倒立位导致脉络膜扩张与前房变浅:OCT研究

Upside-down position leads to choroidal expansion and anterior chamber shallowing: OCT study.
PubMed 文献 · Br J Ophthalmol · 2019-09-13

该研究纳入34名健康志愿者,采用倒立位1.5分钟作为模型,通过OCT观察脉络膜厚度和前房的动态变化。倒立位后脉络膜厚度显著增加、脉络膜血流指数下降、前房深度和巩膜突500微米处房角开放距离减小,眼压升高。结论提示脉络膜扩张与前房变浅相关,可能与房角关闭的发病机制有关。

要点

  • 倒立位1.5分钟后脉络膜厚度从226.39±52.44 µm增至238.34±54.84 µm(p<0.001)。
  • 倒立位后前房深度从3.21±0.22 mm减至3.13±0.21 mm(p<0.001)。
  • 脉络膜厚度增加与基线脉络膜厚度正相关(p=0.001)。

核心结论

  • 从坐位变为倒立位时,出现脉络膜增厚、前房变浅和眼压升高,伴脉络膜血流减少。
  • 数据提供了脉络膜扩张与前房变浅之间关系的证据,可能与房角关闭的发病机制有关。

研究设计与样本

  • 纳入34名健康志愿者,共分析68只眼。
  • 参与者在倒立机上采取倒立位1.5分钟,作为研究脉络膜厚度动态变化的模型。
  • 在基线、倒立位后及休息15分钟后获取眼压、脉络膜和前房的OCT图像。
  • 比较基线与倒立位之间眼压、前房和脉络膜血流的变化。

主要发现

  • 倒立位1.5分钟后脉络膜厚度从226.39±52.44 µm增加至238.34±54.84 µm(p<0.001)。
  • 倒立位时脉络膜血流指数从0.3357±0.0251降至0.3004±0.0190。
  • 前房深度从3.21±0.22 mm减至3.13±0.21 mm(p<0.001)。
  • 巩膜突500 µm处房角开放距离从0.65±0.24 mm减至0.58±0.20 mm(p=0.007)。
  • Pearson相关分析显示脉络膜厚度增加与基线脉络膜厚度正相关(p=0.001)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: To determine whether dynamic changes in choroidal thickness (CT) cause shallowing of the anterior chamber. METHODS: 34 healthy volunteers were enrolled. The participants in our study adopted the upside-down position for 1.5 min, which was the model we used to study the dynamic changes in CT. Intraocular pressure (IOP) elevation, optical coherence tomography images of the choroid and anterior chamber were obtained at baseline, after being in an upside-down position in an inversion machine and after 15 min of rest. The changes in IOP, anterior chamber and choroidal blood flow between the baseline and the upside-down position were compared. RESULTS: Sixty-eight eyes from 34 subjects were analysed. After being in upside-down position for 1.5 min, there was a significant increase in CT from 226.39±52.44 µm to 238.34±54.84 µm (p<0.001). Choroidal flow index decreased from 0.3357±0.0251 to 0.3004±0.0190 in upside-down position, and there was a decrease in anterior chamber depth (3.21±0.22 mm to 3.13±0.21 mm, p<0.001) and angle opening distance at 500 µm from the scleral spur (0.65±0.24 mm to 0.58±0.20 mm, p=0.007). Pearson correlation analysis showed that the increase in CT was positively related with CT at baseline CT (p=0.001). CONCLUSION: When the body position changed from sitting to upside-down position, there was choroidal thickening, anterior chamber shallowing and IOP elevation with reduced choroidal blood flow. The data provide evidence about the relationship between choroid expansion and shallowing of the anterior chamber, which may be of relevance for the pathogenesis of angle closure.