文献速递小儿眼科与斜弱视

儿童斜视的自动化诊断与测量

Automated Diagnosis and Measurement of Strabismus in Children.
PubMed 文献 · Am J Ophthalmol · 2019-12-27

本研究前瞻性纳入69名3至15岁儿童,比较基于眼动追踪与全遮挡眼镜的自动化系统与两名独立盲法检查者手动遮盖-去遮盖试验及棱镜交替遮盖试验的结果。自动化系统与手动测量在水平和垂直偏斜上均高度相关(R=0.9和R=0.91,P<0.001),对偏斜类型的识别正确率为100%,平均测试时长46秒。结论认为该系统可精确检测和测量眼位偏斜,并能区分隐斜与显斜,适用于3岁及以上儿童。

要点

  • 自动化系统与手动测量的水平偏斜相关系数R=0.9(P<0.001),垂直偏斜R=0.91(P<0.001)。
  • 自动化系统对偏斜类型的识别正确率为100%,平均测试时长46秒。
  • Bland-Altman分析显示两种测量方法的平均差值为-2.9棱镜度(PD),95%一致性界限半宽为±11.4 PD。

核心结论

  • 自动化系统可精确检测和测量眼位偏斜,并能区分隐斜与显斜,适用于3岁及以上儿童。

研究设计与样本

  • 前瞻性、盲法、评分者间信度研究,纳入69名3至15岁儿童。
  • 由两名独立盲法检查者及自动化系统在50 cm距离、第一眼位分别进行遮盖-去遮盖试验和棱镜交替遮盖试验(PACT)。

主要发现

  • 自动化与手动测量结果高度相关:水平偏斜R=0.9(P<0.001),垂直偏斜R=0.91(P<0.001)。
  • 偏斜类型识别正确率为100%。
  • 自动化测试平均时长为46秒。
  • Bland-Altman图比较两种方法,平均值为-2.9棱镜度(PD),95%一致性界限半宽为±11.4 PD。

临床怎么用

  • 该自动化系统可作为儿童斜视手动测量的快速、客观且易用的替代方案,适用于3岁及以上儿童。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Manual measurements of strabismus are subjective, time consuming, difficult to perform in babies, toddlers, and young children, and rely on the examiner's skill and experience. An automated system, based on eye tracking and dedicated full occlusion glasses, was developed to provide a fast, objective, and easy-to-use alternative to the manual measurements of strabismus. This study tested the efficacy of the system in determining the presence of strabismus in children, as well as its type and the amount of deviation, in addition to differentiating between phorias and tropias. DESIGN: A prospective, masked, inter-rater reliability study. METHODS: A prospective, masked, cross-sectional study included 69 children, 3-15 years of age. A cover-uncover test and a prism alternating cover test (PACT) for the primary gaze, at a distance of 50 cm, were performed by 2 independent, masked examiners and by the automated system. RESULTS: A high correlation was found between the automated and the manual test results (R = 0.9 and P < 0.001 for the horizontal deviation, and R = 0.91 and P < 0.001 for the vertical deviations, with 100% correct identification of the type of deviation). The average automated test duration was 46 seconds. The Bland-Altman plot, used to compare the 2 measurement methods, showed a mean value of -2.9 prism diopters (PD) and a half-width of the 95% limit of agreement of ±11.4 PD. CONCLUSION: The automated system provides precise detection and measurements of ocular misalignment and differentiated between phorias and tropias. It can be used in participants from the ages of 3 years old and up.

来源:Am J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Yehezkel O, Belkin M, Wygnanski-Jaffe T, et al. Automated Diagnosis and Measurement of Strabismus in Children. Am J Ophthalmol. 2019. PMID: 31887281. doi:10.1016/j.ajo.2019.12.018