文献速递青光眼

现代影像设备在青光眼筛查中的诊断效能

Diagnostic performance of modern imaging instruments in glaucoma screening.
PubMed 文献 · Br J Ophthalmol · 2020-01-16

基于芬兰北部出生队列眼研究,对45至49岁未选择人群3039人6060眼行SD-OCT、GDx和HRT影像检查,以临床青光眼诊断为参考。各设备敏感度31%至61%,特异度88%至96%,AUC 0.73至0.76,阳性后验概率5%至11%,筛查准确性仅中等,单独用于筛查不可靠。

要点

  • OCT联合分析敏感度61%、特异度90%,为各方法中最高敏感度。
  • HRT敏感度31%、特异度96%,敏感度最低但特异度最高。
  • 各方法AUC 0.73至0.76,阳性后验概率仅5%至11%。

核心结论

  • OCT、GDx和HRT的筛查能力相当,所有评估参数准确性仅中等,单独用于青光眼筛查不可靠。

研究设计与样本

  • 基于人群的芬兰北部出生队列眼研究,纳入45至49岁未选择人群,共检查3039人6060眼。
  • 采用SD-OCT(RNFL、GCIPL及联合分析)、GDx和HRT3五种影像方法,以基于视盘和RNFL照片及视野的「2/3法则」临床青光眼诊断为参考标准。

主要发现

  • 临床评估发现33人(1.1%)43眼存在青光眼性损害。
  • RNFL分析敏感度53%、特异度95%,AUC 0.76,阳性后验概率11%。
  • GCIPL分析敏感度50%、特异度92%,AUC 0.73,阳性后验概率7%。
  • OCT联合分析敏感度61%、特异度90%,AUC 0.75,阳性后验概率6%。
  • GDx敏感度56%、特异度88%,AUC 0.75,阳性后验概率5%;HRT敏感度31%、特异度96%,AUC 0.73,阳性后验概率7%。

关键数值与推荐

  • 临床青光眼诊断采用「2/3法则」,即基于视盘和RNFL照片及视野三项评估中两项异常。
  • 各影像方法敏感度31%至61%,特异度88%至96%,AUC 0.73至0.76,阳性后验概率5%至11%。

临床怎么用

  • 在中年未选择人群中,OCT、GDx和HRT自动化分类的筛查准确性仅中等,不宜单独作为青光眼筛查工具。

局限性

  • 原文未报告局限性相关内容。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIM: To evaluate the applicability of imaging devices (spectral-domain optical coherence tomography (Cirrus SD-OCT), scanning laser polarimetry (GDx) and scanning laser ophthalmoscopy (Heidelberg Retinal Tomograph, HRT3)) for glaucoma screening in a middle-aged unselected population. METHODS: Participants of the population-based Northern Finland Birth Cohort Eye Study, aged 45 to 49 years, underwent a comprehensive eye examination including modern imaging with five methods (retinal nerve fibre layer (RNFL) and macular ganglion cell layer +inner plexiform layer (GCIPL) analysis and their combination with SD-OCT, GDx and HRT). The performance of the automated classification of the imaging devices was assessed using a clinical glaucoma diagnosis as reference, that is, the '2 out of 3' rule based on the evaluation of optic nerve head and RNFL photographs and visual fields. RESULTS: We examined 6060 eyes of 3039 subjects; in the clinical evaluation, glaucomatous damage was found in 33 subjects (1.1%) in 43 eyes. The following sensitivities were obtained; RNFL analysis (53%), GCIPL analysis (50%), OCT combination analysis (61%), GDx (56%) and HRT (31%) with corresponding specificities of 95%, 92%, 90%, 88% and 96%. The area under the curve values were 0.76, 0.73, 0.75, 0.75 and 0.73, respectively. Post-test probabilities of glaucoma after positive imaging finding with each of these methods in this unselected population were 11%, 7%, 6%, 5% and 7%, respectively. CONCLUSION: Screening capabilities of the OCT, GDx and HRT were rather similar. The accuracy of all evaluated parameters was only moderate and thus screening with these parameters alone is not reliable.

来源:Br J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Karvonen E, Stoor K, Luodonpää M, et al. Diagnostic performance of modern imaging instruments in glaucoma screening. Br J Ophthalmol. 2020. PMID: 31949097. doi:10.1136/bjophthalmol-2019-314795

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。