新型囊袋固定式人工晶状体的有效晶状体位置与屈光结果比较
囊袋固定式IOL的ELP最短(4.29±0.24 mm),与板式襻(P=.001)和C襻式(P=.000)差异显著;基于晶状体设计的前房深度调整对所有公式和晶状体的屈光及IOL度数预测均有显著影响(P<.05)。
要点
- 囊袋固定式IOL的ELP为4.29±0.24 mm,短于C襻式(4.41±0.42 mm)和板式襻(4.51±0.26 mm)。
- 囊袋固定式IOL与板式襻(P=.001)及C襻式(P=.000)的ELP差异均显著。
- 基于晶状体设计的前房深度调整对所有公式和晶状体的屈光及IOL度数预测有显著影响(P<.05)。
核心结论
- IOL固定方式及其在囊袋内最终位置对有效晶状体位置有显著影响,进而显著影响术后屈光预测。
研究设计与样本
- 回顾性、多中心、比较病例系列,纳入年龄相关性白内障手术并植入板式襻、C襻式或新型囊袋固定式IOL的患者,术前及术后3个月行生物测量和屈光检查,并进行晶状体常数优化。
主要发现
- 囊袋固定式IOL的ELP最短(4.29±0.24 mm),其次为C襻式(4.41±0.42 mm)和板式襻(4.51±0.26 mm)。
- 囊袋固定式IOL与板式襻(P=.001)和C襻式IOL(P=.000)的ELP差异显著。
- 基于晶状体设计的前房深度调整对所有公式和晶状体的屈光及IOL度数预测均有显著影响(P<.05)。
- 囊袋固定式IOL的屈光差异从Hill-Radial Basis Function的-0.039 D到Haigis的-0.096 D;其他两种晶状体平均屈光差异在Hill-Radial Basis Function的+0.046 D到Haigis的+0.097 D之间。
临床怎么用
- 临床进行IOL度数计算和术后屈光预测时,应考虑不同晶状体设计对有效晶状体位置的影响,并可能需基于晶状体设计调整前房深度参数。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: WE sought to evaluate differences in effective lens position (ELP) based on the lens design. Intraocular lenses (IOLs) with plate-haptic, c-loop haptic, and a rhexis-fixated lens were compared. DESIGN: Retrospective, multicenter, comparative case series. METHODS: The study included patients having age-related cataract surgery with implantation of either a plate-haptic, c-loop haptic, or a novel rhexis-fixated IOL. Biometry and refraction measurements were conducted preoperatively and 3 months postoperatively. Lens constant optimization was performed. RESULTS: Seventy eyes of 56 subjects were included. ELP for rhexis-fixated IOL was shortest (4.29 ± 0.24 mm), followed by c-loop haptic (4.41 ± 0.42 mm) and plate-haptic (4.51 ± 0.26 mm) IOL. Difference in ELP was significant between rhexis-fixated IOL and both plate-haptic (P = .001) and c-loop haptic IOL (P = .000). Anterior chamber depth adjustment based on lens design showed a significant effect on refraction and IOL power predictions for all formulas and lenses (P < .05). For the rhexis-fixated IOL the differences in refraction ranged from -0.039 diopters (D) for the Hill-Radial Basis Function to -0.096 D for Haigis. The other 2 lenses showed mean differences in refraction between +0.046 D for Hill-Radial Basis Function and +0.097 D for Haigis. CONCLUSION: The difference in IOL fixation and its resulting position in the capsular bag have a significant effect on the effective lens position and consequently a significant effect on the prediction of postoperative refraction.
引用
- Shajari M, Sonntag R, Niermann T, et al. Determining and Comparing the Effective Lens Position and Refractive Outcome of a Novel Rhexis-Fixated Lens to Established Lens Designs. Am J Ophthalmol. 2020. PMID: 31953058. doi:10.1016/j.ajo.2020.01.009