measurements over 24 hours. METHODS: This prospective, open-label, single-centre, non-randomised study included glaucoma and healthy subjects. IOP and OPA values acquired by the pressure-measuring contact lens (PMCL) device in one patient's eye at the beginning of the measurement were compared with tonometry values (Goldman applanation tonometry (GAT) and dynamic contour tonometry (DCT)) in the same eye just before PMCL placement. Furthermore, IOP and OPA values measured with PMCL on the study eye during a water drinking test (WDT) were compared with DCT values in the fellow eye. Comparisons were performed using t-tests with 95% Confidence Intervals. RESULTS: Twenty-four-hour IOP and OPA curves were obtained for eight subjects. The mean IOP difference between PMCL and tonometry on the same eye was within ±5 mm Hg in 75% (GAT) and 87.5% (DCT) of subjects. IOP variations due to WDT were detected by PMCL and DCT, showing an average increase of 2.43 and 1.85 mm Hg, respectively. Differences between PMCL and DCT for IOP variations in fellow eyes were within ±5 mm Hg for 97.2% of time points. The difference between OPA in fellow eyes was within ±5 mm Hg for 85.5% of the time points. CONCLUSIONS: This first-in-human study is a proof-of-concept for 24-hour continuous measurements of IOP and OPA with the PMCL. This device is non-invasive and has good comparability with standard tonometry.
引用
- Wasilewicz R, Varidel T, Simon-Zoula S, et al. First-in-human continuous 24-hour measurement of intraocular pressure and ocular pulsation using a novel contact lens sensor. Br J Ophthalmol. 2020. PMID: 32075818. doi:10.1136/bjophthalmol-2019-315276
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。