使用接触镜进行无创眼压监测

Non-invasive Intraocular pressure monitoring with contact lens.
PubMed 文献 · Br J Ophthalmol · 2019-07-02

本研究在新鲜离体猪眼上测试一种嵌入微通道的新型接触镜,用摄像头记录微通道内指示液体的移动,并与玻璃体腔微压导管测得的真实眼压比较。结果显示接触镜对眼压变化呈一致的线性响应,指示液体平均移动为28微米/毫米汞柱,且能测量周期性波动中眼压的升高和降低。作者认为这种低成本无创传感器基于猪眼模型能可靠监测眼压变化。

要点

  • 在离体猪眼上,嵌入微通道的接触镜对眼压变化呈一致的线性响应。
  • 指示液体在所有试验中的平均移动为28微米/毫米汞柱。
  • 该装置能测量周期性波动中眼压的升高和降低。

核心结论

  • 作者认为这种低成本无创传感器基于猪眼模型能可靠监测眼压变化。

研究设计与样本

  • 研究在新鲜离体猪眼上进行,将接触镜置于猪眼上,用摄像头记录微通道内液体移动。
  • 将微压导管穿入玻璃体腔中心记录真实眼压,并与微通道内指示液体的位移进行比较。

主要发现

  • 接触镜对眼压变化显示一致的线性响应,且对眼间解剖差异稳健。
  • 指示液体在所有试验中的平均移动为28微米/毫米汞柱。
  • 装置能测量周期性波动中眼压的升高和降低。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Glaucoma is the second leading cause of blindness in the world and the first leading cause of irreversible vision loss. Currently, the primary methodology of testing for the intraocular pressure (IOP) is during clinical office hours, which only provide a limited amount of information on the trends and fluctuations of the IOP. Therefore, a continuous monitoring system is required to properly determine the peaks of pressure and to negate any false results obtained by sparse, clinic hour testing. The objective of this study is to determine the ability of a newly designed contact lens with an embedded microchannel, to accurately measure the fluctuations in the IOP. METHODS: Experimentation was completed on fresh enucleated porcine eyes. The contact lens was placed on the porcine eye and utilising a camera the fluid movement, within the microchannel in the contact lens, was recorded. A micro-pressure catheter, threaded into the centre of the vitreous chamber, recorded the true IOP and was compared with the displacement of the indicator fluid within the microchannel. RESULTS: The contact lenses showed a consistent linear responsiveness to changes in IOP and robust to the effects of anatomical differences among eyes. The indicator fluid had an average fluid movement of 28 um/mm Hg between all the trials. Additionally, the devices showed the ability to measure both increases and decreases in IOP during cyclical fluctuations. CONCLUSION: The described inexpensive and non-invasive sensor is able to reliably monitor the IOP changes based on porcine eye model.

来源:Br J Ophthalmol,查看原文。版权归原作者所有。