确定视盘周围OCT可靠性的循证标准
研究发现SS在10至3之间下降对可靠性影响较小,低于3时影响显著增大,且严重青光眼眼受影响更大;伪影独立于SS降低可靠性。作者据此提出循证指南以指导医生判断OCT扫描结果是否可信。
要点
- SS在10至3之间每降低1点,ΔRNFL为-0.67至-1.25,对OCT可靠性影响较小。
- SS低于3时每降低1点,ΔRNFL为-15.70至-16.34,对可靠性影响显著增大。
- 伪影独立于SS显著影响可靠性(ΔRNFL -4.76),且仅影响其所在象限。
核心结论
- 信号强度下降至3对OCT可靠性的影响相对较小,低于3时进一步下降对可靠性的影响显著增大。
- 信号强度对可靠性的影响在严重青光眼中更大。
- 伪影独立于信号强度导致可靠性下降,且仅影响其所在象限。
- 作者提出循证指南以指导医生判断OCT扫描结果是否可信。
研究设计与样本
- 提取平均RNFL厚度和信号强度,并手动分级扫描伪影。
- 建立多水平线性模型估计真实(预测)RNFL厚度,以模型残差(ΔRNFL)作为可靠性指标。
- 建立含样条和交互项的第二多水平线性模型,以信号强度和伪影为预测因子建模总体及象限特异性可靠性(ΔRNFL)。
主要发现
- 信号强度在10至3之间每降低1点,ΔRNFL为-0.67至-1.25(P < 0.05)。
- 信号强度低于3时每降低1点,ΔRNFL为-15.70至-16.34(P < 0.05)。
- 信号强度在10至3之间每降低1点,严重青光眼眼ΔRNFL为-1.25(P < 0.05),轻中度青光眼眼为-0.67至-0.75(P < 0.05)。
- 伪影对OCT可靠性的影响独立于信号强度(ΔRNFL -4.76;P < 0.05)。
- 伪影仅影响其所在象限的可靠性,一个象限的伪影对相反象限的ΔRNFL无影响(P > 0.05)。
关键数值与推荐
- 信号强度在10至3之间:每降低1点ΔRNFL为-0.67至-1.25(P < 0.05)。
- 信号强度低于3:每降低1点ΔRNFL为-15.70至-16.34(P < 0.05)。
- 严重青光眼眼信号强度在10至3之间每降低1点ΔRNFL为-1.25(P < 0.05),轻中度青光眼眼为-0.67至-0.75(P < 0.05)。
- 伪影:ΔRNFL -4.76(P < 0.05),仅影响其所在象限,对相反象限无影响(P > 0.05)。
临床怎么用
- 作者提出循证指南以指导医生判断OCT扫描结果是否可信。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To assess the impact of OCT signal strength (SS) and artifact on retinal nerve fiber layer (RNFL) measurement reliability and to understand whether glaucoma severity modifies this relationship. DESIGN: Retrospective, longitudinal cohort study. PARTICIPANTS: Two thousand nine hundred ninety-two OCT scans from 474 eyes of 241 patients with glaucoma or glaucoma suspect status. METHODS: We extracted mean RNFL thickness and SS and manually graded scans for artifact. To analyze the effect of SS and artifact on OCT reliability, we (1) created a multilevel linear model using measured RNFL thickness values and demographic and clinical data to estimate the true (predicted) RNFL thickness, (2) calculated model residuals (ΔRNFL) as our reliability measure, and (3) created a second multilevel linear model with splines and interaction terms that modeled overall and quadrant specific reliability (ΔRNFL) as the outcome, using SS and artifact as predictors. MAIN OUTCOME MEASURES: Impact of SS and artifact on ΔRNFL. RESULTS: For SS between 10 and 3, the impact of decreases in SS on OCT reliability is modest (-0.67 to -1.25 ΔRNFL per 1-point decrease in SS; P < 0.05). But at less than 3, changes in SS have a large impact on reliability (-15.70 to -16.34 ΔRNFL per 1-point decrease in SS; P < 0.05). At SS between 10 and 3, decreases in SS tend to have a larger impact on reliability in eyes with severe glaucoma (-1.25 per 1-point decrease in SS; P < 0.05) compared with eyes with mild or moderate glaucoma (-0.67 to -0.75 per 1-point decrease in SS; P < 0.05). The presence of artifact has a significant impact on OCT reliability independent of the effects of SS (-4.76 ΔRNFL; P < 0.05). Artifact affects reliability solely in the quadrant in which it occurs, with artifact in one quadrant showing no impact on ΔRNFL in the opposite quadrant (P > 0.05). CONCLUSIONS: Signal strength decreases down to 3 have relatively mild impacts on OCT reliability. At less than 3, the impact of further decreases in SS on reliability are substantial. The effect of SS on reliability is greater in severe glaucoma. Artifacts result in a decrease in reliability independent of the effect of SS. We propose evidence-based guidelines to guide physicians on whether to trust the results of an OCT scan.