文献速递青光眼角膜与眼表疾病

Corvis ST新型眼压测量与中央角膜厚度及角膜滞后量的关系

Relationship between novel intraocular pressure measurement from Corvis ST and central corneal thickness and corneal hysteresis.
PubMed 文献 · Br J Ophthalmol · 2019-07-30

本研究纳入141只眼(35只健康眼、106只青光眼眼),比较Goldmann压平眼压计、Corvis ST生物力学矫正眼压与眼反应分析仪测量的眼压同中央角膜厚度及角膜滞后量的关系。结果显示生物力学矫正眼压与中央角膜厚度无显著关联(p=0.19),但与角膜滞后量及角膜阻力因子显著关联。

要点

  • 生物力学矫正眼压与中央角膜厚度无显著关联(p=0.19)。
  • 生物力学矫正眼压与角膜滞后量显著关联(p<0.001)。
  • 所有眼压记录均与角膜阻力因子显著关联(p<0.001)。

核心结论

  • Corvis ST测量的生物力学矫正眼压独立于中央角膜厚度,但依赖于角膜滞后量和角膜阻力因子。

研究设计与样本

  • 共纳入141名受试者的141只眼,其中健康眼35只、青光眼眼106只,分别用Goldmann压平眼压计、Corvis ST和眼反应分析仪测量眼压。
  • 采用线性回归分析眼压测量值(ORA-IOPg、ORA-IOPcc、CST-bIOP和GAT IOP)与生物力学特性(中央角膜厚度、角膜滞后量和角膜阻力因子)的关系。

主要发现

  • IOPg、IOPcc和GAT IOP与中央角膜厚度显著关联(p<0.001),而bIOP与中央角膜厚度无显著关联(p=0.19)。
  • IOPg、bIOP和GAT IOP与角膜滞后量显著关联(IOPg:p<0.001;bIOP:p<0.001;GAT IOP:p=0.0054),而IOPcc与角膜滞后量无显著关联(p=0.18)。
  • 所有眼压记录均与角膜阻力因子显著关联(p<0.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: Corvis ST (CST) yields biomechanical corrected IOP (bIOP) which is purported to be less dependent on biomechanical properties. In our accompanied paper, it was suggested that the repeatability of bIOP is high. The purpose of the current study was to assess the relationship between intraocular pressure (IOP) measured with CST and central corneal thickness (CCT) and corneal hysteresis (CH), in comparison with IOP measured with Goldmann applanation tonometry (GAT) and the ocular response analyzer (ORA). METHODS: A total of 141 eyes from 141 subjects (35 healthy eyes and 106 glaucomatous eyes) underwent IOP measurements with GAT, CST and ORA. The relationships between IOP measurements (ORA-IOPg, ORA-IOPcc, CST-bIOP and GAT IOP) and biomechanical properties (CCT, CH and corneal resistance factor (CRF)) were analysed using the linear regression analysis. RESULTS: IOPg, IOPcc and GAT IOP were significantly associated with CCT (p<0.001), whereas bIOP was not significantly associated with CCT (p=0.19). IOPg, bIOP and GAT IOP were significantly associated with CH (IOPg: p<0.001; bIOP: p<0.001; GAT IOP: p=0.0054), whereas IOPcc was not significantly associated with CH (p=0.18). All of IOP records were associated with CRF (p<0.001). CONCLUSION: The bIOP measurement from CST is independent from CCT, but dependent on CH and CRF.

来源:Br J Ophthalmol,查看原文。版权归原作者所有。

引用

  1. Matsuura M, Murata H, Fujino Y, et al. Relationship between novel intraocular pressure measurement from Corvis ST and central corneal thickness and corneal hysteresis. Br J Ophthalmol. 2019. PMID: 31362932. doi:10.1136/bjophthalmol-2019-314370

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。